Skip to main content
. 2021 Sep 1;23(9):601–608. doi: 10.1089/dia.2021.0097

Table 2.

Comparison of Glycemic Outcomes for Users Who Had ≥75% Continuous Glucose Monitoring Use (Baseline vs. 12-Month Use of Control-IQ)

  Baseline (Basal-IQ) 12-mth control-IQ use P
All users
 No. of participants 9010 9010  
 Mean sensor glucose [mg/dL] 164 (146–185) 152 (140–166) <0.001
 Sensor time <54 mg/dL [%] 0.10 (0.00–0.30) 0.15 (0.06–0.30) <0.001
 Sensor time 54–70 mg/dL [%] 0.8 (0.3–1.8) 0.9 (0.4–1.6) 0.053
 Sensor TIR [%] 63.6 (50.0–75.7) 73.6 (64.5–81.8) <0.001
 Sensor time 180–250 mg/dL [%] 25.1 (18.0–31.1) 19.7 (14.2–24.3) <0.001
 Sensor time >250 mg/dL [%] 8.1 (2.9–16.7) 4.6 (1.9–9.5) <0.001
 Coefficient of variation [%] 33.7 (30.0–37.6) 32.9 (29.5–36.3) <0.001
 GMI 7.2 (6.8–7.7) 6.9 (5.6–7.3) <0.001
T1DM users
 No. of participants 7813 7813  
 Mean sensor glucose [mg/dL] 163 (141–190) 151 (134–170) <0.001
 Sensor time <54 mg/dL [%] 0.01 (0.00–0.35) 0.02 (0.00–0.4) <0.001
 Sensor time 54–70 mg/dL [%] 0.9 (0.3–1.9) 0.9 (0.5–1.7) 0.123
 Sensor TIR [%] 63.2 (49.8–75.1) 73.5 (64.4–81.6) <0.001
 Sensor time 180–250 mg/dL [%] 25.2 (18.2–31.0) 19.7 (14.3–24.2) <0.001
 Sensor time >250 mg/dL [%] 8.3 (3.1–16.9) 4.7 (2.0–9.6) <0.001
T2DM users
 No. of participants 378 378  
 Mean sensor glucose [mg/dL] 158 (138–184) 150 (136–169) <0.001
 Sensor time <54 mg/dL [%] 0.00 (0.0–0.07) 0.04 (0.01–0.10) <0.001
 Sensor time 54–70 mg/dL [%] 0.2 (0.0–0.6) 0.2 (0.0–0.6) 0.337
 Sensor TIR [%] 69.9% (55.1–82.6) 78.0% (66.2–86.1) <0.001
 Sensor time 180–250 mg/dL [%] 23.9 (14.6–32.0) 19.0 (12.4–25.5) <0.001
 Sensor time >250 mg/dL [%] 3.6 (0.7–10.4) 2.3 (0.8–6.7) <0.001

Data are expressed as median (IQR) unless otherwise specified.

GMI, glucose management indicator; IQR, interquartile range; T1DM, type 1 diabetes; T2DM, type 2 diabetes; TIR, time in range.