Skip to main content
. 1999 May;37(5):1567–1569. doi: 10.1128/jcm.37.5.1567-1569.1999

TABLE 2.

Determination of the low limit detection level of CA v2.0

No. of HCV RNA copies/mla Detection of:
HCV RNAb (OD660) IC
6 × 103 Pos. (3.861) Pos.
103 Pos. (3.556) Pos.
102 Pos. (0.613) Pos.
Pos. (1.633) Pos.
Pos. (0.498) Pos.
30 Neg. (0.008) Pos.
Neg. (0.007) Pos.
Neg. (0.007) Pos.
a

Series of diluted specimens containing known amounts of HCV RNA were obtained by diluting a calibrated serum (Accurun HCV RNA). 

b

Crude readings of the optical density at 660 nm (OD660) are furnished by Cobas. Samples are positive (Pos.) with optical densities of >0.15 and negative (Neg.) below this value.