. 1999 Jul;37(7):2223–2229. doi: 10.1128/jcm.37.7.2223-2229.1999

TABLE 2.

Comparison of PACE 2, LCx, and AMP CT assays for detection of C. trachomatis in 738 endocervical swab specimensa

Result by the following assayb:
No. of specimens
Gen-Probe PACE 2 Abbott LCx Gen-Probe AMP CT
Pos Pos Pos  50
Neg Pos Pos  10
Neg Neg Pos   2
Neg Neg Neg 673
a

There were three false-positive specimens, one by each assay, for which data are not presented. 

b

Pos, positive; Neg, negative. By the PACE 2, LCx, and AMP CT assays, 50 (80.6%), 60 (96.8%), and 62 (100%) of the specimens were positive, respectively (percentages were calculated as described in footnote a of Table 1).