Skip to main content
. 2019 Apr 4;2(1):35–42. doi: 10.1159/000497361

Table 2.

LC-MS/MS conditions for quantification of CBD

CBD IS
Mass spectrometry parameters for CBD and IS
Polarity Negative Positive
Q1 mass, Da 313.085 467.161
Q3 mass, Da 245.033 372.900
Declistering potential, V −30 41
Entrance potential, V −10 10
Collision energy, V −28 13
Collision cell exit potential, V −17 10
Retention time, min 11.8 5.71

>Optimized source parameters
Curtain gas, psi 20 40
Collision gas, psi 3 5
IonSpray voltage, V −4,500 5,500
Source temperature, ° C 400 400
Ion source gas 1 40 80
Ion source gas 2 70 70

CBD, cannabidiol; IS, internal standard (clobetasol propionate).