. 2021 Sep 8;46(3):211–220. doi: 10.4103/jmp.JMP_12_21

Table 1.

Novalis Tx and TrueBeam STx dosimetric match analysis

Energy (MV) Data Field size (cm×cm) Novalis Tx 5445 Novalis Tx 5479 Novalis Tx 6140 TrueBeam STx 3169
6 TPR 3×3 0.539 0.535 0.537 0.535
20/10 10×10 0.578 0.575 0.571 0.574
30×30 0.630 0.634 0.640 0.630
SRS-6 TPR 6×6 0.547 0.542 0.547
20/10 10×10 0.571 0.568 0.569
15×15 0.590 0.594 0.594
10 TPR 6×6 0.615 0.610 0.613
20/10 10×10 0.632 0.630 0.633
30×30 0.674 0.678 0.671

TPR: Tissue-phantom ratio