Table 2: Summary of imaging settings and whole body doses over a typical 37 fraction treatment course measured on the Clinac iX machines. TrueBeam imaging dose parameters are included, however CTDI phantom measurements were not obtained. Instead air kerma measurements were taken for each imaging protocol on both Clinac and TrueBeam machines and a further dose reduction of 50% for TrueBeam is inferred from these results (see Table 3).
| Group | kV | mA | mS | mAs | CTDIw100 | Dose (cGy) | Dose diff (%) | Dose (37#) (Gy) |
|---|---|---|---|---|---|---|---|---|
| (mGy/mAs) | ||||||||
| Clinac iX Imaging Parameters | ||||||||
| Spotlight | 125 | 80 | 25 | 704 | 4.09 | 2.9 | 0 | 1.1 |
| Pelvis S | 100 | 40 | 12 | 609 | 1.59 | 0.5 | −83.2 | 0.2 |
| Pelvis M | 100 | 63 | 20 | 1014 | 1.57 | 1.3 | −56.5 | 0.5 |
| Pelvis L | 110 | 80 | 20 | 1013 | 2.3 | 2.3 | −19.2 | 0.9 |
| Pelvis XL | 125 | 80 | 25 | 1266 | 3.17 | 4 | 39.4 | 1.5 |
| TrueBeam Imaging Parameters | ||||||||
| Pelvis S | 100 | 40 | 12 | 317 | ||||
| Pelvis M | 100 | 63 | 20 | 832 | ||||
| Pelvis L | 125 | 80 | 16 | 845 | ||||
| Pelvis XL | 125 | 80 | 25 | 1320 | ||||
CTDI, CT dose index.