TABLE 2.
Determination of the low limit detection level of CA v2.0
No. of HCV RNA copies/mla | Detection of:
|
|
---|---|---|
HCV RNAb (OD660) | IC | |
6 × 103 | Pos. (3.861) | Pos. |
103 | Pos. (3.556) | Pos. |
102 | Pos. (0.613) | Pos. |
Pos. (1.633) | Pos. | |
Pos. (0.498) | Pos. | |
30 | Neg. (0.008) | Pos. |
Neg. (0.007) | Pos. | |
Neg. (0.007) | Pos. |
Series of diluted specimens containing known amounts of HCV RNA were obtained by diluting a calibrated serum (Accurun HCV RNA).
Crude readings of the optical density at 660 nm (OD660) are furnished by Cobas. Samples are positive (Pos.) with optical densities of >0.15 and negative (Neg.) below this value.