TABLE 2.
ICE group N = 175 | TEE group N = 49 | p | |
---|---|---|---|
Procedure | |||
Procedural success | 171 (97) | 49 (100) | 0.895 |
Procedural SAE | 2 (2) | 2 (4.1) | 0.590 |
Total time in the lab, min | 66 ± 21 | 110 ± 27 | 0.004 |
Fluoroscopy time, min | 21 ± 11 | 24 ± 12 | 0.803 |
Contrast used (ml) | 48 ± 42 | 70 ± 28 | 0.016 |
Mean number of devices used | 1.15 ± 0.49 | 1.20 ± 0.46 | 0.556 |
ACP device | 154 (88) | 46 (94) | 0.394 |
16 mm | 5 | 2 | |
18 mm | 4 | 3 | |
20 mm | 14 | 5 | |
22 mm | 33 | 14 | |
24 mm | 25 | 13 | |
26 mm | 25 | 4 | |
28 mm | 31 | 4 | |
31 mm | 11 | 1 | |
34 mm | 6 | 0 | |
Watchman device | 21 (12) | 3 (6) | 0.398 |
21 mm | 4 | 0 | |
24 mm | 3 | 0 | |
27 mm | 9 | 3 | |
30 mm | 4 | 0 | |
33 mm | 1 | 0 | |
In‐Hospital MACE | |||
SAE prior to discharge | 9 (5) | 5 (10) | 0.689 |
Vascular complication | 1 (1) | 2(4) | 0.755 |
Cardiac tamponnade | 2 (2) | 2 (4) | 0.532 |
Device migration | 0 (0) | 0 (0) | 1.00 |
Device thrombus | 3 (2) | 1 (2) | 0.610 |
Stroke/TIA | 1 (0.5) | 0 (0) | 0.597 |
Death | 0 | 0 | 1.00 |
Post‐procedure CT or TEE | |||
Device thrombus | 1 (0.5) | 1 (2) | 0.654 |
No peri‐device leak | 134 (77) | 39 (80) | 0.349 |
Small leak (<5 mm) | 29 (16) | 4 (8) | 0.306 |
Large leak (>5 mm) | 9 (5) | 3 (6) | 0.999 |
Note: Data are presented as n (%),or Mean ± SD. SAE: Death, stroke, embolism, major bleed, device migration, major vascular complication.