. 2021 Sep 23;21(19):6363. doi: 10.3390/s21196363

Table 2.

Validity criteria of GAR-derived SO2 values compared to those obtained with a medical device (COV) and with arterial blood gas analysis (ABG) at 4559 m.

ICC MAPE [%] Pearson’s r p-Value
GAR vs. COV (n = 49) 0.661 6.81 0.537 0.011 *
GAR vs. ABG (n = 37) 0.549 9.77 0.380 <0.001 *
COV vs. ABG (n = 26) 0.883 6.15 0.904 0.979

ICC = intraclass correlation coefficient; MAPE = mean absolute percentage error; GAR = Garmin Fēnix® 5X Plus; COV = Covidien Nellcor Portable SpO2 Patient Monitoring; ABG = Radiometer ABL 90 Flex; * p < 0.05; p-values were derived via unpaired t-test.