Table 2: Summary of imaging settings and whole body doses over a typical 37 fraction treatment course measured on the Clinac iX machines. TrueBeam imaging dose parameters are included, however CTDI phantom measurements were not obtained. Instead air kerma measurements were taken for each imaging protocol on both Clinac and TrueBeam machines and a further dose reduction of 50% for TrueBeam is inferred from these results (see Table 3).

Group kV mA mS mAs CTDIw100 Dose (cGy) Dose diff (%) Dose (37#) (Gy)
(mGy/mAs)
Clinac iX Imaging Parameters
Spotlight 125 80 25 704 4.09 2.9 0 1.1
Pelvis S 100 40 12 609 1.59 0.5 −83.2 0.2
Pelvis M 100 63 20 1014 1.57 1.3 −56.5 0.5
Pelvis L 110 80 20 1013 2.3 2.3 −19.2 0.9
Pelvis XL 125 80 25 1266 3.17 4 39.4 1.5
TrueBeam Imaging Parameters
Pelvis S 100 40 12 317
Pelvis M 100 63 20 832
Pelvis L 125 80 16 845
Pelvis XL 125 80 25 1320

CTDI, CT dose index.