Table 1.

Plan quality metrics comparing 60Co‐compensators plans to baseline clinical 6 MV‐MLC plans. Five head‐and‐neck plans (top panel) and five gynecological plans (bottom panel) were generated for a 6 MV‐MLC, a 60Co‐compensator device with a 63‐cm SCD, and a 60Co compensator device with 53‐cm SCD. All compensator plans used 2 mm compensator resolution, 2 TVL maximum thickness, and 9 beams. Values shown indicate the percent deviation of the 60Co‐compensator plans from the 6 MV‐MLC plan ± one standard deviation

Difference from 6 MV‐linac plan
63 cm SCD 53 cm SCD
Head and neck endpoint
PTV D2% +0.5 ± 1.7% +1.5 ± 1.5%
PTV D99% +2.0 ± 5.4 −0.7 ± 6.3
L parotid mean dose +8.6 ± 8.9 +14.0 ± 9.2
R parotid mean dose +7.4 ± 12.2 +11.1 ± 15.1
Cord max dose +0.9 ± 6.9 +0.9 ± 5.6
Gynecological endpoint
PTV D2% −0.8 ± 0.7% −0.4 ± 0.8%
PTV D99% +2.1 ± 3.0 −0.9 ± 0.9
Rectum D60% +7.0 ± 13.1 +7.1 ± 12.4
Bladder D35% +0.2 ± 1.2 +0.2 ± 2.0