Skip to main content
. 2021 Oct 7;11:19970. doi: 10.1038/s41598-021-99431-5

Figure 1.

Figure 1

Schematic, dose response and exposure characterization of the proposed UVC decontamination device. (A) Schematic of the operational germicidal device. The UVC germicidal device comprises a conveyor belt which is scalable. (B) Real image of the prototype germicidal device. (C) Changes to the dose response curve as a response belt speed. Total dose changes in response to the conveyor belt speed and exposure time. (D) Schematic of the experimental setup lamp and surface positioning. The germicidal lamp bulbs contain 3.8 cm between them and the distance from the lamp to the surface is 15.2 cm. Exposure changes as a result of positioning was monitored. (E) Exposure profile of the lamp based on the surface placement as indicated in (D). The UVC 254 nm detector was placed in positions 1 cm apart, covering a total distance of 7 cm. Measurements were taken at 3 min at each time point after initial lamp warm up.