Table 3.

Image Quality and Lesion Detection

SD-IR LD-IR LD-DLIR p-value
Image quality
 Sharpness (ĸ value) 0.31 0.24 0.68 N.A.
 Reader 1 4.0 ± 0.2 (4.0–5.0) 3.2 ± 0.4a (3.0–4.0) 4.1 ± 0.3 (4.0–5.0) <0.001
 Reader 2 4.1 ± 0.3 (4.0–5.0) 3.4 ± 0.5a (3.0–4.0) 4.1 ± 0.3 (4.0–5.0) <0.001
 Image noise (ĸ value) 0.41 0.66 0.61 N.A.
 Reader 1 3.9 ± 0.3 (3.0–4.0) 3.0 ± 0.2a (2.0–3.0) 4.1 ± 0.4b (3.0–5.0) <0.001
 Reader 2 3.9 ± 0.3 (3.0–5.0) 2.9 ± 0.3a (2.0–4.0) 4.2 ± 0.5b (3.0–5.0) <0.001
 Diagnostic acceptability (ĸ value) 0.65 0.59 0.42 N.A.
 Reader 1 4.0 ± 0.2 (3.0–4.0) 3.1 ± 0.3a (2.0–4.0) 4.1 ± 0.3 (3.0–5.0) <0.001
 Reader 2 4.0 ± 0.3 (3.0–5.0) 3.1 ± 0.4a (2.0–4.0) 4.0 ± 0.3 (3.0–5.0) <0.001
Lesion detection
 Thyroid nodule
 Reader 1 100% (22/22) 100% (22/22) 100% (22/22) 1.00
 Reader 2 91% (20/22) 86% (19/22) 86% (19/22) 1.00
 Breast mass
 Reader 1 100% (2/2) 100% (2/2) 100% (2/2) 1.00
 Reader 2 100% (2/2) 100% (2/2) 100% (2/2) 1.00
 Lymphadenopathy
 Reader 1 100% (4/4) 100% (4/4) 100% (4/4) 1.00
 Reader 2 100% (4/4) 100% (4/4) 100% (4/4) 1.00
 Liver cyst
 Reader 1 100% (34/34) 97% (33/34) 97% (33/34) 1.00
 Reader 2 91% (31/34) 91% (31/34) 91% (31/34) 1.00
 Hepatic hemangioma
 Reader 1 100% (8/8) 100% (8/8) 100% (8/8) 1.00
 Reader 2 88% (7/8) 88% (7/8) 88% (7/8) 1.00
 Liver metastasis
 Reader 1 100% (3/3) 100% (3/3) 100% (3/3) 1.00
 Reader 2 100% (3/3) 100% (3/3) 100% (3/3) 1.00
 Pancreatic cyst
 Reader 1 100% (2/2) 100% (2/2) 100% (2/2) 1.00
 Reader 2 100% (2/2) 100% (2/2) 100% (2/2) 1.00
 Peritoneal dissemination
 Reader 1 100% (2/2) 100% (2/2) 100% (2/2) 1.00
 Reader 2 100% (2/2) 100% (2/2) 100% (2/2) 1.00
 Adrenal nodule
 Reader 1 100% (2/2) 100% (2/2) 100% (2/2) 1.00
 Reader 2 100% (2/2) 100% (2/2) 100% (2/2) 1.00
 Renal cyst
 Reader 1 95% (35/37) 89% (33/37) 95% (35/37) 0.73
 Reader 2 100% (37/37) 97% (36/37) 97% (36/37) 1.00
 Uterine mass
 Reader 1 86% (6/7) 86% (6/7) 86% (6/7) 1.00
 Reader 2 86% (6/7) 86% (6/7) 86% (6/7) 1.00
 Ovarian mass
 Reader 1 100% (4/4) 100% (4/4) 100% (4/4) 1.00
 Reader 2 100% (4/4) 100% (4/4) 100% (4/4) 1.00

N.A., not applicable. LD-DLIR, low-dose with deep learning image reconstruction.LD-IR, low-dose with adaptive statistical iteration reconstruction Veo; SD-IR, standard-dose with hybrid-iterative reconstruction technique.

Data are means ± 1 standard deviation.

a

Value was significantly lower (p < 0.017) than those with SD-IR and LD-DLIR.

b

Value was significantly greater (p < 0.017) than those with SD-IR.