. 2021 Oct 20;196:113689. doi: 10.1016/j.bios.2021.113689

Table 1.

Clinical testings for 30 positive and 68 negative cases by qRT-PCR and NESBA.a

Diagnostic parameter qRT-PCRb
NESBA
E gene N gene E gene N gene
Positive True 30 29 30 30
False - - - -
Negative True 68 68 68 68
False - 1 - -
Sensitivity (%) (95% CIc%) 100 (88.4 to 100) 96.67 (82.8 to 99.9) 100 (88.4 to 100) 100 (88.4 to 100)
Specificity (%) (95% CI%) 100 (94.7 to 100) 100 (94.7 to 100) 100 (94.7 to 100) 100 (94.7 to 100)
a

The cases were confirmed as positive and negative based on the clinical sample analysis in Table S4.

b

The qRT-PCR was carried out using AllplexTM 2019-nCoV assay kit (Seegene, Seoul, Korea).

c

Confidence interval (MedCalc software, version 20.011).