. 2021 Oct 8;279(2):825–834. doi: 10.1007/s00405-021-07105-9
Mean difference (95% CI)
Background Coughing
Cold instruments
 Total particle concentration 1.039 (0.648 to 1.461) − 0.072 (− 0.575 to 0.431)
 < 1 µm particle concentration 1.003 (0.582 to 1.455) − 0.097 (− 0.605 to 0.411)
 1–5 µm particle concentration 0.612 (0.202 to 1.046) − 0.485 (− 1.002 to 0.032)
 > 5 µm particle concentration 0.226 (-0.305 to 0.857) 0.398 (− 0.149 to 0.944)
Microdebrider
 Total particle concentration 1.016 (0.718 to 1.315) − 0.273 (− 1.029 to 0.483)
 < 1 µm particle concentration 0.921 (0.544 to 1.298) − 0.317 (− 1.076 to 0.441)
 1–5 µm particle concentration 1.129 (0.294 to 1.965) − 0.443 (− 1.248 to 0.361)
 > 5 µm particle concentration 1.427 (0.597 to 2.256) 0.783 (− 0.029 to 1.596)
CO2 laser
 Total particle concentration 2.065 (1.852 to 2.279) 0.822 (0.461 to 1.182)
 < 1 µm particle concentration 2.091 (1.873 to 2.309) 0.867 (0.505 to 1.229)
 1–5 µm particle concentration 1.591 (1.457 to 1.724) 0.245 (− 0.102 to 0.592)
 > 5 µm particle concentration 0.294 (− 0.147 to 0.735) 0.341 (− 0.088 to 0.770)

Mean difference and 95% CI calculated from log-normalized values

CI confidence interval