Table 6.

Electronic portal image device gamma index comparison total arcs in Novalis Tx and TrueBeam STx “beam-matched”

Criteria (%/mm) Novalis Tx 5445 (%) Novalis Tx 5479 (%) Novalis Tx 6140 (%) TrueBeam STx 3169 (%)
Prostate
 3/2 97.1 (3.0) 97.8 (1.3) 99.4 (0.6)
P 0.143 0.027
 3/1 92.9 (7.1) 96.3 (1.6) 99.2 (0.7)
P 0.123 0.012
 2/2 95.3 (3.6) 95.9 (1.2) 98.4 (1.7)
P 0.164 0.019
Breast
 3/2 98.4 (0.6) 98.8 (0.4) 98.6 (0.6) 99.4 (0.4)
P 0.119 0.571 0.001
 3/1 94.3 (1.9) 95.6 (1.0) 94.9 (1.6) 97.7 (0.9)
P 0.119 0.150 0.008
 2/2 96.5 (0.9) 97.0 (0.6) 96.3 (1.3) 98.0 (0.8)
P 0.065 0.269 0.0001
H&N
 3/2 96.0 (2.5) 97.1 (1.8) 97.0 (1.5) 98.0 (1.9)
P 0.279 0.911 0.306
 3/1 89.6 (5.4) 93.4 (3.3) 93.2 (2.6) 95.2 (3.6)
P 0.169 0.421 0.508
 2/2 92.5 (3.9) 94.6 (2.3) 93.8 (2.0) 95.5 (3.5)
P 0.073 0.888 0.283

H and N: Head-and-neck