Skip to main content
. 2021 Oct 15;11:20553. doi: 10.1038/s41598-021-99831-7

Table 3.

PMT voltage independent measure of spread: two different sets of PMT voltages were applied to set the MFI of the positive population to 110,000 (110 K) or 50,000 (50 K).

graphic file with name 41598_2021_99831_Tab3_HTML.jpg

ISS values are shown in Italics.

Six single stained bead samples were run using these two voltage sets, and the ISS and SQI values were calculated. The SQI values were similar at the different voltages, whereas the ISS values were voltage different. ISS values are shown in Italics.