Skip to main content
. 2021 Sep 8;46(3):211–220. doi: 10.4103/jmp.JMP_12_21

Table 4.

Novalis Tx SN-5479 and TrueBeam STx SN-3169 output factor local differences for 6-MV mode

M Jaw field size

mm 8 12 22 32 42 60 80 100 140 220 254
L 5 2.3% 2.0% 2.1% 2.2% 2.3% 2.2% 2.3% 2.5% 2.4% 2.7% 2.7%
C 10 −2.4% −1.1% −1.1% −1.1% −1.2% −1.1% −0.9% −0.9% −0.8% −0.5% −0.5%
F 20 −2.6% −1.6% −1.2% −1.1% −0.9% −0.9% −0.9% −0.7% −0.5% −0.3%
I 30 −1.2% −0.7% −0.6% −0.5% −0.4% −0.3% −0.2% −0.1% −0.1%
E 40 −0.7% −0.3% −2.0% −1.4% −0.6% −0.4% −0.4% 0.1%
L 60 −0.7% −0.4% −0.5% −0.2% 0.0% 0.3% 0.3%
D 80 −0.1% 0.0% 0.1% 0.4% 0.4%
S 100 0.0% 0.1% 0.4% 0.3%
I 140 0.4% 0.4% 0.3%
Z 220 0.7% 0.7%
E 254 1.2%