Skip to main content
. 2021 Oct 7;9:722275. doi: 10.3389/fbioe.2021.722275

TABLE 3.

Quantitative assessment of stabilization methods on the tibial plateau fracture stability.

Screw Cement injection Fragment interactions Maximum interfragmentary strains Bone volume with stress above local yield strength [mm3]
Depressed fragment Separated fragment All bone geometries
Bi-cortical Asnis III cannulated screw, Stryker Yes Mobile 0.84% 0.58% 0.00
Bonded 0.65% 0.12% 0.00
Fused 0.00
No Mobile 5.84% 1.15% 16.18
Bonded 0.74% 0.13% 0.00
Fused 0.00
Uni-cortical Asnis III cannulated screw, Stryker Yes Mobile 1.22% 0.72% 0.00
Bonded 0.74% 0.12% 0.00
Fused 0.00
No Mobile 3.42% 1.04% 21.66
Bonded 0.78% 0.07% 0.00
Fused 0.00