Delta4 gamma index comparison between Novalis Tx and TrueBeam STx “beam-matched”

Case Criteria (%/mm) Gamma index Novalis Tx 6140 (%) Gamma index TrueBeam STx 3169 (%) P
Prostate 3/2 98.9 (0.8) 99.5 (0.7) 0.086
3/1 94.4 (2.0) 96.1 (2.5) 0.104
2/2 97.0 (1.4) 98.1 (1.0) 0.060
H&N 3/2 98.5 (1.4) 98.4 (2.1) 0.889
3/1 93.9 (3.7) 96.2 (4.1) 0.270
2/2 95.0 (3.0) 95.4 (4.7) 0.877