Table 1.

Precision Studies of the Elecsys Anti–SARS-CoV-2 S-Ab Assaya

SARS-CoV-2 Controls Within Run (n = 20) Between Day (n = 23)
Mean, U/mL SD, U/mL CV, % Mean, U/mL SD, U/mL CV, %
Low positive control 0.96 0.014 1.50 1.02 0.03 3.2
High positive control 8.95 0.081 0.91 8.98 0.22 2.4

CV, coefficient of variation; S-Ab, spike protein antibody.

aA positive S-Ab assay result was defined by a measured S-Ab concentration ≥0.8 U/mL. CV (%) = (SD/mean) × 100.