TABLE 2.

Procedure and post‐procedure characteristics

ICE group N = 175 TEE group N = 49 p
Procedure
Procedural success 171 (97) 49 (100) 0.895
Procedural SAE 2 (2) 2 (4.1) 0.590
Total time in the lab, min 66 ± 21 110 ± 27 0.004
Fluoroscopy time, min 21 ± 11 24 ± 12 0.803
Contrast used (ml) 48 ± 42 70 ± 28 0.016
Mean number of devices used 1.15 ± 0.49 1.20 ± 0.46 0.556
ACP device 154 (88) 46 (94) 0.394
16 mm 5 2
18 mm 4 3
20 mm 14 5
22 mm 33 14
24 mm 25 13
26 mm 25 4
28 mm 31 4
31 mm 11 1
34 mm 6 0
Watchman device 21 (12) 3 (6) 0.398
21 mm 4 0
24 mm 3 0
27 mm 9 3
30 mm 4 0
33 mm 1 0
In‐Hospital MACE
SAE prior to discharge 9 (5) 5 (10) 0.689
Vascular complication 1 (1) 2(4) 0.755
Cardiac tamponnade 2 (2) 2 (4) 0.532
Device migration 0 (0) 0 (0) 1.00
Device thrombus 3 (2) 1 (2) 0.610
Stroke/TIA 1 (0.5) 0 (0) 0.597
Death 0 0 1.00
Post‐procedure CT or TEE
Device thrombus 1 (0.5) 1 (2) 0.654
No peri‐device leak 134 (77) 39 (80) 0.349
Small leak (<5 mm) 29 (16) 4 (8) 0.306
Large leak (>5 mm) 9 (5) 3 (6) 0.999

Note: Data are presented as n (%),or Mean ± SD. SAE: Death, stroke, embolism, major bleed, device migration, major vascular complication.