Table 4.
Statistical reliability test results and corresponding standard errors for three test conditions (first: Fpz+EOG test-retest comparing results of Fpz+EOG 1st Rating and Fpz+EOG 2nd Rating; second: Fpz+EOG vs. cEEGrid comparing results of Fpz+EOG 2nd Rating and cEEGrid Rating; third: Fpz+EOG vs. cEEGrid+EOG comparing results of Fpz+EOG 2nd Rating and cEEGrid+EOG Rating).
|
Fpz+EOG
test-retest |
Fpz+EOG vs.
cEEGrid |
Fpz+EOG vs.
cEEGrid+EOG |
|
|---|---|---|---|
| W | 0.78 ± 0.01 | 0.71 ± 0.01 | 0.71 ± 0.01 |
| N1 | 0.46 ± 0.02 | 0.37 ± 0.02 | 0.42 ± 0.02 |
| N2 | 0.77 ± 0.01 | 0.62 ± 0.01 | 0.75 ± 0.01 |
| N3 | 0.90 ± 0.01 | 0.85 ± 0.01 | 0.88 ± 0.01 |
| REM | 0.86 ± 0.01 | 0.69 ± 0.01 | 0.83 ± 0.01 |
| all | 0.78 ± 0.01 | 0.67 ± 0.01 | 0.75 ± 0.01 |
Combining all participants, the tests results are shown for either every single sleep stage vs. the rest or all sleep stages combined.