Retinal morphological changes and number of functioning electrodes after Argus II implantation and their extent

Retinal morphological changes Time since the onset of changes, months Extent of changes (number of electrodes involved/covered by scanned OCT images)
Functioning electrodes, n
1 month 3 months 6 months 12 months baseline 12 months
Patient 1
 Fibrosis 12 NA NA NA 41/58 59 59
 Schisis NA NA NA NA NA
Patient 2
 Fibrosis NA NA NA NA NA 54 54
 Schisis NA NA NA NA NA
Patient 3
 Fibrosis 3 NA 24/56 33/59 34/60 59 59
 Schisis 6 NA NA 30/59 30/60
Patient 4
 Fibrosis 3 NA 42/60 56/60 56/60 59 59
 Schisis NA NA NA NA NA
Patient 5
 Fibrosis NA NA NA NA NA 59 59
 Schisis NA NA NA NA NA