Fig. 3. Selectivity and stability of the biosensing device in different environments.

Fig. 3

a Cyclic voltammograms of the biosensing device in PBS containing 5 mM glucose. Scanning rates: 30, 50, 70, and 100 mV/s. b Selectivity of the biosensing device to different interfering substances. c Measurement repeatability of the biosensing device. Glucose concentration: 5 mM in PBS. d Temperature stability of the biosensing device from 20 to 45 °C. Glucose concentration: 5 mM in PBS. e The pH stability of the biosensing device from pH 6 to 9. Glucose concentration: 5 mM in PBS. f The storage stability of the biosensing device over 7 days of storage. Glucose concentration: 5 mM in PBS (The sensor was measured three times under each condition, as shown in df)