Table 2.

Postoperative Vault, Anterior Chamber Depth and Pupil Size Changes with Myopic and Toric ICL under Different Lighting Conditions

Characteristics Myopic ICL Toric ICL
Mesopic Photopic Mesopic Photopic
ICLVault (μm)
 Mean ± SD 703.5 ± 259.9 587.8 ± 251.8 744.8 ± 244.7 626.2 ± 230.0
 Range 266 to 1430 171 to 1209 297 to 1465 228 to 1287
 △ (Mean/SD/range)

115.7 ± 64.9 (14 to 288)

±

118.1 ± 61.7 (19 to 326)

to

ACD (mm)
 Mean ± SD 3.17 ± 0.22 3.17 ± 0.22 3.23 ± 0.22 3.24 ± 0.23
 Range 2.79 to 3.70 2.64 to 3.71 2.66 to 3.65 2.67 to 3.72
 △ (Mean/SD/range) 0.01 ± 0.06 (− 0.17 to 0.26) ± − 0.01 ± 0.05 (− 0.31 to 0.11) to
Pupil Size (mm)
 Mean ± SD 5.04 ± 0.74 3.14 ± 0.57 4.80 ± 0.82 3.15 ± 0.55
 Range 3.27 to 6.562 1.86 to 4.41 3.11 to 7.04 1.93 to 4.77
 △ (Mean/SD/range)

1.91 ± 0.70 (0.50 to 3.45)

±

1.66 ± 0.78 (0.41 to 3.65)

to

ICL Implantable collamer lens, ACD Anterior chamber depth, △: Change; SD Standard deviation

27.50 ± 6.12

31.89 ± 6.17

− 4.38 ± 4.65