[Table 3:].
Selected Ultrasound Parameters Used for Therapy.
| US Parameter | In vivo-optimized | Phantom-optimized |
|---|---|---|
| PRF, Hz | 1 | 100 |
| Voltage, V | 70 | 90 |
| PNP, MPa | 3.4 | 4.4 |
| MI | 1.22 | 1.54 |
| Therapy time, sec | 100 | 100 |
| No of Therapy Pulses | 100 | 10,000 |