Visualization of the renal artery in kidney transplant patients using time-resolved computed tomography angiography

Svensson-Marcial Anders 1,2,, Genberg Helena 2,3, Brehmer Katharina 1,2, Themudo Raquel 1,4, Brismar B Torkel 1,2
PMCID: PMC8489769  PMID: 34616566

Abstract

Background

Transplant renal artery stenosis (TRAS) is a post-operative complication which most often occurs between 3 months and 2 years after transplantation. TRAS is associated with kidney failure and hypertension and, thereby, with an increased risk of cardiovascular events.

Purpose

The purpose of this retrospective study was to report our experience of perfusion computed tomography angiography (P-CTA) to identify a 50% lumen reduction (as compared to digital subtraction angiography, DSA), assess its subjective image quality and evaluate if contrast-induced acute kidney injury (CI-AKI) occurred.

Material and Methods

All 13 patients who had undergone P-CTA for suspected TRAS at our institution were retrospectively evaluated. At P-CTA, eight or 12 g of iodine were administered intravenously, and five to seven scan sequences were merged into time-resolved images after motion correction. Eight patients underwent subsequent DSA.

Results

The average patient weight was 76 kg (range 55–97 kg). Image quality was rated as good or excellent for all patients, and pathological changes were shown in 10 of 13 patients undergoing P-CTA. Two patients had a serum creatinine increase of >26 μmol/L during the first 3 days, but serum creatinine was significantly lower in all patients 1 month after P-CTA (165+/−69 μmol/L versus 232+/−66 μmol/L, P < .01). The diagnosis at P-CTA was verified in all eight patients who underwent DSA. However, in two cases with suspected stenosis, renal function was restored without angioplasty.

Conclusion

Anatomy and blood flow of the transplant renal artery can be visualized using less than a third of the standard contrast media dose by using P-CTA technique.

Keywords: Computed tomography, time-resolved perfusion computed tomography angiography, transplant kidney, transplant renal artery stenosis, contrast agents-intravenous

Introduction

Transplant renal artery stenosis (TRAS) is a post-operative complication which most often occurs between 3 months and 2 years after transplantation. TRAS is associated with kidney failure and hypertension and, thereby, with an increased risk of cardiovascular events. The reported risk of TRAS varies among different studies, ranging from 1% to 23% in all kidney transplantations.13 Early detection of TRAS is important to prompt adequate treatment and to preserve renal function. In clinical practice, TRAS is diagnosed primarily by Doppler ultrasound (DUS). The reference test for the definitive diagnosis of arterial stenosis is digital subtraction invasive angiography (DSA).4,5 Alternative non-invasive techniques for diagnosing TRAS are magnetic resonance angiography (MRA) and intravenous contrast media (CM)–enhanced computed tomography angiography (CTA).6,7 However, the nephrotoxicity of intravenous iodine contrast media (CM) is still a challenge to CM-enhanced imaging. Gadolinium-enhanced MRA is also contraindicated in patients with severe renal failure due to the risk of nephrogenic systemic fibrosis (NSF).8,9

The rapid evolution of CT technology has enabled the development of perfusion computed tomography angiography (P-CTA), by which multiple passages/sequences over the organ of interest are acquired. The technique can also be used for minimizing the CM dose at CT angiography (CTA).1013 The high temporal resolution guarantees that a short CM bolus will be detected with optimal arterial enhancement. Using post-processing of image data, the rendered images can be shown as a movie, and the inflow of the CM in the vessels can be visualized as high-resolution time-resolved images.

The purpose of this retrospective study was to report our experience of P-CTA to identify a 50% lumen reduction (as compared to DSA), assess its subjective image quality, and evaluate if contrast-induced acute kidney injury (CI-AKI) occurred.

Material and methods

This retrospective study was approved by the Regional Ethics Board, and informed consent was waived. All methods used in this study were carried out in accordance with local clinical radiology guidelines. The study complies with the Declaration of Helsinki. The inclusion criteria were a) patients in whom DUS had difficulty visualizing the renal artery and b) referred for P-CTA examination of the transplant kidney renal artery between April 2012 and Feb 2019. In total, 13 patients with clinical signs of TRAS were retrospectively included in the study. There were seven men and six women, aged 26–67 (mean 50) years and the time between kidney transplantation and P-CTA ranged between 12 and 270 days. All 13 patients had acute onset of hypertension and reduced kidney function (Table 1).

Table 1.

Patient characteristics, estimated glomerular filtration rate per minute (eGFR, mL/min 1.73 m2) at the day of P-CTA, 3 days after and 30 days after CT. The amount of intravenous contrast media administered is stated in grams. The diagnosis (dx) at P-CTA and the number of days after the P-CTA that the DSA was performed and what interventions performed is also written out.

Pat. No Age Body weight (kg) Body height (cm) BMI (kg/m2) Day of the P-CTA Iodine given (g) 3 days after P-CTA 30 days after P-CTA P-CTA dx Day of DSA Treatment at DSA
eGFR sCr eGFR sCr eGFR sCr
1 46 56 153 24 34 155 12 56 104 53 108 Stenosis
2 62 97 178 31 22 257 12 22 260 34 242 Stenosis 22 Ballon
3 46 78 185 23 21 294 12 33 203 37 185 Stenosis 53 Ballon
4 36 82 182 25 28 245 12 31 226 45 168 None 13 None
5 26 60 156 25 20 263 10 14 348 80 84 Stenosis 2 Stent
6 52 65 167 23 21 230 10 16 285 29 175 Not significant 14 None
7 54 91 191 25 32 204 10 29 217 51 137 None
8 55 61 165 22 36 141 10 28 173 45 118 Stenosis
9 59 93 173 31 28 218 12 53 127 56 121 Stenosis 1 Ballon
10 63 92 173 31 14 365 12 15 350 15 350 Kinked artery 62 Stent
11 36 71 154 30 14 307 8 25 215 52 118 None
12 53 90 183 27 42 161 12 38 174 43 169 Stenosis 6 Ballon
13 67 55 160 21 26 173 10 29 160 27 170 Atero-sklerosis

Notes: Patient: No.1 received anti-rejection therapy, No. 4 had angiotensin II-receptor blocker discontinued, No. 5 had DSA 2 days after P-CTA, No. 7 underwent P-CTA within 12 days from transplantation and had a recovery of transplantation-induced acute tubular necrosis in the early post-operative period, No. 9 had DSA the day after P-CTA, and No. 11 had a biopsy bleeding and subcapsular hematoma. This patient underwent surgical decompression.

All patients were scanned with a dual source 2 x 64 detector row MDCT (Siemens Somatom Definition Flash, Forchheim, Germany). The position of the transplant kidney was defined by non-CM scanning over the pelvis using automatic dose modulation, 100 kVp and 45 reference mAs. During P-CTA, a total volume of 20–30 mL (8–12 g (g) iodine) of iomeprol 400 mg I/mL, (Iomeron-400, Bracco Imaging SpA, Milan, Italy) was administered at 6 mL per second, followed by a saline chaser with the same injection flow through a 16-gauge peripheral venous access inserted in the antecubital vein. A fixed dose of 12 g of iodine was administered to the first four patients. From patient five and onward, the dosage was adjusted through visual assessment according to the patient’s body constitution. The scan delay was set to 15 s after the start of the CM injection. The P-CTA consisted of 5–7 scan repetitions over a range of 15 cm covering the transplant kidney, using 100 kVp and 70 mAs, giving a radiation dose below <9 millisievert (mSv).14 The CM enhancement of the iliac and renal arteries was monitored visually while scanning the different series, and the scan was manually ended when the enhancement started to decrease. Time resolution between series was 1.5 s.

All reconstructed series were sent to a dedicated post-processing workstation for motion correction and creation of a time-resolved merged data set (Syngovia, Siemens Healthcare, Forchheim, Germany). All multiplanar reformats (MPR), 3D volume rendering (VR), and 4D post-processing were carried out at a separate workstation (Advantage workstation 4.6, GE Healthcare, Milwaukee, WI, USA). The total post-processing time, including image transfer time between the CT and SyngoVia workstation, image motion correction, 3D reconstruction, 4D movie and image reformates, was approximately 45 min. With regard to retrospective analysis, two radiologists with 8 and 14 years of experience rated image quality on a five-grade scale (absent, poor, adequate, good, and excellent). The five-degree assessment of image quality summarized the visualization of the renal arteries, vascular anastomosis, and iliac arteries. The transplant renal artery was evaluated for the absence or the presence of luminal diameter reduction. From the thin axial slices, multiplanar reconstructions (MPR) were performed in order to visualize the artery in its long axis and in cross-sectional section perpendicular to the vessel long axis. When luminal reduction was present, the stenosis was graded as above or below 50 per cent luminal diameter narrowing.15,16 Luminal stenosis was assessed by visual assessment. Both reviewers were blinded to the DSA results. Reader disagreement was solved in consensus. Eight patients underwent DSA after the P-CTA. In six patients, DSA was performed within 1 month from P-CTA (1, 2, 6, 13, 14, and 22 days after P-CTA, respectively) and in two patients 53 and 62 days after P-CTA. The consensus from the P-CTA was compared with the findings reported by the interventionist at DSA.

Kidney function was evaluated by measurement of serum creatinine (sCr) on the day of the P-CTA examination, 3 days after and approximately >30 days after the P-CTA examination by calculating the estimated glomerular filtration rate (eGFR) using the Lund-Malmö formula. Student’s paired t-test, without Bonferroni correction, was used to test for changes in sCr, that is, to evaluate whether contrast-induced acute kidney injury (CI-AKI) had occurred. CI-AKI is defined as an increase in sCr of ≥26 μmol/l in the 48–72 h following CM administration or a sCr increase of ≥1.5–1.9 times baseline.17

Results

Average body weight was 76 kg (range 55–97 kg). The renal artery of the transplanted kidney was successfully imaged with image quality graded as good or excellent in all 13 patients by both reviewers (Table 2). All arterial anastomoses were done end-to-side to the external iliac artery. For three patients, a donor aortic cuff was used, and for the other 10 patients, the anastomosis was done without cuff. Intravenous CM inflow could be demonstrated with time-resolved images in all patients (Figure 1(a)–(d)), and pathological changes were demonstrated in 10 patients; nine with renal artery stenosis (Figures 25), of which one was due to renal atherosclerosis, and one with a kinked renal artery. The diagnosis was verified by DSA in eight patients, of whom six received balloon dilatation and one received a stent. In one patient, a normal artery was verified. In two patients, with suspected stenosis at P-CTA, the kidney function was stabilized and no DSA was performed. In the remaining three patients, no further imaging or intervention was done.

Figure 3.

Figure 3.

Significant stenosis of the renal artery (arrow). After CT examination, the patient was referred for acute percutaneous transluminal angioplasty (PTA), confirming the renal CTA finding.

Table 2.

Attenuation in iliac and transplant kidney artery (HU).

Image quality assessment
Pat. No Iliac artery Transplant kidney artery Reader 1 Reader 2
1 384 337 5 5
2 160 141 4 4
3 154 140 5 5
4 234 215 5 5
5 318 257 5 5
6 263 200 5 5
7 212 201 4 4
8 359 235 5 5
9 260 293 5 4
10 228 222 5 4
11 264 290 5 4
12 230 200 5 5
13 348 584 5 5

Figure 1. (a-d).

Figure 1. (a-d)

3D volume rendered (VR) images of the arterial anatomy adjacent to the transplant kidney. The images are reconstructed from 4 different P-CTA series, 1.5 time resolution per series.

Figure 2.

Figure 2.

Stenosis of the renal artery at the anastomosis (arrow).

Figure 5.

Figure 5.

(a–f). Volume rendered CT image (a) showing stenosis of the renal artery. Color-coded CT perfusion (b) using the same data set acquired from the P-CTA showing decreased perfusion of the caudal renal cortex (ROI 2), perfusion defect confirmed by scintigraphy (c, d). Patient was referred for digital subtraction angiography (DSA) and angioplasty (e, f).

As a by-product of the P-CTA, it was also possible to demonstrate whether the vessel stenosis was accompanied by a decreased parenchymal perfusion [Figure 5(b)].

A transient sCr of >25% was observed in two patients 3 days after the examination. In both patients, the P-CTA showed severe stenosis of the renal artery, which was successfully treated with angioplasty (balloon dilatation and stenting). When analyzing the whole group of patients, there was no significant change in average sCr 3 days after the examination, 219 ± 76 μmol/l, compared with before, 232 ± 66 μmol/l. At 30-day follow-up, sCr had significantly decreased (i.e., normalized) to 165 ± 69 μmol/l (P < .01) (Figure 6) (Table 1). The attenuation of the iliac artery ranged between 154 and 384 Hounsfield units (HU, average 263 HU) and of the renal artery between 141 and 584 HU (average 255 HU, Table 2).

Figure 6.

Figure 6.

Creatinine levels (μmol/liter) in the 13 patients before, after 3 and 30 days after time-resolved renal CTA.

Discussion

The renal artery was successfully visualized by P-CTA in all 13 kidney transplant patients. The image quality was rated as good or excellent in all patients. Six out of eight diagnosed significant stenosis, and two arteries without significant stenosis, were verified at subsequent DSA. In two patients, a stenosis was suspected, but kidney function was regained before DSA. These might constitute false positive cases.

When a small amount of CM is injected, as in our study, the CM bolus duration will be very short, which means an increased risk of scanning at the wrong time-point. Through repeated scanning and the use of the same software as that used for reconstruction and analysis of CT-perfusion studies, it was possible to obtain approximately 2000 low-resolution images at five to seven time-points, which eliminates the risk of scanning at the wrong time-point [Figures 7(a) and (b)].

Figure 7.

Figure 7.

(a, b). Low resolution CT images reconstructed from several time-points (a). By merging data from several time-points, a time-resolved merged data set with high image quality can be reconstructed (b).

Volume rendering technique (VRT) enabled the imaging of the vascularity of the kidney to be demonstrated with a so-called 4D-movie [Figure 1(a)–(d)]. Such a movie makes it easier for the surgeon and nephrologist to comprehend the anatomy and pathological changes. The data set can also be rotated freely to visualize the anatomy and pathology in the most optimal projection (Figures 24). The time-resolved data can also be merged into a single data set of grey-scale images or 3D VRT images with high image quality [Figure 7(b)]. This procedure compensates for the poor image quality of the time-resolved images caused by the low tube current used (70 mAs compared to approximately 150 mAs for standard images). In this study, the resulting image quality was rated good or excellent by the two radiologists. Moreover, scan data can be loaded into the perfusion software to visualize renal perfusion [Figure 5(b)].

Figure 4.

Figure 4.

Significant stenosis (arrow a) and calcifications (arrow b, c) of the renal artery.

Several other reference tests are available to evaluate the renal arteries. The least non-invasive and simplest for the patient is DUS.18 In most patients with renal transplants, the kidney is easily imaged as it is located rather superficially and not affected by patient habitus and bowel gas. However, DUS is also user-dependent, and the lack of images that can easily be interpreted by the clinician makes alternative imaging and diagnostic techniques valuable. Gadolinium-enhanced MRA is an appealing alternative,8,9 but as mentioned previously, the risk of NSF in patients with severe renal failure and scanner availability limit its use. Non-contrast MRA has shown promising results,19,20 but the technique is not yet widely available and there is a risk of occasional false positive results with overestimation of the severity of vascular stenosis.21 Standard CTA is not an alternative due to the high CM dose needed, almost 40 g of iodine.22,23 A technique to combine CTA with invasive placement of a micro-catheter in the supplying artery has successfully reduced the dose to 5–20 mL,24 but this is an invasive technique and requires the interaction of an interventionist. A one-stop-shop alternative is to perform DSA without prior imaging. This would make it possible to diagnose and treat any stenosis at the same time. However, availability, high cost, radiation, and contrast media dose, combined with the invasiveness of the procedure, are limiting factors. P-CTA has the great advantage of using only 8–12 g of iodine, which is less than both DSA, typically 18–33 g of iodine,25 and standard CTA.

The risk of causing contrast-induced AKI (CI-AKI) is increased in patients with deteriorating renal function. Some authors argue that the risk of CI-AKI is greatly exaggerated,26,27 whereas others argue that there is a considerable risk of CI-AKI in certain patient groups.17,28 However, in clinical practice, it is considered that the risk of kidney damage increases with the dose of injected CM, which is especially important to consider in kidney transplant patients.29,30 When developing the technique of P-CTA for TRAS, the dosage of CM was based on our previous experience from perfusion CT of the liver using 20 gI (50 mL). As there was no intention to evaluate renal parenchyma enhancement, it was considered that 12 g (30 mL) would suffice. After scanning the first five patients, it became clear that the dose could be further reduced depending on patient habitus. Still, despite the low CM dose, two patients had a serum creatinine increase of >26 μmol/l during the first 3 days. Both patients were diagnosed with severe stenosis of the renal artery [Figure 5(a)–(f)], providing a plausible alternative cause of the kidney injury. Also, one of the patients had no diuresis on the day prior to CTA. Moreover, in all 13 patients, the sCr was significantly lower >30 days after the examination than before, indicating that the CM given at P-CTA did not induce any long-standing kidney damage. However, the retrospective design of the study and the small number of cases make any conclusions on renal function after low contrast dose P-CTA merely speculative.

This study has several limitations. A direct comparison with DSA or MRA to verify the findings in all patients would have been desirable. However, performing DSA in patients with normal findings or stabilized renal function was considered unethical. Furthermore, the reason for developing the low contrast media technique was due to the low availability of MRA. There is also a risk of false positive stenosis at MRA. A possible reason for misinterpretation might be the strong curvature at the anastomosis to the iliac artery that is often observed. Unfortunately, this is also where stenosis most often develops. However, the lack of comparative DSA or MRA means that definite conclusions about the ability of P-CTA to determine if stenosis is present or absent cannot be drawn regarding the remaining five patients.

As this study describes a new technique in a vulnerable group of patients, the patients referred had a high risk of TRAS. This might have biased the reviewers’ evaluations.

It might be possible to use even lower doses of iodine than those used in this study. However, as only two patients out of 13 had possible CI-AKI, it was considered that the risk of non-diagnostic imaging by reducing the dose further would exceed the eventual benefit. The radiation dose during P-CTA was below <9 mSv, equivalent to a standard parenchymal phase liver CT scan. However, the local organ/tissue dose over the transplant kidney and its surroundings was higher due to the repeated scanning over the 15 cm scan length. Regarding availability, the technique of P-CTA should be increasingly available with the greater spread of modern multi-detector CT (MDCT) scanners, provided that it is accepted among radiologists and clinicians. In conclusion, perfusion technique CTA enables imaging of the transplanted kidney by using only a low dose of iodine, reducing the risk of CI-AKI.

Footnotes

Declaration of conflicting interests: The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.

Funding: The author(s) disclosed receipt of the following financial support for the research, authorship, and/or publication of this article: The study was financed by grants from the Swedish state under the agreement between the Swedish government and the county councils, the ALF-agreement (20190655).

ORCID iD

Svensson-Marcial Anders https://orcid.org/0000-0001-5685-7255

References

  • 1.Bruno S, Remuzzi G, Ruggenenti P. Transplant renal artery stenosis. J Am Soc Nephrol 2004; 15: 134–141. [DOI] [PubMed] [Google Scholar]
  • 2.Buturović-Ponikvar J. Renal transplant artery stenosis. Nephrol Dial Transpl 2003; 18(Suppl 5): 74–77. [DOI] [PubMed] [Google Scholar]
  • 3.Seratnahaei A, Shah A, Bodiwala K, et al. Management of transplant renal artery stenosis. Angiology 2011; 62: 219–224. [DOI] [PubMed] [Google Scholar]
  • 4.Biederman DM, Fischman AM, Titano JJ, et al. Tailoring the endovascular management of transplant renal artery stenosis. Am J Transplant 2015; 15: 1039–1049. [DOI] [PubMed] [Google Scholar]
  • 5.Rossi AP, Vella JP. Hypertension, living kidney donors, and transplantation: where are we today? Adv Chronic Kidney Dis 2015; 22: 154–164. [DOI] [PubMed] [Google Scholar]
  • 6.Tuna IS, Tatli S. Contrast-enhanced CT and MR imaging of renal vessels. Abdom Img 2014; 39: 875–891. [DOI] [PubMed] [Google Scholar]
  • 7.Gaddikeri S, Mitsumori L, Vaidya S, et al. Comparing the diagnostic accuracy of contrast-enhanced computed tomographic angiography and gadolinium-enhanced magnetic resonance angiography for the assessment of hemodynamically significant transplant renal artery stenosis. Curr Probl Diagn Radiol 2014; 43: 162–168. [DOI] [PubMed] [Google Scholar]
  • 8.Luk SH, Chan JHM, Kwan TH, et al. Breath-hold 3D gadolinium-enhanced subtraction MRA in the detection of transplant renal artery stenosis. Clin Radiol 1999; 54: 651–654. [DOI] [PubMed] [Google Scholar]
  • 9.Chan YL, Leung CB, Yu SCH, et al. Comparison of non-breath-hold high resolution gadolinium-enhanced MRA with digital subtraction angiography in the evaluation on allograft renal artery stenosis. Clin Radiol 2001; 56: 127–132. [DOI] [PubMed] [Google Scholar]
  • 10.Dowlatshahi D, Wasserman JK, Momoli F, et al. Evolution of computed tomography angiography spot sign is consistent with a site of active hemorrhage in acute intracerebral hemorrhage. Stroke 2014; 45: 277–280. [DOI] [PubMed] [Google Scholar]
  • 11.Lehmkul L, Andres C, Lucke C, et al. Dynamic CT angiography after abdominal aortic endovascular aneurysm repair: influence of enhancement patterns and optimal bolus timing on endoleak detection. Radiology 2013; 268: 890–899. [DOI] [PubMed] [Google Scholar]
  • 12.Fisher MA, Leidner B, Kartalis N, et al. Time-resolved computed tomography of the liver: Retrospective, multi-phase image reconstruction derived from volumetric perfusion imaging. Euro Radiol 2014; 24: 151–161. [DOI] [PubMed] [Google Scholar]
  • 13.Helck A, Bamberg F, Sommer WH, et al. Optimized contrast volume for dynamic CT angiography in renal transplant patients using a multiphase CT protocol. Eur J Radiol 2011; 80: 692–698. [DOI] [PubMed] [Google Scholar]
  • 14.The Measurement, Reporting, and Management of Radiation dose in CT. AAPM report no. 96, Report of AAP task group 23: CT dosimetry, 2008. [Google Scholar]
  • 15.Eriksson P, Mohammed AA, De Geer J, et al. Non-invasive investigations of potential renal artery stenosis in renal insufficiency. Nephrol Dial Transplant 2010; 25: 3607–3614. [DOI] [PubMed] [Google Scholar]
  • 16.Andersson M, Jägervall K, Eriksson P, et al. How to measure renal artery stenosis—a retrospective comparison of morphological measurement approaches in relation to hemodynamic significance. BMC Med Img 2015; 15: 42. DOI: 10.1186/s12880-015-0086-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.van der Molen AJ, Reimer P, Dekkers IA, et al. Post contrast acute kidney injury part 1: definition, clinical features, incidence, role of contrastmedium and risk factors: recomendations for updated ESUR contrast medium safety committee guidelines. Eur Radiol 2018; 28(7): 2845–2855. DOI: 10.1007/s00330-017-5246-5255. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Zeisbrich M, Kihm LP, Drüschler F, et al. When is contrast-enhanced sonography preferable over conventional ultrasound combined with Doppler imaging in renal transplantation? Clin Kidney J 2015; 8: 606–614. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Angeretti M, Lumia D, Canì A, et al. Non-enhanced MR angiography of renal arteries: comparison with contrast-enhanced MR angiography. Acta Radiologica 2013; 54: 749–756. [DOI] [PubMed] [Google Scholar]
  • 20.Liang K-W, Chen J-W, Huang H-H, et al. The performance of noncontrast magnetic resonance angiography in detecting renal artery stenosis as compared with contrast enhanced magnetic resonance angiography using conventional angiography as a reference. J Comput Assist Tomography 2017; 41: 619–627. [DOI] [PubMed] [Google Scholar]
  • 21.Bley TA, François CJ, Schiebler ML, et al. Non-contrast-enhanced MRA of renal artery stenosis: validation against DSA in a porcine model. Eur Radiol 2016; 26: 547–555. [DOI] [PubMed] [Google Scholar]
  • 22.Eun-Suk C, Jeong-Sik Y, Jhii-Hyun A, et al. CT angiography of the renal arteries: comparison of lower-tube-voltage CTA with moderate-concentration iodinated contrast material and conventional CTA. AJR Am J Roentg 2012; 199: 96–102. [DOI] [PubMed] [Google Scholar]
  • 23.Morita S, Matsuzaki Y, Yamamoto T, et al. Mid-term outcome of transarterial embolization of renal artery pseudoaneurysm and arteriovenous fistula after partial nephrectomy screened by early postoperative contrast-enhanced CT. CVIR Endovascular 2020; 3: 68. DOI: 10.1186/s42155-020-00160-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Althoff CE, Günther RW, Hamm B, et al. Intra-arterial ultra low iodine CT angiography of renal transplant arteries. CardioVascular Interv Radiol 2014; 37: 1062–1067. [DOI] [PubMed] [Google Scholar]
  • 25.Lanzman RS, Voiculescu A, Walther C, et al. ECG-gated nonenhanced 3D steady-state free precession MR angiography in assessment of transplant renal arteries: comparison with DSA. Radiology 2009; 252: 914–921. [DOI] [PubMed] [Google Scholar]
  • 26.Katzberg RW, Newhouse JH. Intravenous contrast medium-induced nephrotoxicity: is the medical risk really as great as we have come to believe? Radiology 2010; 256: 21–28. [DOI] [PubMed] [Google Scholar]
  • 27.Ehrmann S, Badin J, Savath L, et al. Acute kidney injury in the critically ill. Crit Care Med 2013; 41: 1017–1026. [DOI] [PubMed] [Google Scholar]
  • 28.Cheungpasitporn W, Thongprayoon C, Mao MA, et al. Contrast-induced acute kidney injury in kidney transplant recipients: A systematic review and meta-analysis. World J Transplant 2017; 7: 81–87. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Lee J, Cho JY, Lee HJ, et al. Contrast-induced nephropathy in patients undergoing intravenous contrast-enhanced computed tomography in Korea: a multi-institutional study in 101487 patients. Korean J Radiol 2014; 15: 456–463. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Nyman U, Almén T, Jacobsson B, et al. Are intravenous injections of contrast media really less nephrotoxic than intra-arterial injections? Eur Radiol 2012; 22: 1366–1371. [DOI] [PubMed] [Google Scholar]

Articles from Acta Radiologica Open are provided here courtesy of SAGE Publications