Table 1:

Correction factors needed for determination of dose rate in Protocol 1.

Correction Factor Explanation
Nκ Air kerma calibration factor
[(μen/ρ)Wair]water Ration of mass energy absorption coefficients of water to air; approximately 1.05
Pq,Cham Correction accounting for chamber stem affecting photon fluence perterbation by chamber; approximately 1.022
Psheath Correction accounting for sheath protecting ionization chamber; value of 1, as chamber is waterproof
Ppol Correction factor accounting for polarity; determined in Protocol 1
Pion Correction factor accounting for ion recombination; determined in Protocol 1
PTp Correction factor acocunting for temerpature and pressure on day of experiment; determined in Protocol 1