Table 1.
Novalis Tx and TrueBeam STx dosimetric match analysis
Energy (MV) | Data | Field size (cm×cm) | Novalis Tx 5445 | Novalis Tx 5479 | Novalis Tx 6140 | TrueBeam STx 3169 |
---|---|---|---|---|---|---|
6 | TPR | 3×3 | 0.539 | 0.535 | 0.537 | 0.535 |
20/10 | 10×10 | 0.578 | 0.575 | 0.571 | 0.574 | |
30×30 | 0.630 | 0.634 | 0.640 | 0.630 | ||
SRS-6 | TPR | 6×6 | 0.547 | 0.542 | 0.547 | |
20/10 | 10×10 | 0.571 | 0.568 | 0.569 | ||
15×15 | 0.590 | 0.594 | 0.594 | |||
10 | TPR | 6×6 | 0.615 | 0.610 | 0.613 | |
20/10 | 10×10 | 0.632 | 0.630 | 0.633 | ||
30×30 | 0.674 | 0.678 | 0.671 |
TPR: Tissue-phantom ratio