Table 1.

Quantitative image analysis of pediatric CT with different reconstruction techniques in comparison with 50% ASIR-V

Parameters ASIR-V 50 ASIR-V 100 DLR-M DLR-H
Chest CT without contrast enhancement (n = 16)
Attenuation (HU) lung − 795.9 ± 91.1 − 796.2 ± 94.2 0.872 − 797.6 ± 94.3 0.872 − 796.7 ± 93.9 0.872
paraspinal muscle 56.1 ± 11.1 54.6 ± 12.1 0.552 56.3 ± 10.7 0.552 54.9 ± 9.8 0.552
Noise 21.8 ± 3.7 15.6 ± 10.9 0.159 20.2 ± 3.8 0.455 14.6 ± 2.5  < 0.001
CNR Lung 11.4 ± 3.7 22.9 ± 6.9  < 0.001 25.8 ± 13.3  < 0.001 28.4 ± 11.4  < 0.001
SNR Lung 10.7 ± 3.5 21.4 ± 6.4  < 0.001 24.1 ± 12.5  < 0.001 26.6 ± 10.8  < 0.001
Chest CT with contrast enhancement (n = 12)
Attenuation (HU) Lung − 718.9 ± 139.4 − 718.7 ± 144.0 0.989 − 720.5 ± 145.4 0.989 − 719.1 ± 142.8 0.989
Paraspinal muscle 65.0 ± 8.5 64.6 ± 5.8 0.425 64.3 ± 5.2 0.425 62.6 ± 6.8 0.425
Noise 24.5 ± 6.1 12.6 ± 3.8  < 0.001 21.3 ± 4.9 0.172 14.8 ± 4.7  < 0.001
CNR Lung 10.4 ± 4.0 17.8 ± 8.1 0.010 20.0 ± 6.9 0.001 21.4 ± 8.6  < 0.001
SNR Lung 9.6 ± 3.7 16.4 ± 7.6 0.011 18.4 ± 6.6 0.001 19.8 ± 8.1  < 0.001
Abdomen CT with contrast enhancement (n = 23)
Attenuation (HU) Liver 131.4 ± 28.4 131.7 ± 28.7 0.369 125.9 ± 39.1 0.369 132.6 ± 28.6 0.369
Aorta 185.7 ± 45.2 184.9 ± 45.9 1.000 185.3 ± 45.1 0.227 187.6 ± 45.4 0.085
Paraspinal muscle 71.2 ± 8.4 71.3 ± 6.6 1.000 72.7 ± 6.0 0.535 71.1 ± 6.2 1.000
Noise 19.9 ± 3.7 11.1 ± 3.6  < 0.001 16.3 ± 3.1 0.002 12.2 ± 2.4  < 0.001
CNR Liver 3.2 ± 1.7 5.3 ± 3.0  < 0.001 3.2 ± 2.4 1.000 4.9 ± 2.5  < 0.001
Aorta 5.3 ± 2.2 9.5 ± 4.3  < 0.001 5.8 ± 1.9 0.012 8.0 ± 2.9  < 0.001
SNR Liver 6.8 ± 2.1 11.5 ± 3.9  < 0.001 7.6 ± 2.7 0.583 10.7 ± 2.9  < 0.001
Aorta 8.5 ± 2.4 15.6 ± 5.1  < 0.001 9.7 ± 2.0 0.002 13.0 ± 3.2  < 0.001

The p-values < 0.05 were marked as bold

Values are presented as the mean ± standard deviation

ASIR-V 50, 50% adaptive statistical iterative reconstruction-V

ASIR-V 100 100% ASIR-V, DLR-M medium strength deep learning reconstruction, DLR-H, high strength DLR, CNR contrast to noise ratio, SNR signal to noise ratio. Image noise is based on standard deviation of paraspinal muscle attenuation