. 2021 Aug 30;9(9):1124. doi: 10.3390/healthcare9091124

Updated Clinical Evaluation of the CLUNGENE® Rapid COVID-19 Antibody Test

Christopher C Lamb 1,2,3,*, Fadi Haddad 4,5,6, Christopher Owens 7, Alfredo Lopez-Yunez 7, Marion Carroll 8, Jordan Moncrieffe 8
Editor: Joaquim Carreras
PMCID: PMC8470863  PMID: 34574897

Abstract

Background: COVID-19 antibody testing has been shown to be predictive of prior COVID-19 infection and an effective testing tool. The CLUNGENE® SARS-COV-2 VIRUS (COVID-19) IgG/IgM Rapid Test Cassette was evaluated for its utility to aide healthcare professionals. Method: Two studies were performed by using the CLUNGENE® Rapid Test. (1) An expanded Point-of-Care (POC) study at two clinical sites was conducted to evaluate 99 clinical subjects: 62 positive subjects and 37 negative subjects were compared to RT-PCR, PPA, and NPA (95% CI). Sensitivity was calculated from blood-collection time following symptom onset. (2) A cross-reactivity study was performed to determine the potential for false-positive results from other common infections. Results: The specificity of subjects with confirmed negative COVID-19 by RT-PCR was 100% (95% CI, 88.4–100.0%). The sensitivity of subjects with confirmed positive COVID-19 by RT-PCR was 96.77% (95% CI, 88.98–99.11%). In the cross-reactivity study, there were no false-positive results due to past infections or vaccinations unrelated to the SARS-CoV-2 virus. Conclusion: There is a need for a rapid, user-friendly, and inexpensive on-site monitoring system for diagnosis. The CLUNGENE® Rapid Test is a useful diagnostic test that provides results within 15 min, without high-complexity laboratory instrumentation.

Keywords: COVID-19 antibody testing, COVID-19 immunity, COVID-19 serology, CLUNGENE®

1. Introduction

The COVID-19 pandemic SARS-CoV2 COVID-19 has infected over 140 million people worldwide and has caused approximately 3.89 million deaths as of 23 June 2021 [1]; however, some studies suggest that the actual number of global COVID-19 deaths may be about 6.9 million, which is more than double the recorded amount [2,3,4]. In response to the pandemic, the US Food and Drug Administration (FDA) authorized the use of COVID-19 serological tests through Emergency Use Authorizations (EUAs) to make COVID-19 in vitro diagnostic tests widely available to help identify individuals with an adaptive immune response indicating recent or prior infection [5]. Serology tests, or immunoassays, play a significant role in the fight against COVID-19 [6,7]. A prior history of SARS-CoV-2 infection is associated with a lower risk of infection, with an estimated protective effect of up to seven (7) months following primary infection; this supports the conclusion that convalescent plasma with specific antibodies to SARS-CoV-2 has powerful antiviral activity, which can reduce the viral load and mortality in patients with active COVID-19 infection [8,9,10].

There is an urgent need for a rapid, user-friendly, and inexpensive on-site monitoring system for diagnosis [11]. The CLUNGENE® SARS-COV-2 VIRUS (COVID-19) IgG/IgM Rapid (15 min) Test Cassette has been commercially available in the US under an FDA-approved Emergency Use Authorization (EUA201121) [12] and Europe (CE Mark reference 02PBJ267 dated 9 March 2020). The CLUNGENE® test has been previously studied, including the use of the test in the offices of general practitioners, evaluating the presence of antibodies in convalescent plasma donors, and how the test performs at a point of care facility [13,14,15]. The aim of this research was to better understand the sensitivity and specificity of the CLUNGENE® assay and the potential for false-positive results related to infections or vaccinations not linked to the SARS-CoV-2 virus.

2. Materials and Methods

2.1. Study #1

2.1.1. Design

In an initial study, a single Point-of-Care (POC) facility was used to estimate the sensitivity and specificity of the CLUNGENE® test [16]. The study was expanded to a second independent site. The two sites were Sharp Healthcare, a not-for-profit multi-center regional healthcare group located in San Diego, CA; and Alivio Medical Center, an urgent care/primary care center in Indianapolis, IN. Samples used for RT-PCR were nares swabs. Positive subjects were symptomatic for SAR-CoV-2 Virus infected and confirmed with RT-PCR positive tested nares swabs. Negative subjects were asymptomatic, from high-risk areas, and confirmed with negative RT-PCR tested nares swabs. Finger-prick whole-blood samples were used for SARS-COV-2 Virus IgG/IgM detection. The comparator method for RT-PCR was either Cobas Roche SARS-COV2 RT-PCR (Roche Diagnostics, 9115 Hague Road PO Box 50457, Indianapolis, IN 46250) or Thermo Fisher TaqPath COVID-19 Combo Kit (Thermo Fisher Scientific, 168 3rd Ave, Waltham, MA 02451, USA). Trained operators with no prior information about each subject drew samples. Subject inclusion criteria included individuals with a confirmed COVID-19 test result by SARS-CoV-2 RT-PCR. Subjects were excluded if they were unable to provide informed consent due to mental or cognitive disabilities.

2.1.2. Methods

The CLUNGENE® Point-of-Care test was run according to the manufacturer’s instructions (see Figure 1). The test result was read after 15 min. Days from symptom onset were captured from an electronic medical record which documented self-reported data from patients reporting on the number of days they had been sick at the time of study enrollment.

Figure 1.

Figure 1

CLUNGENE® Point-of-Care test manufacturer’s instructions.

Categorical variables were compared using the chi-squared or Fisher exact test, and continuous variables were compared using the Student t-test or Mann-Whitney U test, as appropriate. All tests were two tailed, and p < 0.05 was considered statistically significant. SPSS Statistics, IBM SPSS software, version 27.0 (SPSS, Inc., Chicago, IL, USA) was used for all calculations.

The CLUNGENE® test can be stored at ambient temperature between 4 and 30 degrees Celsius. The product is packaged in a box with 25 test kits, with each kit wrapped in an individual foil patch for proper storage and transportation. The test kit itself should be kept away from direct sunlight.

2.2. Study #2

2.2.1. Design

An initial cross-reactivity study was performed in which various common infectious agents were tested for potential false-positive results (Table 1). All tests were negative. In addition, as recommended by the US Food and Drug Administration [17], a follow-up study tested the cross-reactivity of the CLUNGENE® device to antibodies to common coronaviruses that are not SARS-CoV-2: anti-229E, anti-NL63, anti-OC43, and anti-HKU1; and those for which there is a high rate of vaccinations and/or infection in the US, i.e., anti-Haemophilus influenzae IgG and IgM. The testing was performed at Medcomp Sciences, an independent clinical medical laboratory [18].

Table 1.

Common infectious agents used in cross-reactivity study.

Analog
Anti-HBe, anti-HBc and HBsAg
Anti-Hepatitis C Virus (HCV)
Anti-Human Immunodeficiency Virus (HIV)-1
Anti-HIV-2
Anti-influenza A IgG
Anti-influenza B IgG
Anti-influenza A IgM
Anti-influenza B IgM
Anti-respiratory syncytial virus (RSV) IgG
Anti-respiratory syncytial virus (RSV) IgM
Anti-Mycoplasma pneumoniae (MP) IgM
Anti-Chlamydia pneumoniae (CP) IgM
Human parainfluenza virus PCR positive (Paired Convalescent)
Anti-Treponema pallidum (TP)
Rheumatoid factor (RF) 521.00 IU/mL
RF 342.50 IU/mL
RF 347.00 IU/mL
RF 310.00 IU/mL
RF 565.00 IU/mL
RF 125.00IU/mL
RF 796.00 IU/mL
RF 500.00 IU/mL
RF 825.00 IU/mL
RF 144.00 IU/mL
RF 158.00 IU/mL
RF 122.00 IU/mL
RF 197.00 IU/mL
RF 146.00 IU/mL
Antinuclear antibodies (ANA) 1:240
Common human pathogenic coronaviruses:
HCoV-HKU1
HCoV-NL63
HCoV-OC43
HCoV-229E

2.2.2. Methods

The CLUNGENE® Point-of-Care test was run according to same manufacturer’s instructions as was performed with Study #1. Samples (50) used for the cross-reactivity study were obtained from Trina Bioreactives, Ag; Trina Bioreactives, Ag, (Grabenstrasse 8, 8606 Nänikon, Switzerland) is an ISO certified company providing specialized Sars-COV-2 in vitro diagnostic biomaterials [19]. The samples were collected under a protocol approved by the ethics committee of the National Medical Association, Baden-Wurttemburg, Germany (file #F-2012-027, “Plasma Samples for Studies”).

Test results were read after 15 min. Five (5) serum samples that were positive for IgM and five for IgG were analyzed in three (3) separate CLUNGENE® Point-of-Care test lots with each of the below antibodies:

  • Anti-Haemophilus influenzae IgM,

  • Anti-Haemophilus influenzae IgG,

  • IgG, anti-coronavirus 229E IgG,

  • Anti-coronavirus NL63 IgG,

  • Anti-coronavirus OC43 IgG,

  • Anti-coronavirus HKU1 IgG,

  • Anti-coronavirus 229E IgM,

  • Anti-coronavirus NL63 IgM,

  • Anti-coronavirus OC43 IgM,

  • Anti-coronavirus HKU1 IgM.

Cross-reactivity was determined by using Beckman Coulter UniCel DxI Access Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids (Table 2).

Table 2.

Reagents: reference numbers C58961 and C58957 Beckman Coulter Immunoassay System.

Reagent Reference Number Material Type
Access SARS-CoV-2 IgG Calibrator Ref. No. C58963
Access SARS-CoV-2 IgG Calibrator Ref. No. C58958
Access SARS-CoV-2 IgG QC Ref. No. C58964 Quality Control (QC)
Access SARS-CoV-2 IgM QC Ref. No. C58959 Quality Control (QC)
Access Substrate Ref. No. 81906 Quality Control (QC)
Access Wash Buffer II Ref. No. A16792 Quality Control (QC)
UniCel DxI Wash Buffer II Ref. No. A16793 Quality Control (QC)

All QC materials were supplied and used as indicated by the manufacturer. A minimum of 2 levels of QC were processed per run in accordance with Westgard rules for 6-sigma quality requirements. QC results must fall within 3 standard deviations of historical data, as recorded in Levy–Jennings plots by the system. All QC results were documented and verified by Clinical Laboratory Scientist before processing samples (Table 3).

Table 3.

Quality control for DXI600 analysis of cross-reactivity.

Rack/Pos. Sample ID Patient/Lot ID Type Dilution Test Calibrator Results 1 RLUs 2 Completion
207/1 COV-2IgG QC1 922605 Serum COV2G Non-Reactive 0.02 S/CO 7143 27 April 2021 10:20 AM
207/2 COV-2IgG QC2 922605B Serum COV2G Reactive 2.49 S/CO 995255 27 April 2021
10:20 AM
207/3 COV-2IgM QC1 922821 Serum COV2M Non-Reactive 0.16 S/CO 11220 27 April 2021
10:27 AM
COV2M(2) Non-Reactive 0.16 S/CO 11053 27 April 2021
10:27 AM
207/4 COV-2IgM
QC2 922821
Serum COV2M Reactive 4.23 S/CO 301471 27 April 2021
10:28 AM
COV2M(2) Reactive 2.62 S/CO 186854 27 April 2021
10:27 AM

1 S/CO: Signal/Cutoff. 2 RLUs: Relative Light Units.

Qualitative results are reported as Positive/Reactive or Negative/Non-Reactive, depending on whether the analyte in question is above or below the established signal cutoff value (S/CO). Each analyte tested may have a different cutoff value, but each cutoff value is determined during calibration of the instrument, using the value of the calibrator.

3. Results

3.1. Results of Study #1

An analysis was run on 99 patients who completed the study (Table 4).

Table 4.

Antibody percent agreement.

Group Days Post
Symptom Onset
#PCR Total Candidate Device Results
Results PPA 95% CI
IgG Positive 0~7 4 2 50.00% 15.00~85.00%
8~14 23 23 100.00% 85.69~100.00%
≥15 35 33 94.28% 81.39~98.42%
Total 62 58 93.55% 84.55~97.46%
IgM Positive 0~7 4 3 75.00% 30.06~95.44%
8~14 23 21 91.30% 73.21~97.58%
≥15 35 27 77.14% 60.98~87.93%
Total 62 51 82.26% 70.96~89.79%
IgG/IgM Combined Antibody Positive 0~7 4 3 75.00% 30.06~95.44%
8~14 23 23 100.00% 85.69~100.00%
≥15 35 34 97.14% 85.47~99.49%
Total 62 60 96.77% 88.98~99.11%
IgG Negative N/A 37 37 100.00% 90.60~100.00%

Thirty-seven (37) patients who had negative COVID-19 RT-PCR were tested and found antibody negative by using the CLUNGENE® test (95% CI, 90.60~100.00%). Three (3) out of four (4), or 75%, of RT-PCR positive COVID-19 subjects tested prior to day 7 of symptom onset were antibody positive (95% CI, 30.06~95.44%). Twenty-three (23), or 100%, of RT-PCR positive COVID-19 subjects tested positive between day 8 and 14 from symptom onset (95% CI, 85.69~100.00%). A total of 33 out of 35, or 94.28%, of RT-PCR positive COVID-19 subjects tested positive after day 14 from symptom onset were antibody positive (95% CI, 30.06~95.44%). In all 62 patients with confirmed COVID-19 with RT-PCR, the combined sensitivity of IgM and IgG was 96.77% (95% CI, 88.98–99.11%), meaning that there was 96.77 positive agreement between a positive RT-PCR test and a positive antibody test. The specificity was 100% (95% CI, 88.4–100.0%), meaning there was 100% agreement between a negative RT-PCR test and 100% negative antibody result. These results are displayed in Table 5.

Table 5.

Line Table Data.

# and Site Subject ID Age Gender
(F/M)
Whole Blood Specimen Collection Date Days after Symptom Onset CLUNGENE® Rapid Test Result
(Pos/Neg)
PCR Test Date PCR
Confirmation Result
IgM IgG
1a 007sgh 66 F 1 June 2020 Neg Neg 30 May 2020 Neg
2a 008sgh 65 F 1 June 2020 Neg Neg 29 May 2020 Neg
3a 009sgh 72 M 1 June 2020 Neg Neg 31 May 2020 Neg
4a 010sgh 67 F 5 June 2020 Neg Neg 27 May 2020 Neg
5a 011sgh 44 M 5 June 2020 Neg Neg 1 June 2020 Neg
6a 012sgh 31 F 5 June 2020 Neg Neg 1 June 2020 Neg
7a 013sgh 77 F 5 June 2020 Neg Neg 1 June 2020 Neg
8a 024smh 37 M 5 June 2020 11 Neg Pos 2 June 2020 Pos
9a 025sgh 89 M 5 June 2020 Neg Neg 11 June 2020 Neg
10a 023sgh 91 F 6 June 2020 Neg Neg 9 June 2020 Neg
11a 024sgh 81 M 8 June 2020 Neg Neg 9 June 2020 Neg
12a 026sgh 59 M 8 June 2020 Neg Neg 11 June 2020 Neg
13a 019sgh 69 F 9 June 2020 20 Neg Pos 23 May 2020 Pos
14a 028sgh 58 M 10 June 2020 Neg Neg 11 June 2020 Neg
15a 021sgh 63 F 11 June 2020 13 Neg Pos 6 June 2020 Pos
16a 027sgh 77 F 11 June 2020 Neg Neg 1 June 2020 Neg
17a 029sgh 29 M 12 June 2020 8 Pos Pos 8 June 2020 Pos
18a 033sgh 69 M 12 June 2020 Neg Neg 5 June 2020 Neg
19a 034sgh 53 M 12 June 2020 Neg Neg 11 June 2020 Neg
20a 035sgh 77 M 12 June 2020 Neg Neg 11 June 2020 Neg
21a 036sgh 74 M 12 June 2020 Neg Neg 10 June 2020 Neg
22a 037sgh 30 F 12 June 2020 Neg Neg 9 June 2020 Neg
23a 038sgh 22 M 12 June 2020 Neg Neg 6 June 2020 Neg
24a 001sgh 38 M 15 June 2020 26 Pos Pos 23 May 2020 Pos
25a 006sgh 53 M 18 June 2020 23 Neg Pos 29 May 2020 Pos
26a 027smh 47 M 18 June 2020 14 Pos Pos 14 June 2020 Pos
27a 039sgh 60 M 18 June 2020 13 Pos Pos 14 June 2020 Pos
28a 043sgh 22 F 18 June 2020 18 Neg Pos 14 June 2020 Pos
29a 040sgh 70 M 19 June 2020 17 Pos Pos 12 June 2020 Pos
30a 044sgh 32 F 19 June 2020 Neg Neg 17 June 2020 Neg
31a 045sgh 66 M 19 June 2020 Neg Neg 10 June 2020 Neg
32a 046sgh 65 M 19 June 2020 Neg Neg 9 June 2020 Neg
33a 050sgh 31 M 26 June 2020 8 Pos Pos 18 June 2020 Pos
34a 051sgh 44 M 26 June 2020 10 Pos Pos 21 June 2020 Pos
35a 052sgh 35 M 26 June 2020 7 Pos Pos 19 June 2020 Pos
36a 055sgh 57 M 26 June 2020 Neg Neg 21 June 2020 Neg
37a 056sgh 69 F 26 June 2020 Neg Neg 22 June 2020 Neg
38a 057sgh 67 M 26 June 2020 Neg Neg 21 June 2020 Neg
39a 058sgh 59 F 26 June 2020 Neg Neg 17 June 2020 Neg
40a 059sgh 39 M 26 June 2020 6 Neg Neg 20 June 2020 Pos
41a 060sgh 28 F 26 June 2020 Neg Neg 24 June 2020 Neg
42a 061sgh 51 F 26 June 2020 Neg Neg 23 June 2020 Neg
43a 062sgh 71 F 26 June 2020 11 Pos Pos 23 June 2020 Pos
44a 072sgh 62 M 2 July 2020 17 Pos Pos 15 June 2020 Pos
45a 031smh 31 F 2 July 2020 22 Pos Pos 11 June 2020 Pos
46a 066sgh 43 F 2 July 2020 15 Pos Neg 24 June 2020 Pos
47a 068sgh 60 M 2 July 2020 12 Pos Pos 25 June 2020 Pos
48a 075sgh 62 F 2 July 2020 7 Pos Pos 25 June 2020 Pos
49a 081sgh 73 M 9 July 2020 9 Pos Pos 4 July 2020 Pos
50a 082sgh 63 M 9 July 2020 11 Pos Pos 3 July 2020 Pos
51a 084sgh 68 M 9 July 2020 15 Pos Pos 2 July 2020 Pos
52a 088sgh 56 F 9 July 2020 10 Pos Pos 7 July 2020 Pos
53a 090sgh 88 F 9 July 2020 20 Neg Pos 7 July 2020 Pos
54a 089sgh 56 F 17 July 2020 16 Pos Pos 4 July 2020 Pos
55a 094sgh 19 F 17 July 2020 15 Pos Pos 11 July 2020 Pos
56a 096sgh 36 M 17 July 2020 27 Pos Pos 14 July 2020 Pos
57a 097sgh unknown F 17 July 2020 Neg Neg 14 July 2020 Neg
58a 098sgh unknown M 17 July 2020 Neg Neg 8 July 2020 Neg
59a 099sgh 48 F 21 July 2020 9 Pos Pos 18 July 2020 Pos
60a 100sgh 48 F 21 July 2020 14 Pos Pos 17 July 2020 Pos
61a 101sgh 72 M 21 July 2020 16 Neg Neg 25 July 2020 Pos
62a 053sgh 50 F 23 July 2020 27 Neg Pos 26 June 2020 Pos
63a 073sgh 38 M 23 July 2020 34 Pos Pos 30 June 2020 Pos
64a 103sgh 29 M 24 July 2020 12 Pos Pos 15 July 2020 Pos
65a 104sgh 40 M 24 July 2020 21 Pos Pos 11 July 2020 Pos
66a 106sgh 45 M 28 July 2020 15 Neg Pos 22 July 2020 Pos
67a 113sgh 46 F 31 July 2020 7 Pos Neg 25 July 2020 Pos
68a 114sgh 79 F 4 August 2020 12 Pos Pos 30 July 2020 Pos
69a 110sgh 51 F 5 August 2020 15 Pos Pos 29 July 2020 Pos
70a 109sgh 63 M 7 August 2020 19 Pos Pos 25 July 2020 Pos
71a 119sgh 60 F 8 August 2020 10 Pos Pos 7 August 2020 Pos
72a 105sgh 38 M 12 August 2020 31 Pos Pos 20 July 2020 Pos
73a 120sgh 73 M 13 August 2020 22 Pos Pos 26 July 2020 Pos
74a 123sgh 47 M 13 August 2020 12 Pos Pos 10 August 2020 Pos
75a 124sgh 80 M 13 August 2020 30 Pos Pos 20 July 2020 Pos
76a 118sgh 94 F 14 August 2020 15 Neg Pos 31 July 2020 Pos
77a 117sgh 28 F 14 August 2020 14 Pos Pos 5 August 2020 Pos
78a 121sgh 62 F 17 August 2020 14 Pos Pos 12 August 2020 Pos
1b C011 29 F 12 December 2020 14 Pos Pos 30 November 2020 Pos
2b A005 44 F 12 December 2020 Neg Neg 6 December 2020 Neg
3b C020 30 F 12 December 2020 14 Pos Pos 28 November 2020 Pos
4b C012 52 M 14 December 2020 14 Pos Pos 5 December 2020 Pos
5b C032 46 M 14 December 2020 Neg Neg 10 December 2020 Neg
6b C033 18 M 14 December 2020 Neg Neg 10 December 2020 Neg
7b C004 74 M 19 December 2020 26 Pos Pos 3 December 2020 Pos
8b C030 28 M 19 December 2020 Neg Neg 16 December 2020 Neg
9b C016 39 F 21 December 2020 19 Pos Pos 10 December 2020 Pos
10b C031 18 F 21 December 2020 Neg Neg 16 December 2020 Neg
11b C009 53 F 29 December 2020 18 Pos Pos 19 December 2020 Pos
12b C010 55 M 29 December 2020 19 Pos Pos 19 December 2020 Pos
13b C028 46 M 29 December 2020 24 Pos Pos 18 December 2020 Pos
14b C005 19 F 31 December 2020 21 Pos Pos 18 December 2020 Pos
15b C006 19 F 31 December 2020 21 Pos Pos 18 December 2020 Pos
16b C007 42 F 31 December 2020 21 Pos Pos 18 December 2020 Pos
17b C019 42 F 31 December 2020 15 Pos Pos 18 December 2020 Pos
18b C026 51 F 31 December 2020 Neg Neg 31 December 2020 Neg
19b C027 43 F 31 December 2020 Neg Neg 31 December 2020 Neg
20b C029 54 M 31 December 2020 15 Pos Pos 26 December 2020 Pos
21b C013 40 M 4 January 2021 28 Pos Pos 15 December 2020 Pos

The positive predictive value can be calculated, but the result is dependent on the prevalence of disease in the community. If a test for a disease has 96.77% sensitivity and 100% specificity, and the disease prevalence is 10%, the positive predictive value (PPV) is 100%, and the negative predictive value (NPV) is 99.64%. At the time of this publication, the positivity rate for the nares SARS CoV2 RT-PCR tests was between 8.2% and 10% in San Diego, CA (San Diego County), and Indianapolis, IN (Marion Country) [20].

3.2. Results of Study #2

The test results of negative quality-control samples were all negative, and the test results of the positive quality-control samples were all positive. The consistency rate of cross-reactivity of negative samples was 100%.

4. Discussion

In the first study, the performance characteristics of CLUNGENE® were evaluated and showed a specificity of 100% and a sensitivity of 96.77%. In the second study, there were no false-positive results due to past infections or vaccinations unrelated to the SARS-CoV-2 virus. These results are in line with the new European Commission’s Medical Device Coordination Group requirements for rapid COVID-19 antibody tests and consistent with previously published results [21,22,23,24,25].

Antibody testing is a useful aid to confirm past infection [26]. Recent findings confirm that antibody testing is predictive of prior COVID-19 infection, and rapid screening methods—even from finger pricks—are effective testing tools [27]. However, we see the potential for a much broader use and recommend a combined approach that uses both RT-PCR and serological testing. The advantage of the CLUNGENE® antibody test is its simplicity, since there is no need for specialized laboratory personnel to perform and interpret results. The low rate of false positivity makes this test ideal to rule in disease and eliminate the need for further RT-PCR testing if seroconversion occurs, since the CLUNGENE® antibody test can diagnose most infected COVID-19 patients. If the test is negative, a recommendation should be made to have a follow-up RT-PCR test.

Serology testing also has the potential to monitor the presence of antibodies. Studies confirm that a prior history of SARS-CoV-2 infection is associated with a lower risk of infection, with an estimated seven (7) month protective effect [9,10]. The association of SARS-CoV-2 Seropositive Antibody Test with Risk of Future Infection has now been established [28]. It is clear that titers of IgM and IgG antibodies against the receptor-binding domain (RBD) of the spike protein of SARS-CoV-2 decrease significantly between one (1) and seven (7) months, and concurrently, neutralizing activity decreases [29]. Monitoring the presence of antibodies from past infection could assist healthcare professionals in assessing the likely presence of neutralizing activity and immunity when managing patient care.

Given widespread availability, COVID-19 serology testing, similar to other infectious diseases, can now become routine [30]. Consideration should be given to adding COVID-19 antibody testing to the WHO List of Essential In Vitro Diagnostics (EDL), which now consists of 122 test categories, including most serious infectious diseases [31]. A rapid 15-min COVID-19 assay offers the ability to do this testing quickly and efficiently—tests are now available for less than $5 and routinely reimbursed in the US by public and private insurers [5].

In addition, the ability of the CLUNGENE® antibody test to detect antibodies to the coronavirus’s spike protein’s receptor binding domain means it has the potential to assess the efficacy of most vaccines, as well as convalescent plasma therapy [32]. Countries in Europe are now using antibody testing to determine if a second COVID-19 vaccine dose is required if a patient has a prior infection based on a positive antibody test [33]. Recently, airports and Blood Banks have been providing COVID-19 antibody testing services to determine whether a person has developed immunity to COVID-19 through vaccination or through contracting the virus previously [34,35]. Some countries, including China, require an antibody test. Limited evaluation of the CLUNGENE® antibody test has confirmed positive antibody test results following patients who have been vaccinated [36]. Pfizer’s recent data suggest that its vaccine is efficacious for only 6 months and that a third shot within 12 months is likely needed [37]. Furthermore, there is a potential issue regarding vaccine efficacy for recipients who do not receive a full dose; the US CDC estimates that 3% of vaccinated people who received a first dose did not receive a recommended second dose [38]. COVID-19 vaccination also fails to stimulate an immune response in many blood-cancer patients or those otherwise immunocompromised [39,40]. Additional studies are needed to confirm the efficacy of serology testing to monitor vaccine effectiveness.

Limitations

Limitations of the study include a small sample size from two geographic areas. The study also did not include special groups, such as pregnant women or children. The subjectivity of symptom reporting by patients can be a confounding factor in determining the duration of illness. Some patients may have been symptomatic for a different time period than they recalled. Lastly, the CLUNGENE® antibody test has not been compared with another test in the study.

5. Conclusions

In a pandemic crisis with significant economic and health implications, this study confirms the utility of serological testing for COVID-19 disease diagnosis providing rapid test results with a relatively high degree of sensitivity and specificity. Furthermore, given recent data regarding the relationship between positive serology and immunity, routine testing can be a useful tool to monitor antibody status for optimal patient care. Tests such as the CLUNGENE® SARS-COV-2 VIRUS (COVID-19) IgG/IgM Rapid Test Cassette can assist healthcare professionals to help identify individuals with an adaptive immune response indicating recent or prior infection, as intended by the US FDA under an EUA.

Acknowledgments

The author acknowledges Ryan Dagenais for editorial support with the manuscript; his efforts were funded by BioSolutions Services LLC.

Author Contributions

Conceptualization, C.C.L.; methodology, C.C.L., F.H. and M.C.; software, F.H., C.O., M.C. and J.M.; validation, F.H., C.O., A.L.-Y. and M.C.; formal analysis, C.C.L., F.H., C.O. and M.C.; investigation, F.H., C.O., A.L.-Y., M.C. and C.C.L.; resources, all; data curation, F.H., C.O., M.C., J.M. and C.C.L.; writing—original draft preparation, C.C.L.; writing—review and editing, M.C., C.O., F.H. and A.L.-Y.; visualization, C.C.L.; supervision, C.C.L.; project administration, C.C.L.; funding acquisition, C.C.L. All authors have read and agreed to the published version of the manuscript.

Funding

This research was partially funded by Clongene and Speranza.

Institutional Review Board Statement

The study was conducted according to the guidelines of the Declaration of Helsinki, and approved by the Institutional Review Board of Sharp Center for Research (protocol ID: IgG/IgM COVID19; IRB#: 2005801; 21 May 2020). This was a formal IRB approved clinical study conducted within Sharp Healthcare, a not-for-profit multicenter regional healthcare group located in San Diego, California. Subjects were included if hospitalized or recently discharged following a SARS-CoV-2 RT-PCR nares test. A study protocol and informed consent were initiated and approved by the Sharp Institutional Review Board. Subjects were included if they were >18 years of age and understood the study and its requirements. Patients who had impairment of cognition or decision-making capacity were excluded. Subjects were screened by research coordinators to determine if they had a nares SARS-CoV-2 RT-PCR test result, and then consent was requested to enroll them in the study.

Informed Consent Statement

Informed consent was obtained from all subjects involved in the study. Written informed consent was obtained from the patient(s) to publish this paper.

Data Availability Statement

The data presented in this study are available on request from the corresponding author. The data are not publicly available.

Conflicts of Interest

Christopher C. Lamb, PhD, has worked with the manufacturers of SARS-CoV-2 tests for Emergency Use Authorization submissions to the US FDA. The funders had no role in the design of the study; in the collection, analyses, or interpretation of data; in the writing of the manuscript; or in the decision to publish the results.

Footnotes

Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The data presented in this study are available on request from the corresponding author. The data are not publicly available.