To 1000 Gy and back again: a systematic review on dose-response evaluation in selective internal radiation therapy for primary and secondary liver cancer

Joey Roosen 1, Nienke J M Klaassen 1,#, Lovisa E L Westlund Gotby 1,#, Christiaan G Overduin 1, Marcel Verheij 2, Mark W Konijnenberg 1,3, J Frank W Nijsen 1,
PMCID: PMC8484215  PMID: 33839892

Abstract

Purpose

To systematically review all current evidence into the dose-response relation of yttrium-90 and holmium-166 selective internal radiation therapy (SIRT) in primary and secondary liver cancer.

Methods

A standardized search was performed in PubMed (MEDLINE), Embase, and the Cochrane Library in order to identify all published articles on dose-response evaluation in SIRT. In order to limit the results, all articles that investigated SIRT in combination with other therapy modalities (such as chemotherapy) were excluded.

Results

A total of 3038 records were identified of which 487 were screened based on the full text. Ultimately, 37 studies were included for narrative analysis. Meta-analysis could not be performed due to the large heterogeneity in study and reporting designs. Out of 37 studies, 30 reported a ‘mean dose threshold’ that needs to be achieved in order to expect a response. This threshold appears to be higher for hepatocellular carcinoma (HCC, 100–250 Gy) than for colorectal cancer metastases (CRC, 40–60 Gy). Reported thresholds tend to be lower for resin microspheres than when glass microspheres are used.

Conclusion

Although the existing evidence demonstrates a dose-response relationship in SIRT for both primary liver tumours and liver metastases, many pieces of the puzzle are still missing, hampering the definition of standardized dose thresholds. Nonetheless, most current evidence points towards a target mean dose of 100–250 Gy for HCC and 40–60 Gy for CRC. The field would greatly benefit from a reporting standard and prospective studies designed to elucidate the dose-response relation in different tumour types.

Supplementary Information

The online version contains supplementary material available at 10.1007/s00259-021-05340-0.

Keywords: SIRT, Radioembolization, TARE, Yttrium, Holmium, Dosimetry

Introduction

Selective internal radiation therapy (SIRT) is a treatment modality for primary and metastatic liver cancer that has been incorporated into clinical practice worldwide for over two decades [1, 2]. During SIRT, microspheres loaded with the beta-emitting isotopes yttrium-90 (90Y) or holmium-166 (166Ho) are administered through a microcatheter positioned in the hepatic artery. The heterogeneous dose distribution that the microspheres provide to the liver has been under investigation ever since the first implementation of SIRT in clinical studies, typically expressed in a tumour to normal tissue ratio (T/N ratio). In the early days of SIRT, the procedure was performed through laparotomy followed by direct injection of 90Y microspheres into the hepatic artery. The dose distribution could then be measured with a beta probe directly at the liver surface, resulting in T/N ratios up to 45:1 [1]. More recently, a second level of dose heterogeneity has become a topic of interest: the heterogeneous dose distribution within a tumour, typically visualized in a dose-volume histogram (DVH) [3, 4]. Finally, the biological effect of SIRT on healthy liver tissue and its impact on toxicity is an important factor. Compared to the dose-response studies in SIRT, relatively few studies have addressed this issue. The dose distribution on healthy tissue in relation to toxicity is however beyond the scope of this review.

Clinical results of SIRT have always been variable between patients, stressing the need for further in-depth dosimetry and dose-response analysis, as the mean tumour dose has often been correlated to both tumour response and survival [5, 6]. Most research has been performed with 90Y microspheres (Eβ-max = 2.28 MeV (100%)), of which two (vastly different) commercial products are available: glass microspheres with a high specific activity (up to 5000 Bq per microsphere) and resin microspheres with a lower specific activity (50 Bq per microsphere) [7, 8]. As a direct result of the difference in specific activity, a much lower number of microspheres is administrated during treatment with glass microspheres compared to resin microspheres (1.2 million vs. 40–80 million) [7]. Despite the differences in microsphere distribution that could be expected between the two different yttrium products as a result of the mentioned characteristics, both have been shown to be similarly efficacious [9, 10].

During planning angiography days to weeks prior to treatment, technetium-99m-labelled albumin macroaggregates (99mTc-MAA) are administered. The resulting 99mTc-MAA SPECT imaging has also been used for dosimetry, even though this is a surrogate for the actual distribution of microspheres and the validity of this proxy has been questioned multiple times [1113]. As 90Y decay lacks gamma radiation, nuclear imaging after SIRT was initially restricted to SPECT imaging of bremsstrahlung with poor spatial resolution. 90Y decay however also involves positron emission in 0.003% of decays [14], and PET-imaging was adopted as an additional imaging modality in 2010, yielding images with a higher spatial resolution that are more fitting for dosimetry, providing a definitely improved quantification accuracy compared to 90Y bremsstrahlung SPECT [15]. Next to the 90Y microspheres, 166Ho microspheres are the third commercially available product for SIRT [1618]. 166Ho emits a slightly less energetic spectrum of beta radiation (Eβ-max = 1.85 MeV (48.8%), 1.77 MeV (49.9%)) but additionally emits gamma radiation suitable for quantitative SPECT imaging (Eγ = 81 keV (6.6%)), and holmium makes the microspheres paramagnetic, allowing for MRI-based post-treatment dosimetry at a very high resolution [1821].

Over the years, multiple clinical studies have been performed in order to grasp the tumour dose-response relation after SIRT for a wide array of primary and metastatic liver malignancies such as hepatocellular carcinoma (HCC) [5], biliary tract cancer (BTC) [22], colorectal cancer (CRC) [6], neuroendocrine tumours (NET) [23], and (ocular) melanoma [24]. Even though the heterogeneity of study designs is rather large, many have described a so-called ‘threshold dose’ that needs to be achieved in order to achieve an objective response or at least local disease control. In 2018, a review article on the physics of SIRT has been published [25], which elaborates on many aspects of SIRT, such as pretreatment dosimetry methods, nuclear imaging strategies, and post-treatment dosimetric models. This systematic review adds to the aforementioned study by aiming to collect all studies that have investigated the tumour dose-response relationship after SIRT and to critically appraise the evidence at hand.

Methods

Search strategy

On the 31st of January 2020, an initial search was conducted in the following electronic databases: PubMed (MEDLINE), Embase, and the Cochrane Library. The full search strategy can be found in Supplementary Table 1. In brief, we searched for all synonyms for SIRT, but not the actual acronym SIRT, as that yielded a lot of extra results which were mainly about the sirtuin gene and not about radionuclide therapy. Even though the scope of the review clearly consists of dose-response evaluation, it was decided not to include this in the search strategy, as it is difficult to capture in a proper search term and this strongly limited the number of results. After the full-text screening was completed, a second search was performed on the 3rd of July 2020, to add articles that were published during the screening process. Reference lists of all included studies after full-text screening were used for manual cross-referencing. In between the completion and submission of this review, the results of the DOSISPHERE-01 trial were published [26]. As this is the only level 1 evidence available, it was decided to incorporate this study as well.

Inclusion criteria

Studies were considered eligible for full-text screening if they presented original research on SIRT of liver malignancies. A second criterion was that the title or abstract had to include at least one of the following terms: overall survival (OS), progression-free survival (PFS), or response criteria (e.g. RECIST, PERCIST, WHO, EASL). If this was not the case but the article was clearly about dosimetry based on the title and abstract, it was also included for full-text screening.

Exclusion criteria

Articles were excluded for full-text screening if they were not about SIRT, studied an isotope other than 90Y or 166Ho, or were about technical aspects of the treatment such as imaging technicalities or materials used. Reviews, case reports, comments, editorials, and study protocols were excluded, as well as all preclinical work. To further narrow down the results, all studies were excluded in which the effect of SIRT was investigated in combination with other therapies such as chemotherapy and immunotherapy.

Full-text screening

The full-text screening consisted of two selection rounds, of which the first was to exclude all articles that did not mention a liver dose or tumour dose at all. In the second round, the remaining full texts were screened more thoroughly on whether dose-response evaluation was performed. Only these articles were included in the final analysis.

All studies were assessed for eligibility independently by two reviewers (JR and NJMK). All full-text screening and data extraction were performed by JR. Disagreements were resolved by consensus or by another reviewer if deemed necessary (JFWN).

Analysis

Due to the extensive heterogeneity in study designs, study populations, outcome measures, and reporting, it was not deemed possible to perform a meta-analysis. Therefore, a narrative analysis was performed.

Quality of evidence

When writing a systematic review, an important aspect is evaluating the quality of evidence, preferably applying validated risk of bias tools such as the Grading of Recommendations Assessment, Development, and Evaluation (GRADE) system [27]. However, considering the high variability in study designs and reporting strategies and as the majority of the results are retrospective studies, we saw no possibilities to apply such validated risk of bias tools. An estimation of the overall quality of evidence was made taking the GRADE principles into account which is reported at the end of the results section.

Results

The screening process is depicted in Fig. 1. Briefly, our search strategy resulted in 3038 hits after the removal of duplicates, of which 487 publications were considered for full-text screening. Ultimately, 37 articles were included for final analysis. The resulting studies have been subdivided into three groups: studies on HCC patients, studies on non-HCC patients, and studies with mixed patient populations. These studies are summarized in Table 1 (HCC treated with glass microspheres), Table 2 (HCC treated with resin microspheres), Table 3 (non-HCC patients), and Table 4 (mixed populations).

Fig. 1.

Fig. 1

Flow chart of the search strategy and inclusion/exclusion process

Table 1.

Included articles on HCC patients treated with glass microspheres

Author Year Patients (n) Lesions (n) PVT (%) Criteria Response rate Activity prescription method Dosimetry modality Dosimetry technique Dose threshold Meaning Effect on median OS
Riaz [30] 2011 84 37%* WHO, EASL 59% (WHO), 81% (EASL) 120 Gy ILD Tumour hypervascularity ratio Suborgan (MIRD) No significant difference between responders and non-responders
Chiesa1 [31] 2011 46 70 EASL 38% 120 Gy ILD 99mTc-MAA SPECT Voxel-based 257 Gy, 400 Gy Predictor of response, 85% sensitivity, 70% specificity (257 Gy), specificity 86% (400 Gy)
Mazzaferro1 [35] 2013 52 65 67% EASL 40% 120 Gy ILD 99mTc-MAA SPECT Voxel-based 500 Gy Predictor of response, AUC = 0.78
Chiesa1 [36] 2015 52 65 67% EASL 40% 120 Gy ILD 99mTc-MAA SPECT Voxel-based 250 Gy, 1000 Gy Tumour control probability ≥50%. Lower dose needed for lesions <10 ml (250 Gy)
Garin [32] 2012 36 58 44% EASL 67% 120 Gy ILD 99mTc-MAA SPECT Suborgan (MIRD) 205 Gy Predictor of response, 100% sensitivity, 75% specificity 18 vs. 9 mo
Garin [34] 2013 71 45% EASL 78.8% One group 120 Gy ILD, other group 205 Gy TD, <120 Gy healthy liver dose 99mTc-MAA SPECT Suborgan (MIRD) 205 Gy Predictor of response, 100% sensitivity, 53% specificity 23.2 vs. 11.5 mo
Garin [5] 2015 41 100% EASL 85.4% 205 Gy TD, <120 Gy healthy liver dose 99mTc-MAA SPECT Suborgan (MIRD) 205 Gy Predictor of response, 100% sensitivity, 90% accuracy 18.2 vs. 4.3 mo
Garin [37] 2017 85 132 36% EASL 80.2% 80–150 Gy ILD 99mTc-MAA SPECT Suborgan (MIRD) 205 Gy Predictor of response, TCP = 89.7% (>205 Gy) vs. TCP = 9.1% (<205 Gy). Also prolonged OS. 21 vs. 6.5 mo
Garin [26] 2020 58 67% EASL, RECIST 1.1 36–71% One group 120 Gy ILD, other group 205 Gy TD, <120 Gy healthy liver dose 99mTc-MAA SPECT Suborgan (MIRD) 205 Gy Predictor of response, TCP = 71% (>205 Gy) response vs. TCP = 36% (<205 Gy). Also prolonged OS 26.6 vs. 10.7 mo
Kokabi [40] 2014 18 100% NA NA 120 Gy ILD 90Y-SPECT Suborgan (MIRD) 100 Gy Predictor of prolonged survival 13.2 vs. 4.3 mo
Srinivas [43] 2014 56 98 (48**) mRECIST 48% 120 Gy ILD 90Y-PET Suborgan (MIRD) No significant difference between responders and non-responders
Haste [12] 2017 62 RECIST, vRECIST, mRECIST 81% (mRECIST) As recommended by manufacturer (probably 120 Gy ILD) 99mTc-MAA SPECT and 90Y-PET Suborgan (MIRD) No significant difference between responders and non-responders
Chan [45] 2018 27 38 52% mRECIST 84% At least 120 Gy ILD 90Y-PET Voxel-based 200 Gy Predictor of response, sensitivity of 66%, specificity of 100%. AUC 0.875.
Ho [46] 2018 62 30% MTB decrease ≥50% 59.70% Calculated based on various parameters, compartment model 90Y-PET Suborgan (MIRD) 152–262 Gy Predictor of response. 11C-acetate avid tumours require a lower dose (152 Gy) than 18F-FDG avid tumours (262 Gy). Sensitivity 89.2%, specificity 88%
Kappadath [41] 2018 34 53 85%* WHO, RECIST, mRECIST 56.5% (mRECIST) 120 Gy ILD 90Y-SPECT Voxel-based 160 Gy TCP160 Gy = 50%

Articles are sorted chronologically and are grouped per research group when possible

Dose thresholds indicate mean dose, unless indicated otherwise

All effects on median overall survival are significant unless indicated otherwise, p values are not reported

HCC, hepatocellular carcinoma; ILD, injected liver dose; MIRD, medical internal radiation dose; MTB, metabolic tumour burden; n.s., not significant; OS, overall survival; PVT, portal vein thrombosis; TCP, tumour control probability; TD, tumour dose

*These represent the percentage of BCLC-C class patients when PVT percentages were not provided

**For this lesion-based dose-response analysis, only 48 lesions were included

1These articles use the same patient population, Chiesa (2011) contains preliminary results of Mazzaferro (2013)

Table 2.

Included articles on HCC patients treated with resin microspheres

Author Year Patients (n) Lesions (n) PVT (%) Criteria Response rate Activity prescription method Dosimetry modality Dosimetry technique Dose threshold Meaning Effect on median OS
Lau [28] 1994 16 ≥50% decrease in tumour volume 50% Not clearly described Beta probe Suborgan (MIRD) 120 Gy Predictor of response (87.5% vs. 12.5% response rate) 12.9 vs. 6.0 mo
Ho [29] 1997 71 ≥50% decrease in tumour volume 22.5% Not clearly described Beta probe Suborgan (MIRD) 225 Gy (for response)

TCP>225 Gy = 37.5%

TCP<225 Gy = 10.3%

11.0 vs. 6.8 mo (300 Gy cut-off, n.s.)
Strigari [39] 2010 73 2.7%* RECIST, EASL 54% (RECIST, 77% (EASL) BSA method 90Y-SPECT Voxel-based 110–120 Gy Predictor of response in at least 50% of patients
Kao [33] 2012 8 50% RECIST 1.1, EASL, WHO 37.50% Partition model 99mTc-MAA SPECT Suborgan (MIRD) 91 Gy Median decrease of index tumour of 58%
Kao [3] 2013 7 29% RECIST 63.6% Not clearly described 90Y-PET Voxel-based D70 ≥ 100 Gy Predictor for complete response
Allimant [44] 2018 38 80% mRECIST 31% BSA method 90Y-PET Voxel-based 61 Gy** Predictor of response, 76.5% sensitivity, 75% specificity
Tabone [42] 2020 24 100% mRECIST, EASL 54%*** Partition model, <40 Gy to healthy tissue 90Y-SPECT Suborgan (MIRD) Responders received a higher median dose (248 vs. 138 Gy) 30 vs. 11 mo
Hermann [38] 2020 109 61%* RECIST 1.1 20.5% Modified BSA method 99mTc-MAA SPECT Voxel-based 100–150 Gy TCP100 Gy = 72%, TCP150 Gy = 92% 14.1 vs. 6.8 mo

Dose thresholds indicate mean dose, unless indicated otherwise

BSA, body surface area; D70, minimum dose that 70% of the voxels received; HCC, hepatocellular carcinoma; MIRD, medical internal radiation dose; n.s., not significant; OS, overall survival; PVT, portal vein thrombosis; TCP, tumour control probability

*These represent percentages of BCLC-C class patients when PVT percentages were not provided

**This study reported the area under the DVH instead of mean dose, which is equal to the mean dose [63]

***This group also includes stable disease

Table 3.

Included articles on non-HCC patients

Author Year Cancer type Micro-spheres Patients (n) Lesions (n) Criteria Response rate Activity prescription method Dosimetry modality Dosimetry technique Dose threshold Meaning Effect on median OS
Lam [48] 2013 CRC Resin 25 RECIST 1.1 26.7% BSA method 99mTc-MAA and 99mTc- SC SPECT Suborgan (MIRD) 55 Gy Predictor of prolonged survival 32.8 mo vs. 7.2 mo
Van den Hoven [49] 2016 CRC Resin 30 133 PET RECIST, ≥50% TLG reduction 46% BSA method 90Y-PET Suborgan (MIRD) 40–60 Gy Predictor of metabolic response (conservative estimation) 11.5 mo vs. 5.3 mo (60 Gy cut-off)
Willowson [4] 2017 CRC Resin 22 63 TLG ≥50% reduction 67% Modified BSA method 90Y-PET Voxel-based 50 Gy Predictor of metabolic response, 91% PPV
Levillain [6] 2018 CRC Resin 24 57 Adapted from PERCIST, ≥50% TLG reduction 64.9% Partition model 90Y-PET Voxel-based 39 Gy, 60 Gy Predictor of metabolic response, 39 Gy for <15% TLG reduction, 80% sensitivity, 95% specificity. 60 Gy for ≥50% TLG reduction, 70% sensitivity, 95% specificity 13 vs. 5 mo (39 Gy cut-off)
Abbott [50] 2020 CRC Resin 23 96 vRECIST 14% Modified BSA method 90Y-SPECT Voxel-based 48.3 Gy Predictor of response, odds ratio 1.09
Van Roekel [51] 2020 CRC Ho-166 PLLA 40 PERCIST, RECIST 1.1 36% (PERCIST) 60 Gy ILD 166Ho-SPECT Suborgan (MIRD) 90 Gy Predictor of response, sensitivity 100%, specificity 38% Improved survival
Bourien [22] 2019 BTC Glass 64 RECIST1.1, Choi 15% (RECIST), 71% (Choi) 80–150 Gy ILD (or higher if segmentectomy) 99mTc-MAA SPECT Suborgan (MIRD) 260 Gy Predictor of response (Choi criteria). No correlation RECIST response 28.2 vs. 11.4 mo
Eaton [24] 2014 Melanoma Resin 7 30 EORTC 60% Self-modified BSA method 90Y-SPECT Voxel-based 50 Gy Predictor of metabolic response
Chansanti [23] 2017 NET Resin 15 55 mRECIST 46.7% Partition model 99mTc-MAA SPECT Suborgan (MIRD) 72.8 Gy, 191.3 Gy Predictor of response (191.3 Gy), 83% sensitivity and 93% specificity. <72.8 Gy predicted no response with 100% sensitivity

Dose thresholds indicate mean dose, unless indicated otherwise

BSA, body surface area; BTC, biliary tract cancer; CRC, colorectal cancer; HCC, hepatocellular carcinoma; ILD, injected liver dose; MIRD, medical internal radiation dose; NET, neuroendocrine tumour; OS, overall survival; SC, sulphur colloid; TLG, total lesion glycolysis

Table 4.

Included articles on mixed populations

Author Year Cancer type Microspheres Patients (n) Lesions (n) Criteria Activity prescription method Response rate Dosimetry modality Dosimetry technique Dose threshold Meaning Effect on median OS
Demirelli [55] 2015 12 HCC, 5 BTC, 18 CRC, 6 breast, 11 others 46/54 resin, 8/54 glass 54 RECIST, EORTC BSA method (resin), 100–110 Gy ILD (glass) 33% (RECIST), 48% (EORTC) 99mTc-MAA SPECT Suborgan (MIRD) 280 Gy Predictor for tumour control (95% TCP, both criteria), AUC = 0.869 17.7 vs. 9 mo
Song [52] 2015 16 HCC, 3 BTC, 4 others Resin 23 RECIST1.1 Partition model, BSA method Not clearly reported 99mTc-MAA SPECT, 90Y-PET Suborgan (MIRD) 200 Gy (90Y-PET) Predictor for prolonged PFS. Could not be found for 99mTc-MAA SPECT based dosimetry
Lam [54] 2015 26 HCC, 18 BTC, 20 NET, 29 CRC 98/122 resin, 24/122 glass 122 mRECIST Modified BSA method (resin), 90–120 Gy ILD (glass) 48% 99mTc-MAA and 99mTc- SC SPECT Suborgan (MIRD) Tumour dose independently associated with survival, stratified for tumour type
Fowler [53] 2016 8 HCC, 4 NET, 9 CRC, 3 others Mainly glass for HCC (6/8), mainly resin for other (14/16) 24 92 RECIST, vRECIST Modified BSA (resin), as manufacturer recommends (glass) 41.3% (RECIST), 73.5% (vRECIST) 90Y-PET Voxel-based 29.8 Gy (CRC) Predictor of response, 76.9% sensitivity, 75.9% specificity. No significant trend for other tumour types
Bastiaannet [56] 2019 21 CRC, 4 breast, 4 cholangiocarcinoma, 4 melanoma, 3 others Ho-166 PLLA 36 98 PERCIST 60 Gy ILD, 80 Gy ILD (3 patients) 50% 166Ho-SPECT Suborgan (MIRD) Responders had received a higher tumour dose than non-responders. Median OS was higher in responders

Dose thresholds indicate mean dose, unless indicated otherwise

BSA, body surface area; BTC, biliary tract cancer; CRC, colorectal cancer; HCC, hepatocellular carcinoma; ILD, injected liver dose; MIRD, medical internal radiation dose; NET, neuroendocrine tumour; OS, overall survival; SC, sulphur colloid

In general, a large variation in study designs and outcome reporting was found. The included studies applied the following spectrum of response criteria: RECIST (6/37), RECIST 1.1 (8/37), mRECIST (8/37), vRECIST (3/37), EASL (12/37), WHO (5/37), Choi (1/37), EORTC (2/37), and criteria based on metabolic response such as PERCIST or more generally a decrease in metabolic tumour burden (6/37). Additionally, the timing of response evaluation varies greatly, as some studies evaluate the response at set time points such as 1, 3, or 6 months, while others choose the best radiological outcome time point or do not describe the timing of response evaluation at all.

Almost all studies (30/37) have defined a ‘mean dose (Dmean) threshold’ based on their research that should be achieved in order to improve the chance of a beneficial treatment outcome. As described later on, there is no consensus on what the clinical value of this threshold should imply. Most of these studies (28/30) have correlated this threshold to an improved radiological or metabolic response rate, although there is heterogeneity among response assessment methods and reporting of the characteristics of the threshold as well.

Hepatocellular carcinoma (HCC)

In total, 23 studies have been performed on HCC patients, most of which have used glass microspheres (15/23 studies; 65%). Notably, there is a large variation in the methods used for dosimetry. Two very early studies have determined the T/N ratio through direct, intraoperative measurement of the 90Y decay with a beta probe [28, 29]. One study has attempted to make an estimation of the T/N ratio by incorporating the hypervascularization status (based on CT imaging) into a volumetric partition model [30]. A total of 11 studies have predicted the tumour doses based on 99mTc-MAA SPECT [5, 12, 26, 3138], four have imaged the microsphere distribution with 90Y-SPECT [3942], and six studies have utilized 90Y-PET [3, 12, 4346]. This has also resulted in a very wide range of reported dose thresholds, varying from 61 Gy to 1000 Gy, with the majority being between 100 and 250 Gy (13/19 studies; 68%). Thresholds for glass microspheres are definitely higher than for resin microspheres (range 100–1000 Gy vs. 61–300 Gy).

The first study by Lau et al. from 1994 [28] has set a dose threshold of 120 Gy in order to predict response, although it is not clearly described in the article how this threshold was chosen. In this article, a reference was made towards earlier studies utilizing iodine-131-lipiodol for the treatment of HCC, but the referenced study does not provide further details regarding the 120 Gy cut-off for efficacious treatment either. The only other study applying a similar dosimetric approach [29] has managed to deliver a dose ≥120 Gy to all tumours but has reached a low response rate of only 22.5%.

Only three studies have failed to demonstrate a dose-response relationship. The study by Riaz and colleagues [30] has incorporated a so-called hypervascularity ratio based on angiographic imaging into their dosimetric considerations, a technique that has not been used in later studies. In the discussion of Srinivas et al. [43], it was mentioned that the study is very likely to have been underpowered (n = 56 patients). In the third study [12], it was pointed out that their high response rate (81%) and therefore small group of non-responders (n = 7) is a possible explanation for the lack of correlation between absorbed dose and radiological response.

As mentioned before, another point of interest is the actual meaning of the defined dose thresholds. For instance, the studies by Garin et al. [5, 26, 32, 34, 37] have all applied the 205-Gy predicted dose threshold that was defined in the first study [32], which was chosen in order to achieve a sensitivity of 100%, or in other words, the author chose to set the response threshold at the minimal dose of responding lesions, corresponding to a tumour control probability (TCP) of 100%. The specificity was 75% in the first study and 53% in the second [34]. Other studies have, for instance, chosen their threshold to obtain a specificity of 100% [45], strived for 50% TCP [41], or have used Youden’s index in order to choose a threshold [44].

Three articles were based on (nearly) the same patient group: the 2011 study of Chiesa and colleagues [31] has presented preliminary data of the publication by Mazzaferro et al. [35], and this patient cohort was re-used for further analysis in 2015 [36]. Interestingly, these studies have proposed vastly different dose thresholds of 257/400 Gy (257 Gy for maximal Youden index, 400 Gy threshold for increased specificity) [31], 500 Gy [35] and 250/1000 Gy (250 Gy for TCP50 for lesions smaller than 10 ml, and 1000 Gy for TCP50 for lesions threshold for lesions larger than 10 ml) [36]. The last study bases the threshold on the dose that is required in order to achieve a tumour control probability of 50%, in which tumour control is defined as a combination of CR and PR based on EASL criteria.

The study by Ho et al. [46] is the only HCC study that has based response evaluation on nuclear imaging through FDG-PET. In this study, it was described that the dose threshold is dependent on the cellular differentiation state of the tumours, which was assessed through 11C-acetate and [18F]-FDG PET/CT imaging. 11C-acetate avidity was used as a surrogate for a well-differentiated state, whereas [18F]-FDG avidity was a surrogate for a poorly differentiated state. The resulting dose thresholds for achieving a metabolic tumour burden reduction of 50% are different, i.e. 152 Gy for 11C-acetate avid tumours (sensitivity 90.5%, specificity 87.5%) and 262 Gy for [18F]-FDG avid tumours (sensitivity 75.0%, specificity 91.7%).

Two studies are especially noteworthy, of which the first is the recently published ancillary study of the (negative) prospective SARAH trial. In this this study, the relationship of 99mTc-MAA-based, predicted tumour dose to survival and response (RECIST 1.1) was investigated. In the original publication on the SARAH trial [47], no difference in survival was found between SIRT with resin 90Y microspheres and sorafenib treatment. However, in this secondary analysis, a clear dose-response relationship has been found, as the highest disease control rate was found in patients of whom the predicted tumour dose exceeded 100 Gy and in whom there was an optimal agreement between 99mTc-MAA SPECT and 90Y-SPECT or 90Y-PET post-treatment [38]. The second study is the DOSISPHERE-01 trial, which is the first prospective study in nuclear medicine therapy designed to elucidate the benefit of personalized dosimetric treatment planning, in this case, a personalized dose of glass 90Y microspheres to patients with HCC. In the intervention group, pretreatment dosimetry was based on tumour uptake on the 99mTc-MAA SPECT, with the goal of achieving a dose >205 Gy to the index lesion. The control group received standard pretreatment dosimetry, i.e. 120 Gy targeted to the perfused liver volume. In this study, it was clearly demonstrated that personalized dosimetric treatment planning resulted in a significant increase in objective response (EASL criteria, 71% vs. 36%) and a survival benefit (26.6 mo vs. 10.7 mo), without an increase in liver toxicity [26].

Eight retrospective studies [5, 28, 32, 34, 37, 38, 40, 42] have also found a significant survival benefit in patients of whom the tumour dose reaches their reported threshold (range 100–205 Gy), most of which are a two to threefold increase in overall survival.

Non-HCC cancers

Six studies have been performed on hepatic metastases of CRC [4, 6, 4851], one on BTC [22], one on melanoma [24], and one on NETs [23]. Almost all of the non-HCC studies have been performed with resin microspheres, except for the CRC study using 166Ho-PLLA microspheres [51] and one study on BTC with glass microspheres [22].

The reported dose thresholds for CRC are more in line with each other than the thresholds for HCC: the total range over all five yttrium-based studies was 40 to 60 Gy. Three studies have used the same microspheres (resin) and the same response criterion of a total lesion glycolysis (TLG) reduction ≥50% on FDG-PET follow-up and have also used 90Y-PET for dosimetry. In the study by Van den Hoven and others [49], the presented Dmean threshold of 40–60 Gy is a conservative estimate. Herein, it was also demonstrated that the baseline TLG is associated with the extent of metabolic response. Willowson et al. found that a Dmean greater than 50 Gy predicted a metabolic response with a positive predictive value of 91% [4]. The third study [6] demonstrated that a Dmean ≥ 60 Gy was a predictor for metabolic response with a specificity of 95% and sensitivity of 70%. Additionally, they have described that a threshold of 39 Gy or lower can be used to predict a non-response with a sensitivity of 80% and specificity of 95%.

Similar to HCC, three studies [4, 48, 49] report that a difference in Dmean is correlated to a prolonged overall survival for CRC patients, most dramatically a fourfold difference in overall survival in the study by Lam et al. (Dmean threshold of 55 Gy) [48].

One recent study has explored the dose-response relationship in patients with CRC that have been treated with 166Ho-poly(L-lactic acid) (PLLA) microspheres. The threshold of 90 Gy (100% sensitivity/TCP100) was, as described by the authors, difficult to compare to the abovementioned thresholds found in yttrium-based studies because of numerous reasons as differences in specific activity and half-life [51].

The study by Eaton et al. [24] on 7 melanoma patients (resin microspheres) has found a correlation between the percentage of a tumour volume that received >50Gy and the extent of [18F]-FDG-PET response (decrease in SUVmax). The studies on BTC [22] (glass microspheres) and NET [23] (resin microspheres) result in Dmean thresholds of 260 Gy (sensitivity 73.7%, specificity 80%) and 191.3 Gy (sensitivity 83%, specificity 93%), respectively, which are more in the range of HCC studies. Similar to the publication by Levillain et al. [6], Chansanti and colleagues describe that a Dmean lower than 72.8 Gy was a predictor of non-response in patients with intrahepatic NETs, with a sensitivity of 100% [23].

Mixed populations

The study by Song et al. [52] mainly studied HCC patients (69.5%) and BTC patients (13%) treated with resin microspheres, which resulted in a Dmean threshold of 200 Gy (seemingly arbitrarily chosen, a predictor for prolonged PFS of 286 vs. 92 days, i.e. 9.4 vs. 3.0 mo). This is in line with the results on HCC and BTC described above, in which these tumour types appear to require a higher dose than, for instance, CRC. Fowler and colleagues [53] have found a significant dose-response relationship only for CRC (resin microspheres), with a reported dose threshold of 29.8 Gy (sensitivity 76.9%, specificity 75.9%). It has to be noted that all patient subgroups in this study were relatively small (<10 patients per group). In the article published by Lam et al., [54] a relatively large patient cohort (n = 122) treated with either glass or resin microspheres was studied, resulting in an independent association between predicted dose (based on 99mTc-MAA) and survival, after stratifying for tumour type (univariate and multivariate analysis). No mean dose thresholds were identified. Only one other study (with either glass or resin microspheres) performed on a mixed population looked into survival as well, in which a difference in overall survival was found between patients of whom the mean tumour doses exceeded 280 Gy and patients with lower tumour doses (TCP95, OS 17.7 mo vs. 9 mo) [55].

One publication using 166Ho microspheres has investigated the dose-response relationship after SIRT in a mixed population [56]. This study has linked the (geometric) mean tumour dose to both local response as well as survival (linear mixed-response model, log-rank test), similar to all work that has been performed with 90Y microspheres. No dose threshold was determined.

Tumour dose heterogeneity

One of the earliest studies reporting the heterogeneity of the dose distribution within a tumour was performed by Kao et al. in 2013 (resin microspheres) [3], in which it was decided to report the D70 (minimum dose delivered to 70% of the tumour) and V100 (percentage of tumour volume receiving ≥100 Gy) values based on the acquired DVHs. A D70 > 100 Gy was (arbitrarily) suggested as a threshold to predict treatment response in HCC. In the study on HCC patients by Kappadath et al. (glass microspheres), Dmean and D20 to D80 were found to be correlated to mRECIST response, but not D10, D90, and D100 [41]. Willowson and colleagues investigated dose-response in CRC patients [4] and found that the use of D70 as a dose metric may be favourable to Dmean as this resulted in a stronger correlation between the dose metric and the outcome, albeit very subtle. Moreover, it is suggested to incorporate a measure of tumour dose heterogeneity such as the coefficient of variance into the dose-response analysis, as this improved the positive predictive value of the prediction model. Last, a study on a relatively small, mixed cohort indicated that DVH-derived dose metrics such as D70 are more important for predicting response in hypovascularized lesions than in hypervascularized lesions [53].

Quality of evidence

It is difficult to correctly apply the GRADE approach [27] to this systematic review, as the scope of the review is not to study the extent of clinical benefit of SIRT directly. The objective of elucidating the dose-response correlation would, for instance, not benefit from a control group incorporated in the study design. However, we can state that almost all included studies have been retrospectively executed. For some studies, it is unclear whether patient cases have been re-used for later publications by the same authors, which is a direct disadvantage of the retrospective study designs. If the quality of this body of evidence were to be rated through the GRADE approach, the level would be low.

The most important GRADE category in this body of evidence is the inconsistency between studies, on many levels of study design: patient populations, response criteria, time to follow-up, and applied dosimetry techniques. However, despite this large heterogeneity in study approaches, the relation between mean tumour dose and response or prolonged overall survival has been demonstrated in over 20 studies. We therefore consider it rather likely that the effect of absorbed tumour dose on response and even survival is genuine, albeit unclear what the true mean tumour dose is that needs to be achieved in order to expect a response after SIRT.

Discussion

In this systematic review, all available evidence on the extent of the tumour dose-response relation in SIRT has been summarized. In the past 25 years, many research groups have investigated this correlation, leading to a broad scale of results. The main finding of this systematic review is that there clearly is a dose-response relation in SIRT (as demonstrated in 34/37 included studies), although it remains difficult to thoroughly characterize this relation, particularly in terms of biological effectiveness. The two most frequently studied cancer types are HCC and CRC and the recommended Dmean for HCC (100–250 Gy) appears to be higher than for CRC (40–60 Gy). Reported thresholds are lower for resin microspheres than when glass microspheres are used (for HCC: range 61–300 Gy vs. 100–1000 Gy).

A second major finding is the lack of a designing and reporting standard between the various studies. Some heterogeneity between different studies speaks for itself, such as the choice for specific response criteria for specific tumour types (e.g. mRECIST for HCC and RECIST 1.1 for CRC), but the heterogeneity in response criteria is much more extensive than that. Other aspects concerning the choice and implementation of response criteria should be noted as well. It has, for instance, been demonstrated that a metabolic response is achieved earlier after SIRT than anatomic response (1 mo vs. 3 mo) [57]. There is however no clear consensus between different studies on the optimal timing of the response evaluation: timing varies between 1, 3, and/or 6 months after treatment, and some studies reported the best response found over multiple evaluation time points. Moreover, some studies strive for disease control, others for an objective decline in tumour volume (i.e. partial response and complete response).

A similar problem arises when comparing the reported dose thresholds, as there is a wide variety of methods through which thresholds were chosen. There is a large variation as it is not clearly defined whether to adhere, for instance, 100% sensitivity or specificity or a completely different outcome prediction parameter. As a TCP curve is a sigmoid and not a step function, another variance is the fact that not all studies express the same TCP that their threshold results in. Such fundamental differences in study designs and reporting make it impossible to combine data from different studies.

Post-treatment dosimetry is a crucial step in establishing the dose-response relation after SIRT. Obviously, the applied dosimetry techniques have improved and therefore changed over the 25 years of research incorporated in this systematic review. The downside is that this is another aspect that impedes the comparison of the included studies. Of the 37 incorporated articles, 17 have not evaluated the dose post-treatment but have estimated the absorbed dose based on 99mTc-MAA SPECT imaging prior to treatment. Even though the validity of this approach has been a point of discussion [1113], the studies by Garin and colleagues [5, 26, 32, 34, 37, 58] have, for instance, demonstrated its clear value in predicting the outcome of HCC patients. Optimization through pretreatment dosimetry (in which activity prescription is based on imaging before treatment) is vital for improving the patient outcome, but only post-treatment dosimetry provides information on the technical success of the treatment and the actual dose distribution. Even though pretreatment dosimetry is helpful in predicting response and could lead to its own tumour-response prediction model, a thorough understanding of the true dose-response relation is still lacking, and that can only be evaluated through post-treatment dosimetry. We therefore argue the added value of basing the evaluation of dose-response in SIRT on 90Y-PET-based and 166Ho-MRI-based dosimetry. These imaging modalities do directly visualize the achieved dose distribution in a high-resolution and are therefore most fit for elucidating the dose-response relation.

In order to truly unravel the dose-response characteristics after SIRT, we require, for each histology and kind of microsphere, a universally adopted, standardized pretreatment and post-treatment dosimetry protocol, as well as a reliable radiological response assessment method, and a well-defined methodology to fix an efficacy threshold. This is also illustrated by the ancillary study of the SARAH trial [38], in which dosimetry was performed based on pretreatment 99mTc-MAA SPECT imaging, because post-treatment imaging was often lacking and there was no consensus on performing 90Y-based PET or SPECT imaging between a large number of cooperating centres. Large prospective clinical trials with dosimetry and response evaluation as a primary end point, such as the recently published DOSISPHERE-01 trial [26], are naturally more valuable in resolving this matter than retrospectively executed studies, even though a meta-analysis of (a selection of) the presented retrospective studies in which the data would be re-evaluated in a standardized manner could potentially also provide a large step in the right direction.

The systematic review has demonstrated that resin microspheres seem to result in a different dose threshold than glass microspheres. This may partially be explained by the heterogeneity of the distribution of microspheres as a direct result of the injected number of microspheres and therefore the heterogeneity of the resulting dose distribution, which can be analysed through simulations of DVH’s and DVH-derived dose metrics at a microscopic scale [59]. Several studies have incorporated measures such as D70 as opposed to the Dmean that results from MIRD-based suborgan dosimetry. It has been suggested that DVH-derived dose metrics are more important in predicting the outcome of hypovascularized lesions such as CRC metastases [53]. One can imagine that the highly vascularized status of lesions such as HCC is intrinsically resulting in a more homogeneous microsphere distribution in which the microspheres are positioned closer to each other, simply as a result of the higher density of arterioles in which the microspheres can lodge. In a simulation study by Pasciak et al., it has been demonstrated that a decrease in the number of microspheres will lead to a decrease in D70 and will decrease the steepness of the slope of the DVH [60]. This effect can be compensated for by increasing the total administered dose, i.e. the specific activity of the microspheres, which is the exact difference between the glass and resin 90Y microspheres. This may partially explain the differences in Dmean thresholds that are found, in which glass microsphere treatments appear to require a higher Dmean than resin microspheres.

Conclusion

Even though there is extensive evidence for a dose-response relationship in SIRT for both primary liver tumours and liver metastases, many pieces of the puzzle are still missing. This review indicates that the mean absorbed dose threshold to expect a response appears to be higher for HCC (100–250 Gy) than for CRC (40–60 Gy) and also depends on the type of microsphere used (for HCC: range 61–300 Gy vs. 100–1000 Gy). Other than two prospective, randomized trials (DOSISPHERE-01 and SARAH), of which only the first had dosimetry as a primary focus, the quality of evidence is low, precluding any definitive conclusions. Therefore, the field would greatly benefit from a reporting standard and prospective studies designed to further elucidate the dose-response relation in different tumour types. In the past 2 years, two expert panels have formulated recommendations for personalized dosimetry for glass microspheres for HCC treatment [61] and resin microspheres for HCC and other cancer types [62]. We argue that any prospective trial into the effectiveness of SIRT should incorporate standardized dosimetry, in order to at least evaluate the technical success of the treatments performed. In our opinion, this standardization should at least include voxel-based post-treatment dosimetry with high-resolution imaging, resulting in not only a Dmean but also a measure of dose heterogeneity within the tumours. Secondly, the field would benefit from a standardized response evaluation method and standardized methods through which dose thresholds are defined (such as a TCP of 50%). We expect an instance such as the EANM to formulate guidelines to facilitate and streamline this process in the coming years (Tables 2, 3, 4).

Supplementary information

ESM 1 (13.3KB, docx)

(DOCX 13 kb)

Author contribution

All authors contributed to the study conception and design. Material preparation, data collection, and analysis were performed by JR, NJMK, and LELWG. The first draft of the manuscript was written by JR, and all authors commented on previous versions of the manuscript. All authors read and approved the final manuscript.

Funding

Open access funding provided by Radboud University Medical Center.

Availability of data and material

The datasets generated during and/or analysed during the current study are available from the corresponding author on reasonable request.

Code availability

Not applicable.

Declarations

Ethics approval

Not applicable.

Consent to participate

Not applicable.

Consent for publication

Not applicable.

Conflict of interest

JFW Nijsen is a co-founder and part-time scientific director of Quirem Medical which has been acquired by Terumo Europe NV in July 2020. Nijsen is entitled to certain milestone payments from Terumo which are related to Quirem’s financial, operational, and regulatory performance in the future. Furthermore, Nijsen is an inventor on the patents related to radioactive microspheres that are assigned to the University Medical Center Utrecht Holding BV, Quirem Medical or BASF Corp. The activities of JFW Nijsen within Quirem Medical are approved of and supported by the board of directors of the Radboudumc. The other authors declare that they have no conflict of interest.

Footnotes

This article is part of the Topical Collection on Dosimetry

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Nienke J. M. Klaassen and Lovisa E. L. Westlund Gotby contributed equally to this work.

References

  • 1.Burton MA, Gray BN, Klemp PF, Kelleher DK, Hardy N. Selective internal radiation therapy: distribution of radiation in the liver. Eur J Cancer Clin Oncol. 1989;25:1487–1491. doi: 10.1016/0277-5379(89)90109-0. [DOI] [PubMed] [Google Scholar]
  • 2.Andrews JC, Walker SC, Ackermann RJ, Cotton LA, Ensminger WD, Shapiro B. Hepatic radioembolization with yttrium-90 containing glass microspheres: preliminary results and clinical follow-up. J Nucl Med. 1994;35:1637–1644. [PubMed] [Google Scholar]
  • 3.Kao YH, Steinberg JD, Tay YS, Lim GK, Yan J, Townsend DW, et al. Post-radioembolization yttrium-90 PET/CT - part 2: dose-response and tumor predictive dosimetry for resin microspheres. EJNMMI Res. 2013;3:57. doi: 10.1186/2191-219X-3-57. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Willowson KP, Hayes AR, Chan DLH, Tapner M, Bernard EJ, Maher R, et al. Clinical and imaging-based prognostic factors in radioembolisation of liver metastases from colorectal cancer: a retrospective exploratory analysis. EJNMMI Res. 2017;7:46. doi: 10.1186/s13550-017-0292-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Garin E, Rolland Y, Edeline J, Icard N, Lenoir L, Laffont S, et al. Personalized dosimetry with intensification using 90Y-loaded glass microsphere radioembolization induces prolonged overall survival in hepatocellular carcinoma patients with portal vein thrombosis. J Nucl Med. 2015;56:339–346. doi: 10.2967/jnumed.114.145177. [DOI] [PubMed] [Google Scholar]
  • 6.Levillain H, Duran Derijckere I, Marin G, Guiot T, Vouche M, Reynaert N, et al. (90)Y-PET/CT-based dosimetry after selective internal radiation therapy predicts outcome in patients with liver metastases from colorectal cancer. EJNMMI Res. 2018;8:60. doi: 10.1186/s13550-018-0419-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Cremonesi M, Chiesa C, Strigari L, Ferrari M, Botta F, Guerriero F, et al. Radioembolization of hepatic lesions from a radiobiology and dosimetric perspective. Front Oncol. 2014;4:210. doi: 10.3389/fonc.2014.00210. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Pasciak AS, Abiola G, Liddell RP, Crookston N, Besharati S, Donahue D, et al. The number of microspheres in Y90 radioembolization directly affects normal tissue radiation exposure. Eur J Nucl Med Mol Imaging. 2020;47:816–827. doi: 10.1007/s00259-019-04588-x. [DOI] [PubMed] [Google Scholar]
  • 9.Van Der Gucht A, Jreige M, Denys A, Blanc-Durand P, Boubaker A, Pomoni A, et al. Resin versus glass microspheres for (90)Y Transarterial Radioembolization: comparing survival in Unresectable hepatocellular carcinoma using pretreatment partition model Dosimetry. J Nucl Med. 2017;58:1334–1340. doi: 10.2967/jnumed.116.184713. [DOI] [PubMed] [Google Scholar]
  • 10.Srinivas SM, Nasr EC, Kunam VK, Bullen JA, Purysko AS. Administered activity and outcomes of glass versus resin (90)Y microsphere radioembolization in patients with colorectal liver metastases. J Gastrointest Oncol. 2016;7:530–539. doi: 10.21037/jgo.2016.03.09. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Wondergem M, Smits MLJ, Elschot M, De Jong HWAM, Verkooijen HM, Van Den Bosch MAAJ, et al. 99mTc-macroaggregated albumin poorly predicts the intrahepatic distribution of 90Y resin microspheres in hepatic radioembolization. J Nucl Med. 2013;54:1294–1301. doi: 10.2967/jnumed.112.117614. [DOI] [PubMed] [Google Scholar]
  • 12.Haste P, Tann M, Persohn S, LaRoche T, Aaron V, Mauxion T, et al. Correlation of technetium-99m macroaggregated albumin and yttrium-90 glass microsphere biodistribution in hepatocellular carcinoma: a retrospective review of pretreatment single photon emission CT and posttreatment positron emission tomography/CT. J Vasc Interv Radiol. 2017;28:722–30.e1. doi: 10.1016/j.jvir.2016.12.1221. [DOI] [PubMed] [Google Scholar]
  • 13.Elschot M, Nijsen JF, Lam MG, Smits ML, Prince JF, Viergever MA, et al. 99mTc-MAA overestimates the absorbed dose to the lungs in radioembolization: a quantitative evaluation in patients treated with 166Ho-microspheres. Eur J Nucl Med Mol Imaging. 2014;41:1965–1975. doi: 10.1007/s00259-014-2784-9. [DOI] [PubMed] [Google Scholar]
  • 14.Walrand S, Flux GD, Konijnenberg MW, Valkema R, Krenning EP, Lhommel R, et al. Dosimetry of yttrium-labelled radiopharmaceuticals for internal therapy: 86Y or 90Y imaging? Eur J Nucl Med Mol Imaging. 2011;38(Suppl 1):S57–S68. doi: 10.1007/s00259-011-1771-7. [DOI] [PubMed] [Google Scholar]
  • 15.Lhommel R, van Elmbt L, Goffette P, Van den Eynde M, Jamar F, Pauwels S, et al. Feasibility of 90Y TOF PET-based dosimetry in liver metastasis therapy using SIR-spheres. Eur J Nucl Med Mol Imaging. 2010;37:1654–1662. doi: 10.1007/s00259-010-1470-9. [DOI] [PubMed] [Google Scholar]
  • 16.Smits ML, Nijsen JF, van den Bosch MA, Lam MG, Vente MA, Mali WP, et al. Holmium-166 radioembolisation in patients with unresectable, chemorefractory liver metastases (HEPAR trial): a phase 1, dose-escalation study. Lancet Oncol. 2012;13:1025–1034. doi: 10.1016/S1470-2045(12)70334-0. [DOI] [PubMed] [Google Scholar]
  • 17.Prince JF, van den Bosch M, Nijsen JFW, Smits MLJ, van den Hoven AF, Nikolakopoulos S, et al. Efficacy of Radioembolization with (166)Ho-microspheres in salvage patients with liver metastases: a phase 2 study. J Nucl Med. 2018;59:582–588. doi: 10.2967/jnumed.117.197194. [DOI] [PubMed] [Google Scholar]
  • 18.Radosa CG, Radosa JC, Grosche-Schlee S, Zophel K, Plodeck V, Kuhn JP, et al. Holmium-166 Radioembolization in hepatocellular carcinoma: feasibility and safety of a new treatment option in clinical practice. Cardiovasc Intervent Radiol. 2019;42:405–412. doi: 10.1007/s00270-018-2133-7. [DOI] [PubMed] [Google Scholar]
  • 19.van de Maat GH, Seevinck PR, Bos C, Bakker CJ. Quantification of holmium-166 loaded microspheres: estimating high local concentrations using a conventional multiple gradient echo sequence with S(0)-fitting. J Magn Reson Imaging. 2012;35:1453–1461. doi: 10.1002/jmri.23593. [DOI] [PubMed] [Google Scholar]
  • 20.Klaassen NJM, Arntz MJ, Gil Arranja A, Roosen J, Nijsen JFW. The various therapeutic applications of the medical isotope holmium-166: a narrative review. EJNMMI Radiopharm Chem. 2019;4:19. doi: 10.1186/s41181-019-0066-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Smits ML, Elschot M, van den Bosch MA, van de Maat GH, van het Schip AD, Zonnenberg BA, et al. In vivo dosimetry based on SPECT and MR imaging of 166Ho-microspheres for treatment of liver malignancies. J Nucl Med. 2013;54:2093–2100. doi: 10.2967/jnumed.113.119768. [DOI] [PubMed] [Google Scholar]
  • 22.Bourien H, Palard X, Rolland Y, Le Du F, Beuzit L, Uguen T, et al. Yttrium-90 glass microspheres radioembolization (RE) for biliary tract cancer: a large single-center experience. Eur J Nucl Med Mol Imaging. 2019;46:669–676. doi: 10.1007/s00259-018-4199-5. [DOI] [PubMed] [Google Scholar]
  • 23.Chansanti O, Jahangiri Y, Matsui Y, Adachi A, Geeratikun Y, Kaufman JA, et al. Tumor dose response in Yttrium-90 resin microsphere embolization for neuroendocrine liver metastases: a tumor-specific analysis with dose estimation using SPECT-CT. J Vasc Interv Radiol. 2017;28:1528–1535. doi: 10.1016/j.jvir.2017.07.008. [DOI] [PubMed] [Google Scholar]
  • 24.Eaton BR, Kim HS, Schreibmann E, Schuster DM, Galt JR, Barron B, et al. Quantitative dosimetry for yttrium-90 radionuclide therapy: tumor dose predicts fluorodeoxyglucose positron emission tomography response in hepatic metastatic melanoma. J Vasc Interv Radiol. 2014;25:288–295. doi: 10.1016/j.jvir.2013.08.021. [DOI] [PubMed] [Google Scholar]
  • 25.Bastiaannet R, Kappadath SC, Kunnen B, Braat A, Lam M, de Jong H. The physics of radioembolization. EJNMMI Phys. 2018;5:22. doi: 10.1186/s40658-018-0221-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Garin E, Tselikas L, Guiu B, Chalaye J, Edeline J, de Baere T, et al. Personalised versus standard dosimetry approach of selective internal radiation therapy in patients with locally advanced hepatocellular carcinoma (DOSISPHERE-01): a randomised, multicentre, open-label phase 2 trial. Lancet Gastroenterol Hepatol. 2020. 10.1016/S2468-1253(20)30290-9. [DOI] [PubMed]
  • 27.Balshem H, Helfand M, Schunemann HJ, Oxman AD, Kunz R, Brozek J, et al. GRADE guidelines: 3. Rating the quality of evidence. J Clin Epidemiol. 2011;64:401–406. doi: 10.1016/j.jclinepi.2010.07.015. [DOI] [PubMed] [Google Scholar]
  • 28.Lau WY, Leung WT, Ho S, Leung NW, Chan M, Lin J, et al. Treatment of inoperable hepatocellular carcinoma with intrahepatic arterial yttrium-90 microspheres: a phase I and II study. Br J Cancer. 1994;70:994–999. doi: 10.1038/bjc.1994.436. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Ho S, Lau WY, Leung TW, Chan M, Johnson PJ, Li AK. Clinical evaluation of the partition model for estimating radiation doses from yttrium-90 microspheres in the treatment of hepatic cancer. Eur J Nucl Med. 1997;24:293–298. doi: 10.1007/BF01728766. [DOI] [PubMed] [Google Scholar]
  • 30.Riaz A, Gates VL, Atassi B, Lewandowski RJ, Mulcahy MF, Ryu RK, et al. Radiation segmentectomy: a novel approach to increase safety and efficacy of radioembolization. Int J Radiat Oncol Biol Phys. 2011;79:163–171. doi: 10.1016/j.ijrobp.2009.10.062. [DOI] [PubMed] [Google Scholar]
  • 31.Chiesa C, Maccauro M, Romito R, Spreafico C, Pellizzari S, Negri A, et al. Need, feasibility and convenience of dosimetric treatment planning in liver selective internal radiation therapy with (90)Y microspheres: the experience of the National Tumor Institute of Milan. Q J Nucl Med Mol Imaging. 2011;55:168–197. [PubMed] [Google Scholar]
  • 32.Garin E, Lenoir L, Rolland Y, Edeline J, Mesbah H, Laffont S, et al. Dosimetry based on 99mTc-macroaggregated albumin SPECT/CT accurately predicts tumor response and survival in hepatocellular carcinoma patients treated with 90Y-loaded glass microspheres: preliminary results. J Nucl Med. 2012;53:255–263. doi: 10.2967/jnumed.111.094235. [DOI] [PubMed] [Google Scholar]
  • 33.Kao YH, Tan AEH, Burgmans MC, Irani FG, Khoo LS, Lo RHG, et al. Image-guided personalized predictive dosimetry by artery-specific SPECT/CT partition modeling for safe and effective 90Y radioembolization. J Nucl Med. 2012;53:559–566. doi: 10.2967/jnumed.111.097469. [DOI] [PubMed] [Google Scholar]
  • 34.Garin E, Lenoir L, Edeline J, Laffont S, Mesbah H, Poree P, et al. Boosted selective internal radiation therapy with 90Y-loaded glass microspheres (B-SIRT) for hepatocellular carcinoma patients: a new personalized promising concept. Eur J Nucl Med Mol Imaging. 2013;40:1057–1068. doi: 10.1007/s00259-013-2395-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Mazzaferro V, Sposito C, Bhoori S, Romito R, Chiesa C, Morosi C, et al. Yttrium-90 radioembolization for intermediate-advanced hepatocellular carcinoma: a phase 2 study. Hepatology. 2013;57:1826–1837. doi: 10.1002/hep.26014. [DOI] [PubMed] [Google Scholar]
  • 36.Chiesa C, Mira M, Maccauro M, Spreafico C, Romito R, Morosi C, et al. Radioembolization of hepatocarcinoma with (90)Y glass microspheres: development of an individualized treatment planning strategy based on dosimetry and radiobiology. Eur J Nucl Med Mol Imaging. 2015;42:1718–1738. doi: 10.1007/s00259-015-3068-8. [DOI] [PubMed] [Google Scholar]
  • 37.Garin E, Rolland Y, Pracht M, Le Sourd S, Laffont S, Mesbah H, et al. High impact of macroaggregated albumin-based tumour dose on response and overall survival in hepatocellular carcinoma patients treated with (90) Y-loaded glass microsphere radioembolization. Liver Int. 2017;37:101–110. doi: 10.1111/liv.13220. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Hermann AL, Dieudonne A, Ronot M, Sanchez M, Pereira H, Chatellier G, et al. Relationship of tumor radiation-absorbed dose to survival and response in hepatocellular carcinoma treated with transarterial radioembolization with (90)Y in the SARAH study. Radiology. 2020;191606. 10.1148/radiol.2020191606. [DOI] [PubMed]
  • 39.Strigari L, Sciuto R, Rea S, Carpanese L, Pizzi G, Soriani A, et al. Efficacy and toxicity related to treatment of hepatocellular carcinoma with 90Y-SIR spheres: radiobiologic considerations. J Nucl Med. 2010;51:1377–1385. doi: 10.2967/jnumed.110.075861. [DOI] [PubMed] [Google Scholar]
  • 40.Kokabi N, Galt JR, Xing M, Camacho JC, Barron BJ, Schuster DM, et al. A simple method for estimating dose delivered to hepatocellular carcinoma after yttrium-90 glass-based radioembolization therapy: preliminary results of a proof of concept study. J Vasc Interv Radiol. 2014;25:277–287. doi: 10.1016/j.jvir.2013.11.007. [DOI] [PubMed] [Google Scholar]
  • 41.Kappadath SC, Mikell J, Balagopal A, Baladandayuthapani V, Kaseb A, Mahvash A. Hepatocellular carcinoma tumor dose response after (90)Y-radioembolization with glass microspheres using (90)Y-SPECT/CT-based voxel dosimetry. Int J Radiat Oncol Biol Phys. 2018;102:451–461. doi: 10.1016/j.ijrobp.2018.05.062. [DOI] [PubMed] [Google Scholar]
  • 42.Tabone M, Calvo A, Russolillo N, Langella S, Carbonatto P, Lo Tesoriere R, et al. Downstaging unresectable hepatocellular carcinoma by radioembolization using 90-yttrium resin microspheres: a single center experience. J Gastrointest Oncol. 2020;11:84–90. doi: 10.21037/jgo.2019.06.01. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.Srinivas SM, Natarajan N, Kuroiwa J, Gallagher S, Nasr E, Shah SN, et al. Determination of radiation absorbed dose to primary liver tumors and normal liver tissue using post radioembolization 90Y PET. Front Oncol. 2014;4. [DOI] [PMC free article] [PubMed]
  • 44.Allimant C, Kafrouni M, Delicque J, Ilonca D, Cassinotto C, Assenat E, et al. Tumor targeting and three-dimensional voxel-based dosimetry to predict tumor response, toxicity, and survival after yttrium-90 resin microsphere radioembolization in hepatocellular carcinoma. J Vasc Interv Radiol. 2018;29:1662–70.e4. doi: 10.1016/j.jvir.2018.07.006. [DOI] [PubMed] [Google Scholar]
  • 45.Chan KT, Alessio AM, Johnson GE, Vaidya S, Kwan SW, Monsky W, et al. Prospective trial using internal pair-production positron emission tomography to establish the yttrium-90 radioembolization dose required for response of hepatocellular carcinoma. Int J Radiat Oncol Biol Phys. 2018;101:358–365. doi: 10.1016/j.ijrobp.2018.01.116. [DOI] [PubMed] [Google Scholar]
  • 46.Ho CL, Chen S, Cheung SK, Leung YL, Cheng KC, Wong KN, et al. Radioembolization with (90)Y glass microspheres for hepatocellular carcinoma: significance of pretreatment (11)C-acetate and (18)F-FDG PET/CT and posttreatment (90)Y PET/CT in individualized dose prescription. Eur J Nucl Med Mol Imaging. 2018;45:2110–2121. doi: 10.1007/s00259-018-4064-6. [DOI] [PubMed] [Google Scholar]
  • 47.Vilgrain V, Pereira H, Assenat E, Guiu B, Ilonca AD, Pageaux GP, et al. Efficacy and safety of selective internal radiotherapy with yttrium-90 resin microspheres compared with sorafenib in locally advanced and inoperable hepatocellular carcinoma (SARAH): an open-label randomised controlled phase 3 trial. Lancet Oncol. 2017;18:1624–1636. doi: 10.1016/S1470-2045(17)30683-6. [DOI] [PubMed] [Google Scholar]
  • 48.Lam MG, Goris ML, Iagaru AH, Mittra ES, Louie JD, Sze DY. Prognostic utility of 90Y radioembolization dosimetry based on fusion 99mTc-macroaggregated albumin-99mTc-sulfur colloid SPECT. J Nucl Med. 2013;54:2055–2061. doi: 10.2967/jnumed.113.123257. [DOI] [PubMed] [Google Scholar]
  • 49.van den Hoven AF, Rosenbaum CE, Elias SG, de Jong HW, Koopman M, Verkooijen HM, et al. Insights into the dose-response relationship of radioembolization with resin 90Y-microspheres: a prospective cohort study in patients with colorectal cancer liver metastases. J Nucl Med. 2016;57:1014–1019. doi: 10.2967/jnumed.115.166942. [DOI] [PubMed] [Google Scholar]
  • 50.Abbott EM, Falzone N, Lee BQ, Kartsonaki C, Winter H, Greenhalgh TA, et al. The impact of radiobiologically-informed dose prescription on the clinical benefit of yttrium-90 SIRT in colorectal Cancer patients. J Nucl Med. 2020. 10.2967/jnumed.119.233650. [DOI] [PubMed]
  • 51.van Roekel C, Bastiaannet R, Smits MLJ, Bruijnen RC, Braat A, de Jong H, et al. Dose-effect relationships of holmium-166 radioembolization in colorectal cancer. J Nucl Med. 2020. 10.2967/jnumed.120.243832. [DOI] [PubMed]
  • 52.Song YS, Paeng JC, Kim HC, Chung JW, Cheon GJ, Chung JK, et al. PET/CT-based dosimetry in 90Y-microsphere selective internal radiation therapy: single cohort comparison with pretreatment planning on (99m)Tc-MAA imaging and correlation with treatment efficacy. Medicine (Baltimore) 2015;94:e945. doi: 10.1097/MD.0000000000000945. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 53.Fowler KJ, Maughan NM, Laforest R, Saad NE, Sharma A, Olsen J, et al. PET/MRI of hepatic 90Y microsphere deposition determines individual tumor response. Cardiovasc Intervent Radiol. 2016;39:855–864. doi: 10.1007/s00270-015-1285-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54.Lam MG, Banerjee A, Goris ML, Iagaru AH, Mittra ES, Louie JD, et al. Fusion dual-tracer SPECT-based hepatic dosimetry predicts outcome after radioembolization for a wide range of tumour cell types. Eur J Nucl Med Mol Imaging. 2015;42:1192–1201. doi: 10.1007/s00259-015-3048-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 55.Demirelli S, Erkilic M, Oner AO, Budak ES, Gunduz S, Ozgur O, et al. Evaluation of factors affecting tumor response and survival in patients with primary and metastatic liver cancer treated with microspheres. Nucl Med Commun. 2015;36:340–349. doi: 10.1097/MNM.0000000000000257. [DOI] [PubMed] [Google Scholar]
  • 56.Bastiaannet R, van Roekel C, Smits MLJ, Elias SG, van Amsterdam WAC, Doan DT, et al. First evidence for a dose-response relationship in patients treated with 166Ho-radioembolization: a prospective study. J Nucl Med Off Publ Soc Nucl Med. 2019;10. [DOI] [PubMed]
  • 57.Jongen JMJ, Rosenbaum C, Braat M, van den Bosch M, Sze DY, Kranenburg O, et al. Anatomic versus metabolic tumor response assessment after radioembolization treatment. J Vasc Interv Radiol. 2018;29:244–53.e2. doi: 10.1016/j.jvir.2017.09.024. [DOI] [PubMed] [Google Scholar]
  • 58.Garin E, Rolland Y, Boucher E, Ardisson V, Laffont S, Boudjema K, et al. First experience of hepatic radioembolization using microspheres labelled with yttrium-90 (TheraSphere): practical aspects concerning its implementation. Eur J Nucl Med Mol Imaging. 2010;37:453–461. doi: 10.1007/s00259-009-1279-6. [DOI] [PubMed] [Google Scholar]
  • 59.Walrand S, Hesse M, Chiesa C, Lhommel R, Jamar F. The low hepatic toxicity per Gray of 90Y glass microspheres is linked to their transport in the arterial tree favoring a nonuniform trapping as observed in posttherapy PET imaging. J Nucl Med. 2014;55:135–140. doi: 10.2967/jnumed.113.126839. [DOI] [PubMed] [Google Scholar]
  • 60.Pasciak AS, Bourgeois AC, Bradley YC. A microdosimetric analysis of absorbed dose to tumor as a function of number of microspheres per unit volume in 90Y radioembolization. J Nucl Med. 2016;57:1020–1026. doi: 10.2967/jnumed.115.163444. [DOI] [PubMed] [Google Scholar]
  • 61.Salem R, Padia SA, Lam M, Bell J, Chiesa C, Fowers K, et al. Clinical and dosimetric considerations for Y90: recommendations from an international multidisciplinary working group. Eur J Nucl Med Mol Imaging. 2019;46:1695–1704. doi: 10.1007/s00259-019-04340-5. [DOI] [PubMed] [Google Scholar]
  • 62.Levillain H, Bagni O, Deroose CM, Dieudonne A, Gnesin S, Grosser OS, et al. International recommendations for personalised selective internal radiation therapy of primary and metastatic liver diseases with yttrium-90 resin microspheres. Eur J Nucl Med Mol Imaging. 2021. 10.1007/s00259-020-05163-5. [DOI] [PMC free article] [PubMed]
  • 63.Walrand S, Chiesa C, Gabina PM, Chouin N, Gear J, Stokke C, et al. Re: tumor targeting and three-dimensional voxel-based Dosimetry to predict tumor response, toxicity, and survival after Yttrium-90 resin microsphere Radioembolization in hepatocellular carcinoma. J Vasc Interv Radiol. 2019;30:2047–2048. doi: 10.1016/j.jvir.2019.08.030. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

ESM 1 (13.3KB, docx)

(DOCX 13 kb)

Data Availability Statement

The datasets generated during and/or analysed during the current study are available from the corresponding author on reasonable request.

Not applicable.


Articles from European Journal of Nuclear Medicine and Molecular Imaging are provided here courtesy of Springer

RESOURCES