Table 3.

Novalis Tx and TrueBeam STx dosimetric leaf gap and transmission for open field and multileaf collimator

Energy (MV) Data Novalis Tx 5445 Novalis Tx 5479 Novalis Tx 6140 TrueBeam STx 3169
6 DLG (mm) 0.854 0.901 0.872 0.866
Transmission for open field (%) 1.25 1.27 1.20 1.17
Transmission for MLC (%) 1.27 1.18 1.20 1.17
SRS-6 DLG (mm) 0.815 0.901 0.864
Transmission for open field (%) 1.12 1.17 1.14
Transmission for MLC (%) 1.16 1.18 1.14
10 DLG (mm) 0.932 0.957 0.908
Transmission for open field (%) 1.37 1.37 1.40
Transmission for MLC (%) 1.46 1.37 1.40

Novalis Tx 5479 measurements were taken as reference. DLG: Dosimetric leaf gap, MLC: Multileaf collimator