Table 1:
Correction factors needed for determination of dose rate in Protocol 1.
| Correction Factor | Explanation |
|---|---|
| Nκ | Air kerma calibration factor |
| [(μen/ρ)Wair]water | Ration of mass energy absorption coefficients of water to air; approximately 1.05 |
| Pq,Cham | Correction accounting for chamber stem affecting photon fluence perterbation by chamber; approximately 1.022 |
| Psheath | Correction accounting for sheath protecting ionization chamber; value of 1, as chamber is waterproof |
| Ppol | Correction factor accounting for polarity; determined in Protocol 1 |
| Pion | Correction factor accounting for ion recombination; determined in Protocol 1 |
| PTp | Correction factor acocunting for temerpature and pressure on day of experiment; determined in Protocol 1 |