Skip to main content
. 2021;7(4):295–307.

Table 3.

Comparison of patient-specific QA results between SRS MapCHECK and ArcCHECK for 4 different groups. The difference is statistically significant if a p value is <0.05.

  Group A: SRS DCAT
Gamma criteria 3%/3 mm 3%/2 mm 2%/2 mm 2%/1 mm
QA device SRS
MC*
AC° SRS MC AC SRS MC AC SRS MC AC
Gamma passing rate (mean ± SD) (%) 100 99.96 ± 0.29 100 99.88 ± 0.52 99.95 ±
0.47
99.79 ±
0.75
99.54 ±
1.71
96.44 ±
3.56
Statistically significantly different? (Y/N) unavailable unavailable Y
(p value = 0.0269)
Y
(p value <0.0001)
Total No. of detectors irradiated (mean ± SD) SRS MC: 178.0 ± 57.8; AC: 61.0 ± 29.7
  Group B: SRS VMAT
Gamma criteria 3%/3 mm 3%/2 mm 2%/2 mm 2%/1 mm
QA device SRS
MC
AC SRS MC AC SRS MC AC SRS MC AC
Gamma passing rate (mean ± SD) (%) 99.79 ± 0.48 99.42 ± 1.40 99.16 ± 1.09 98.60 ± 2.34 98.42 ± 1.76 98.00 ± 2.60 93.71 ± 4.95 90.64 ± 5.68
Statistically significantly different? (Y/N) N
(p value = 0.0599)
N
(p value = 0.1762)
N
(p value = 0.3264)
Y
(p value = 0.0013)
Total No. of detectors irradiated (mean ± SD) SRS MC: 246.7 ± 110.7; AC: 62.0 ± 26.5
  Group C: SBRT DCAT
Gamma criteria 3%/3 mm 3%/2 mm 2%/2 mm 2%/1 mm
QA device SRS
MC
AC SRS MC AC SRS MC AC SRS MC AC
Gamma passing rate (mean ± SD) (%) 100 100 100 99.80 ±
0.36
100 97.60 ±
2.41
100 86.39 ±
8.69
Statistically significantly different? (Y/N) unavailable unavailable unavailable unavailable
Total No. of detectors irradiated (mean ± SD) SRS MC: 506.4 ± 135.6; AC: 263.4 ± 99.4
  Group D: SBRT VMAT
Gamma criteria 3%/3 mm 3%/2 mm 2%/2 mm 2%/1 mm
QA device SRS
MC
AC SRS MC AC SRS MC AC SRS MC AC
Gamma passing rate (mean ± SD) (%) 99.91 ± 0.35 99.82 ± 0.34 99.83 ± 0.64 99.71 ± 0.51 99.57 ± 1.13 99.00 ± 1.17 98.38 ± 3.27 93.60 ± 4.57
Statistically significantly different? (Y/N) N
(p value = 0.0637)
N
(p value = 0.0917)
Y
(p value = 0.0030)
Y
(p value <0.0001)
Total No. of detectors irradiated (mean ± SD) SRS MC: 536.3 ± 119.6; AC: 180.8 ± 86.2

*SRS MapCHECK

°ArcCHECK

Dynamic conformal arc therapy

Volumetric modulated arc therapy