Skip to main content
. 2021 Oct 14;12:1–6. doi: 10.1016/j.artd.2021.08.003

Table 3.

Postdynamic static leak test (phase 3) results.

Pressure (mmHg) STRATAFIX (n = 10, mL/min, ±SD) VICRYL (n = 10, mL/min, ±SD) P value
30 1.02 (±1.17) 2.15 (±2.95) .12
40 0.78 (±0.90) 2.12 (±2.70) .10
50 0.80 (±0.90) 2.11 (±3.24) .19
60 0.90 (±0.84) 2.64 (±4.51) .22
70 0.85 (±0.75) 3.64 (±6.58) .20
80 0.97 (±1.13) 4.57 (±8.88) .22