Table 1:
Sensitivity of direct RT-LAMP, RT-RPA, and Closed-Tube Penn-RAMP for the detection of SARS-CoV-2 eluted from nasal swabs (contrived samples)*.
Swab elution media | Spiked samples | RT-LAMP** | RT-RPA** | Penn-RAMP** |
---|---|---|---|---|
VTM | 10000 particles/reaction | 4/4 | 4/4 | 4/4 |
1000 particles/reaction | 4/4 | 4/4 | 4/4 | |
100 particles/reaction | 2/4 | 1/4 | 4/4 | |
10 particles/reaction | 0/4 | 0/4 | 0/4 | |
0 particles/reaction | 0/4 | 0/4 | 0/4 | |
| ||||
Water | 10000 particles/reaction | 4/4 | 4/4 | 4/4 |
1000 particles/reaction | 4/4 | 4/4 | 4/4 | |
100 particles/reaction | 4/4 | 4/4 | 4/4 | |
10 particles/reaction | 0/4 | 0/4 | 3/4 | |
0 particles/reaction | 0/4 | 0/4 | 0/4 |
We submerged a nasal swab in 3 mL of VTM and water and spike SARS-CoV-2 into each of them to prepare contrived nasal swab samples.
The table documents the fraction of positive results.