TABLE 1.
Analysis of Digene HC II CT-ID and AMPLICOR CT PCR performance with endocervical specimens before and after discrepant analysis (n = 587)
Test and result | No. of samples with result:
|
|||
---|---|---|---|---|
Before discrepant analysis
|
After discrepant analysisa
|
|||
Culture positive | Culture negative | Confirmed positive | Confirmed negative | |
Digene HC II CT-ID | ||||
Positive | 50 | 17 | 62 | 5b |
Negative | 3 | 517 | 3 | 517 |
AMPLICOR CT PCR | ||||
Positive | 47 | 14 | 59 | 2b |
Negative | 6 | 520 | 6 | 520 |
A specimen was considered true positive in discrepant analysis if it was culture positive or positive by both Digene HC II CT-ID and Amplicor CT PCR.
Specimens were also negative by DFA staining.