Skip to main content
. 2021 Oct 11;10:144. doi: 10.1186/s13756-021-00993-w

Table 1.

Decontaminating N95/FFP2 masks for reuse: results of the systematic review 2011–June 2020

Title Authors Intervention Microorganisms tested Quantification method Outcome: log reduction compared to control Data on physical integrity/fit/filtration/residue Sterilization/probable disinfection (minimum 3-log reduction for disinfection, 6-log for sterilization) Method recommended (see text for the definition of this column) Comments
Liquids
Cleaning of filtering facepiece respirators contaminated with mucin and Staphylococcus aureus [2] Heimbuch et al Benzalkonium chloride (Wipes) Mucin or Staphylococcus aureus CFU assay 3–5 Fail Disinfection No
Relative survival of Bacillus subtilis spores loaded on filtering facepiece respirators after five decontamination methods [3] Lin et al Ethanol 50–95% Bacillus subtilis spores CFU assay N/A, but culture results were positive N/A Failed No
Effect of various decontamination procedures on disposable N95 mask integrity and SARS-CoV-2 infectivity [4] Smith et al Ethanol 70% SARS-CoV-2 RT-PCR likely > 3 log Fail Disinfection No The initial contamination was at 3 log as it was coming from human sample
Effectiveness of N95 Respirator Decontamination and Reuse against SARS-CoV-2 Virus [5] Fischer et al Ethanol 70% for 10 min SARS-CoV-2 (HCoV-19 nCoV-WA1-2020 (MN985325.1)) TCID50 4 Fail Disinfection No
Relative survival of Bacillus subtilis spores loaded on filtering facepiece respirators after five decontamination methods [3] Lin et al Hypochlorite in a 0.54% solution Bacillus subtilis spores CFU assay N/A, but culture results were negative N/A Disinfection N/A No colonies but no info on log reduction
Cleaning of filtering facepiece respirators contaminated with mucin and Staphylococcus aureus [2] Heimbuch et al Hypochlorite (Wipes) in a 0.9% solution S. aureus CFU assay 4–5 except for nose pads Fail Disinfection and failure No All masks had good disinfection except for on the nose pads (less than 2 log reduction)
Cleaning of filtering facepiece respirators contaminated with mucin and Staphylococcus aureus [2] Heimbuch et al Nonantimicrobial detergent wipes Mucin or Staphylococcus aureus CFU assay 1 Fail Failure No
Heat
Effectiveness of N95 Respirator Decontamination and Reuse against SARS-CoV-2 Virus [5] Fischer et al Dry heat at 70 °C for 10–60 min SARS-CoV-2 (HcoV-19 nCoV-WA1-2020 (MN985325.1) TCID50  > 1to > 3 depending on time Pass (max 3 rounds) Disinfection and failure Yes Ability to disinfect was time dependent
Effectiveness of Ultraviolet-C Light and a High-Level Disinfection Cabinet for Decontamination of N95 Respirators [6] Cadnum et al Dry heat at 70 °C for 30 min Staphylococcus aureus (MRSA) and bacteriophages MS2 and Phi6 CFU assay, Plaque assay Bacteriophages < 1, MRSA > 4 N/A Disinfection and failure No Failure for bacteriophages
It’s not the heat, it’s the humidity: Effectiveness of a rice cooker-steamer for decontamination of cloth and surgical face masks and N95 respirators [7] Li et al Dry heat at 100 °C for 15 min MS2 ad MRSA Plaque assay  < 3 log10 reduction Pass (visual) Failure No
Validation of N95 filtering facepiece respirator decontamination methods available at a large university hospital [8] Wigginton et al Dry heat at 82 °C for 30 min Staphylococcus aureus and Geobacillus stearothermophilus Plaque assay S. aureus: < 1.0 log10, G. stearotherophilus: < 0.3 log10 Pass Failure No
Decontamination of face masks and filtering facepiece respirators via ultraviolet germicidal irradiation, hydrogen peroxide vaporization, and use of dry heat inactivates an infectious SARS-CoV-2 surrogate virus [9] Ludwig-Begall et al Dry heat at 102 °C for 60 min Porcine respiratory coronavirus (PRCV) TCID50  ≥ 4 N/A Disinfection N/A
Decontamination of Surgical Face Masks and N95 Respirators by Dry Heat Pasteurization for One Hour at 70°C [10] Xiang et al Dry heat at 60 °C and 70 °C 60 min E. coli, S. aureus, P. aeruginosa, K. pneumonia, A.baumannii, C. pseudodiphtheria, and C.albicans, Inf A virus (H1N1) TCID 50 N/A, but culture results were negative Pass Disinfection Yes
Dry Heat as a Decontamination Method for N95 Face Respirator Reuse [11] Oh et al Dry heat at 120 °C for 50 min

Tulane virus, rotavirus, adenovirus, transmissible gastroenteritis

virus

Plaque assay Tulane > 5.2, rotavirus > 6.6, adenovirus > 4.0, gastroenteritis > 4.7 Pass Sterilization and disinfection Yes All disinfection except for rotavirus which reached sterilization
Thermal Disinfection Inactivates SARS-CoV-2 in N95 Respirators while Maintaining Their Protective Function [12] Daeschler et al Dry heat at 70 °C for 60 min SARS-CoV-2 and E. coli TCID 50 mixed: > 4 for SARS CoV-2, < 1 for E.coli Pass Disinfection and failure Depends Failure for E.Coli. Recommendation dependent on microorganism present
Relative survival of Bacillus subtilis spores loaded on filtering facepiece respirators after five decontamination method [3] Lin et al Dry Heat (electric rice cooker) 149–164 °C Bacillus subtilis spores CFU assay N/A, culture results were mixed N/A Failure N/A Possibly disinfection after 24 h, but not immediately
Efficacy of moist heat decontamination against various pathogens for the reuse of N95 respirators in the COVID-19 emergency [13] Oral et al Moist heat at 60 °C at 80% humidity for 30 min Bovine viral diarrhea virus (BVDV), Porcine Parvovirus (PPV) and Influenza A Virus, S. aureus, P. Aeruginosa and A. Baumanii cell culture S. aureus: 5.32 P.aeruginosa:5.7 A. Baumannii:5.92 InfA:4.35 modestly BVDV:1.62 PPV:0 N/A Disinfection and failure N/A Failure for BVDV and PPV
A pandemic influenza preparedness study: Use of energetic methods to decontaminate filtering facepiece respirators contaminated with H1N1 aerosols and droplets [14] Heimbuch et al Moist heat at 65 °C at 85% humidity for 30 min H1N1 TCID50  > 4 Pass Disinfection Yes There was only a visual examination for fit and integrity
Biological Aerosol Test Method and Personal Protective Equipment (PPE) Decon [15] Hinrichs et al Moist heat at 62 °C at 85% humidity for 20 min influenza virus A (H5N1) RT-PCR and TCID50  ≥ 4 N/A Disinfection N/A
Effect of moist heat decontamination on methicillin-sensitive S. aureus for the reuse of N95 respirators in the COVID-19 emergency [16] Gil et al Moist heat at 60 °C at 80% humidity for 30 min S. aureus (methicillin sensitive) CFU assay 5.31 N/A Disinfection N/A
Thermal Disinfection Inactivates SARS-CoV-2 in N95 Respirators while Maintaining Their Protective Function [12] Daeschler et al Moist heat at 70 °C at 50% humidity for 60 min SARS-CoV-2 and E. coli TCID 50 mixed: > 4 for SARS CoV-2, < 3 for E.coli Pass Disinfection and failure Depends Failure for E.Coli because initial contamination was too low, but was probably disinfection. Recommendation dependent on microorganism present
Effectiveness of three decontamination treatments against influenza virus applied to filtering facepiece respirators [17] Lore et al Moist heat at 65 °C for 20 min H5N1 TCID50  ≥ 4.62 and ≥ 4.65 Pass Disinfection Yes
It’s not the heat, it’s the humidity: Effectiveness of a rice cooker-steamer for decontamination of cloth and surgical face masks and N95 respirators [7] Li et al Moist heat for 12–15 min MS2 ad MRSA Plaque assay  > 5 log10 reduction Pass Disinfection Yes There was only a visual examination for fit and integrity
Validation of N95 filtering facepiece respirator decontamination methods available at a large university hospital [8] Wigginton et al Moist heat at 80 °C at 60% humidity for 30 min MS2, phi6, influenza A virus S aureus, G. stearotherophilus Plaque assay MS2: > 6.8, Phi6: > 6.6, influenza virus: > 3.4, and MHV > 1.4, S disinfection > 2.9, G. stearotherophilus < 0.3 Pass Sterilization, disinfection and failure No MS2: Sterilization, Phi6: Sterilization, Influenza virus: Disinfection, MHV: failed, S aureus: failed, G. stearotherophilus: failed
A pandemic influenza preparedness study: Use of energetic methods to decontaminate filtering facepiece respirators contaminated with H1N1 aerosols and droplets [14] Heimbuch et al Moist heat (microwave-generated) for 2 min H1N1 TCID50  > 4 Pass Disinfection Yes There was only a visual examination for fit and integrity
Biological Aerosol Test Method and Personal Protective Equipment (PPE) Decon [15] Hinrichs et al Moist heat (microwave-generated) for 2 min influenza virus AH5N1 RT-PCR and TCID50  ≥ 4 N/A Disinfection N/A
Effectiveness of three decontamination treatments against influenza virus applied to filtering facepiece respirators [17] Lore et al Moist heat (microwave-generated) for 2 min H5N1 TCID50  ≥ 4.81 and ≥ 4.79 Pass Disinfection Yes
Evaluation of microwave steam bags for the decontamination of filtering facepiece respirators [18] Fisher et al Moist heat (microwave-generated) for 1.5 min MS2 bacteriophage CFU assay 3.10 – 4.64 Mixed Disinfection Mixed Can be only recommended in some cases, depending on model of the mask and how much water is absorbed. Some failure for physical integrity/fit/filtration
Microwave-Generated Steam Decontamination of N95 Respirators Utilizing Universally Accessible Materials [19] Zulauf et al Moist heat (microwave-generated) for 3 min for 1, 5, or 20 cycles Escherichia coli MS2 bacteriophage Plaque assay 5–6 Pass Sterilization and disinfection Yes Average 6-log10 PFU and a minimum 5-log10 PFU reduction after a single three-minute microwave treatment
Steam treatment for rapid decontamination of N95 respirators and medical face masks [20] Li et al Steam (autoclave, short cycle) at 100 °C for 10–30 s S. aureus (MRSA), G. stearothermophilus spores, bacteriophage MS2 CFU assay MS2 and MRSA > 3, G. stearothermophilus spores: fail Pass Disinfection and failure No Failure for G. stearothermophilus spores. Authors also tested a 2 s cycle, but the test failed
Relative survival of Bacillus subtilis spores loaded on filtering facepiece respirators after five decontamination methods [3] Lin et al Steam (autoclave) at 121 °C for 15 min Bacillus subtilis spores CFU assay N/A, but culture results were negative N/A Disinfection N/A No colonies but no info on log reduction
N95 mask decontamination using standard hospital sterilization technologies [21] Kumar et al Steam (autoclave) at 121 °C for 40 min Vesicular stomatitis virus, Indiana serotype (VSV) or SARSCoV-2 (contaminated group) TCID50 VSV: > 6, SARSCoV-2: 5.2–6.3 Pass Disinfection and Sterilization Yes Some FFRs may have had too low a level of contamination to ensure a 6-log reduction
Gases
N95 mask decontamination using standard hospital sterilization technologies [21] Kumar et al Ethylene oxide (EtO) for 60 min Vesicular stomatitis virus, Indiana serotype (VSV) TCID50 VSV: > 6 Pass Sterilization No
Validation of N95 filtering facepiece respirator decontamination methods available at a large university hospital [8] Wigginton et al Ethylene oxide (EtO) 55 °C for 60 min at 45% RH MS2 Plaque assay  > 5.8 Pass Disinfection No
Validation of N95 filtering facepiece respirator decontamination methods available at a large university hospital [8] Wigginton et al Hydrogen peroxide (gaseous HPGP) in a 59% solution for 24 min MS2, phi6, influenza A virus Plaque assay Phi6: > 7.9, influenza virus > 3.8, MS2: 5.6 Pass Sterilization and disinfection No Sterilization for Phi6, disinfection for influenza virus and MS2
Effect of various decontamination procedures on disposable N95 mask integrity and SARS-CoV-2 infectivity [4] Smith et al Hydrogen peroxide (gaseous HPV) in a 30% solution (500 ppm) at humidity between 38–99.5% for 20 min SARS-CoV-2 RT-PCR 2 masks: ~ five log10 reduction < 3 log Pass Disinfection No The initial contamination was at 3 log as it was coming from human sample
Aerosolized Hydrogen Peroxide Decontamination of N95 Respirators, with Fit-Testing and Virologic Confirmation of Suitability for Re-Use During the COVID-19 Pandemic [22] Derr et al Hydrogen Peroxide (Gaseous-aHP) in a 7% solution for 12 min SARS-CoV-2, Herpes simplex virus 1, Coxsackie virus B3, Pseudomonas phi6 bacteriophage Plaque assay N/A, but culture results were negative Pass Sterilization Yes
Vapor H2O2 sterilization as a decontamination method for the reuse of N95 respirators in the COVID-19 emergency [23] Oral et al Hydrogen Peroxide (Gaseous-HPV) 410 ppm for 180 min SARS-CoV-2 Plaque assay > 2.6 Pass Disinfection Yes The initial contamination was too low to be able to detect sterilization
N95 mask decontamination using standard hospital sterilization technologies [21] Kumar et al Hydrogen peroxide (gaseous-HPV) in a 35% solution (750 ppm) for 60 min Vesicular stomatitis virus, Indiana serotype (VSV) or SARSCoV-2 (contaminated group) TCID50 VSV: > 6, SARSCoV-2: 5.2–6.3 Pass Sterilization and disinfection Yes Some FFRs may have had too low a level of contamination to ensure a 6-log reduction
Effectiveness of N95 Respirator Decontamination and Reuse against SARS-CoV-2 Virus [5] Fischer et al Hydrogen peroxide (gaseous-HPV) (1000 ppm) for 10 min SARS-CoV-2 (HCoV-19 nCoV-WA1-2020 (MN985325.1)) TCID50  > 4 Pass Disinfection Yes
Validation of N95 filtering facepiece respirator decontamination methods available at a large university hospital [8] Wigginton et al Hydrogen peroxide (gaseous-HPV) 446–659 ppm MS2, phi6, influenza A virus, murine hepatitis virus, E. coli, S. aureus, G. stearothermophilus, A. niger Plaque assay  > 2 Pass Failure No
Decontamination of face masks and filtering facepiece respirators via ultraviolet germicidal irradiation, hydrogen peroxide vaporization, and use of dry heat inactivates an infectious SARS-CoV-2 surrogate virus [9] Ludwig-Begall et al Hydrogen peroxide (gaseous-HPV) in a 59% solution (750 ppm) for 28 min Porcine respiratory 38 coronavirus (PRCV) TCID50  ≥ 5 N/A Disinfection N/A
Hydrogen Peroxide Vapor sterilization of N95 respirators for reuse [24] Kenney et al Hydrogen peroxide (gaseous-HPV) 30–40-min gassing phase at 16 g/min Phages phi-6, T7 and T1 Plaque assay, TCID50 N/A, but complete eradication of phages from masks Pass Sterilization Yes Limit of detection was 5PFU, lower than infectious dose, and authors used the term "sterilization"
Disinfection of N95 respirators by ionized hydrogen peroxide during pandemic coronavirus disease 2019 (COVID-19) due to SARS-CoV-2 [25] Cheng et al Hydrogen Peroxide (Gaseous-iHP) in a 7.8% solution Influenza A virus subtype H1N1 TCID 50 N/A, but culture results were negative N/A Disinfection N/A No growth, but no specific log reduction mentioned, paper uses term "disinfection"
N95 mask decontamination using standard hospital sterilization technologies [21] Kumar et al Hydrogen peroxide (gaseous-LT-HPGP) in a 59% solution for 47 min Vesicular stomatitis virus, Indiana serotype (VSV) TCID50 VSV: > 6 Pass Sterilization No LT-HPGT-treated masks failed testing beyond the first cycle
Effectiveness of Ultraviolet-C Light and a High-Level Disinfection Cabinet for Decontamination of N95 Respirators [6] Cadnum et al Hydrogen Peroxide (Gaseous) and Peracetic acid for 1–3 cycles of 21 min, and a single cycle of 31 min S. aureus (MRSA) and bacteriophages MS2 and Phi6 CFU assay, Plaque assay 1 cycle: > 2.1, 2 cycles: > 3.6, 3 cycles > 6 log10 N/A Sterilization, disinfection, failure N/A Outcome was dependent on the number of cycles (3 cycles resulted in sterilization)
Scalable In-hospital Decontamination of N95 Filtering Facepiece Respirator with a Peracetic Acid Room Disinfection System [26] John et al Hydrogen Peroxide (Gaseous) and Peracetic acid in a 18% solution at at 20 °C for 12–19 min MS2 bacteriophage and G. stearothermophilus spores CFU assay 6/6/4 Pass Sterilization and disinfection Yes Shorter cycle led to disinfection. Can't be used with masks containing cellulose
Enveloped Virus Inactivation on Personal Protective Equipment by Exposure to Ozone [27] Blanchard et al Ozone at 20 ppm and 70% humidity for 40 min

Influenza virus A

A/WSN/33, RSV A2

Plaque assay 4 Mixed Disinfection Mixed Although the facepiece was unaffected for fit/filtration, the elastic band failed
Fast and easy disinfection of coronavirus-contaminated face masks using ozone gas produced by a dielectric barrier discharge plasma generator [28] Lee et al Ozone at 120 ppm for 1 and 5 min HCoV-229E TCID 50 3 Pass Disinfection Yes The initial contamination was too low to be able to detect sterilization
Disinfection of N95 Respirators with Ozone [29] Manning et al Ozone at 450 ppm and 75–90% humidity for 120 min Pseudomonas aeruginosa CFU assay > 7–> 9, (one sample failed disinfection-1.38 log reduction) Mixed Sterilization and failure Mixed There was a single failure, but one needs to verify why one mask failed the test. Although the facepiece was unaffected for fit/filtration, the elastic band failed
N95 mask decontamination using standard hospital sterilization technologies [21] Kumar et al Peracetic acid dry fogging system (PAF) at 80–90% humidity for 60 min Vesicular stomatitis virus, Indiana serotype (VSV) or SARSCoV-2 (contaminated group) TCID50 VSV: > 6, SARSCoV-2: 5.2–6.3 Pass Disinfection and sterilization Yes Some FFRs may have had too low a level of contamination to ensure a 6 log reduction
Ultra violet light
Relative survival of Bacillus subtilis spores loaded on filtering facepiece respirators after five decontamination methods [3] Lin et al UVA at 365 nm and 1.87–37.44 J/cm2 for 1–20 min Bacillus subtilis spores CFU assay N/A, but culture results were positive N/A Failed No
A Scalable Method for Ultraviolet C Disinfection of Surgical Facemasks Type IIR and Filtering Facepiece Particle Respirators 1 and 2 [30] Lede et al UVGI at 253.7 nm and 6 lamps, each 0.6 J/cm2 for 40 min S. aureus CFU assay 7 Pass Sterilization Yes
Ultraviolet germicidal irradiation of influenza-contaminated N95 filtering facepiece respirators [31] Mills et al UVGI at 254 nm and 1.1 J/cm2 and 48% humidity for 40 min Influenza virus (H1N1) TCID 50  ≥ 3 log on 12 of 15 FFR models and straps from 7 of 15 FFR models N/A Disinfection and failure N/A
A method to determine the available UV-C dose for the decontamination of filtering facepiece respirators [32] Fisher and Shaffer UVGI at 254 nm and 0.15–1.5 J/cm2 for 1–10 min MS2 coliphage Plaque assay minimum IFM dose of 1000 J m^-2: log reduction >  = 3 Mixed disinfection No

Model dependent outcomes. Model-specific exposure times to achieve this IFM dose

Ranged from 2 to 266 min. Mostly failure for physical integrity/fit/filtration

The Effect of Ultraviolet C Radiation Against SARS-CoV-2 Inoculated N95 Respirators [33] Ozog et al UVGI at 254 nm and 1.5 J/cm2 for 60–70 s/side SARS-CoV-2 TCID 50 N/A, culture results were mixed N/A Disinfection and failure N/A Disinfection but not for all models of masks (5 models of N95 tested)
Effects of relative humidity and spraying medium on UV decontamination of filters loaded with viral aerosols [34] Woo et al UVGI at 254 nm and 1.8 and 3.6 J/cm2 and 30,60, and 90% humidity for 30 and 60 min MS2 Plaque assay mixed, highest inactivation efficiency: 5.8 log N/A Disinfection and failure N/A Disinfection, but not for all masks and conditions
Relative survival of Bacillus subtilis spores loaded on filtering facepiece respirators after five decontamination methods [3] Lin et al UVGI at 254 nm and 1.13–22.68 J/cm2 for 1–20 min Bacillus subtilis spores CFU assay N/A, culture results were mixed N/A Failure N/A Possible disinfection after 24 h, but not immediately
Effectiveness of three decontamination treatments against influenza virus applied to filtering facepiece respirators [17] Lore et al UVGI at 254 nm and 18 kJ/m2 for 15 min H5N1 TCID50  ≥ 4.54 and ≥ 4.65 Pass Disinfection Yes
A pandemic influenza preparedness study: Use of energetic methods to decontaminate filtering facepiece respirators contaminated with H1N1 aerosols and droplets [14] Heimbuch et al UVGI at 254 nm and 18 kJ/m2 for 15 min CHECK SAME AS LORE) H1N1 TCID50  > 4 Pass Disinfection Yes There was only a visual examination for fit and integrity
Effectiveness of N95 Respirator Decontamination and Reuse against SARS-CoV-2 Virus [5] Fischer et al UVGI at 260-285 nm and 0.33 J/cm2, 0.99 J/cm2, and 1.98 J/cm2 for 10, 30, and 60 min SARS-CoV-2 (HCoV-19 nCoV-WA1-2020 (MN985325.1)) TCID50 between 1 and 3, depending on the time Pass Disinfection and failure Yes Time-dependent: failure for masks below 60 min, probable disinfection at 60 min
Effectiveness of Ultraviolet-C Light and a High-Level Disinfection Cabinet for Decontamination of N95 Respirators [6] Cadnum et al UVGI for 1 and 30 min S. aureus (MRSA) and bacteriophages MS2 and Phi6 CFU assay, Plaque assay 0–4 N/A Disinfection and failure No Outcome depended on model of mask and pathogen, only 1 of 9 masks qualified as disinfected
Validation of N95 filtering facepiece respirator decontamination methods available at a large university hospital [8] Wigginton et al UVGI at 200-315 nm for 5 min MS2, phi6, influenza A virus, murine hepatitis virus, E. coli, S. aureus, G. stearothermophilus, A. niger Plaque assay MS2: 0.7 – 1.3, Phi6: 0.2 – 1.8, influenza: 1.4 – 1.7, MHV > 1.4, S. aureus < 1.0, G. stearotherophilus < 0.3 log10 Pass Failure No
Effect of various decontamination procedures on disposable N95 mask integrity and SARS-CoV-2 infectivity [4] Smith et al UVGI at 254 nm and 0.63 J/cm2 for 33 min SARS-CoV-2 RT-PCR  < 3 log Fail Neither No The initial contamination was at 3 log as it was coming from human sample
Decontamination of face masks and filtering facepiece respirators via ultraviolet germicidal irradiation, hydrogen peroxide vaporisation, and use of dry heat inactivates an infectious SARS-CoV-2 surrogate virus [9] Ludwig-Begall et al UVGI at 254 nm and 5.2 J/cm [2] for 4 min Porcine respiratory 38 coronavirus (PRCV) TCID50  ≥ 4 N/A Disinfection N/A
Disinfection effect of pulsed xenon ultraviolet irradiation on SARS-CoV-2 and implications for environmental risk of COVID-19 transmission [35] Simmons et al UVGI for 5 min SARS CoV-2 Plaque assay  > 4.79 N/A Disinfection N/A
Biological Aerosol Test Method and Personal Protective Equipment (PPE) Decon [15] Hinrichs et al UVGI at 254 nm and 18 kJ/m2 for 15 min influenza virus AH5N1 RT-PCR and TCID50  ≥ 4 log10 TCID50 N/A Disinfection N/A
Reusability of filtering facepiece respirators after germicidal UV irradiation [36] Vernez et al UVGI + dry heat (Dry Heat at 70 °C for 15 min and then UVGI at 254 nm and 60 mJ/cm2 for 4 min) vB_HSa_2002 and P66 phages Plaque assay  > 3 Pass Disinfection Yes
Validation of N95 filtering facepiece respirator decontamination methods available at a large university hospital [8] Wigginton et al UVGI + dry heat (Dry heat at 82 °C and UVGI at 200–315 nm) MS2, phi6, influenza A virus, murine hepatitis virus, Staphylococcus aureus Plaque assay The influenza virus: > 3.9, the mouse coronavirus: 1.1, Phi6 deposited in PBS < 1.5 when heated to 82C and at ~ 8% RH. S. aureus: 1.2 Pass Disinfection and failure No Influenza virus: Disinfection MHV: Failed MS2: Failed, Phi6: Failed S. aureus: failed
Validation of N95 filtering facepiece respirator decontamination methods available at a large university hospital [8] Wigginton et al UVGI + medium humidity heat (Heat at 80 °C, RH at 62–66% and UVGI at 200–315 nm for 15 min) MS2, phi6, influenza A virus Plaque assay influenza virus: > 3.9, mouse coronavirus MHV > 1.1, MS2 > 6.8, Phi6 > 6.6 Pass Sterilization, disinfection, failure No MS2: Sterilization, Phi6: Sterilization, Influenza virus: Disinfection, MHV: failed