TABLE II:
CSL Errors for Scatterers (τl = 0) and Microbubbles with Varying Lifetime (τl ≠ 0) from Field II Simulations
Method | Units | Source Localization Error, Δr |
||||
---|---|---|---|---|---|---|
τl = 0μs | τl = 1μs | τl = 3μs | τl = 10μs | τl = 0 to 10μs | ||
Anderson-Trahey | mm | 1.77 ± 1.12 | 2.00 ± 1.22 | 2.81 ± 1.35 | 6.00 ± 2.20 | 4.24 ± 1.73 |
Anderson-Trahey, x-transformation | mm | 0.41 ± 0.18 | 0.29 ± 0.16 | 0.90 ± 0.29 | 3.30 ± 0.96 | 1.93 ± 0.58 |
Hyperbolic Fit | mm | 0.13 ± 0.14 | 0.12 ± 0.12 | 0.12 ± 0.11 | 0.11 ± 0.10 | 0.12 ± 0.11 |
| ||||||
Anderson-Trahey | λ0 | 2.11 ± 1.34 | 2.39 ± 1.46 | 3.43 ± 1.62 | 7.17 ± 2.63 | 5.07 ± 2.07 |
Anderson-Trahey, x-transformation | λ0 | 0.49 ± 0.22 | 0.35 ± 0.19 | 1.08 ± 0.34 | 3.94 ± 1.14 | 2.31 ± 0.69 |
Hyperbolic Fit | λ0 | 0.16 ± 0.17 | 0.14 ± 0.14 | 0.14 ± 0.14 | 0.13 ± 0.12 | 0.14 ± 0.13 |