Table 4.
Comparison of ArcCHECK patient-specific QA results between dose grid sizes of 1 mm × 1 mm and 2 mm × 2 mm for SBRT plans. The difference is statistically significant if a p value is < 0.05.
Group C: SBRT DCAT† | ||||||||
---|---|---|---|---|---|---|---|---|
Gamma criteria | 3%/3 mm | 3%/2 mm | 2%/2 mm | 2%/1 mm | ||||
Dose grid size | 2 mm | 1 mm | 2 mm | 1 mm | 2 mm | 1 mm | 2 mm | 1 mm |
Gamma passing rate (mean ± SD) (%) | 100 | 100 | 99.71 ± 0.40 | 99.80 ± 0.36 |
97.22 ± 2.62 | 97.60 ± 2.41 |
85.51 ± 9.32 | 86.39 ± 8.69 |
Statistically significantly different? (Y/N) | unavailable | N (p value = 0.4375) |
Y (p value = 0.0313) |
Y (p value = 0.0098) |
||||
Group D: SBRT VMAT‡ | ||||||||
Gamma criteria | 3%/3 mm | 3%/2 mm | 2%/2 mm | 2%/1 mm | ||||
Dose grid size | 2 mm | 1 mm | 2 mm | 1 mm | 2 mm | 1 mm | 2 mm | 1 mm |
Gamma passing rate (mean ± SD) (%) | 98.83 ± 3.67 | 99.82 ± 0.34 | 97.73 ± 5.88 | 99.71 ± 0.51 | 95.96 ± 6.68 | 99.00 ± 1.17 | 86.85 ± 13.55 | 93.60 ± 4.57 |
Statistically significantly different? (Y/N) | Y (p value <0.0001) |
Y (p value <0.0001) |
Y (p value <0.0001) |
Y (p value <0.0001) |
†Dynamic conformal arc therapy
‡Volumetric modulated arc therapy