Table 1:

Sensitivity of direct RT-LAMP, RT-RPA, and Closed-Tube Penn-RAMP for the detection of SARS-CoV-2 eluted from nasal swabs (contrived samples)*.

Swab elution media Spiked samples RT-LAMP** RT-RPA** Penn-RAMP**
VTM 10000 particles/reaction 4/4 4/4 4/4
1000 particles/reaction 4/4 4/4 4/4
100 particles/reaction 2/4 1/4 4/4
10 particles/reaction 0/4 0/4 0/4
0 particles/reaction 0/4 0/4 0/4

Water 10000 particles/reaction 4/4 4/4 4/4
1000 particles/reaction 4/4 4/4 4/4
100 particles/reaction 4/4 4/4 4/4
10 particles/reaction 0/4 0/4 3/4
0 particles/reaction 0/4 0/4 0/4
*

We submerged a nasal swab in 3 mL of VTM and water and spike SARS-CoV-2 into each of them to prepare contrived nasal swab samples.

**

The table documents the fraction of positive results.