Skip to main content
. 2021 Aug 31;29(3):125–134. doi: 10.12793/tcp.2021.29.e13

Table 3. Stability of rosuvastatin on volumetric absorptive microsampling under various stability conditions of 3 replicates.

Validation assessments Stability Condition Accuracy (%) %CV
LQC HQC LQC HQC
Bench top RT (5 hr) 110.33 104.33 0.43 3.16
Autosampler 4°C (24 hr) 100.53 97.10 7.12 3.09
Freeze and thaw −70°C (3 cycles) 112.00 101.90 0.73 1.72
Short-term RT (1 hr) 102.17 102.77 4.77 5.77
Long-term RT (30 days) 90.67 99.37 6.99 0.21