. 1999 Aug;37(8):2639–2647. doi: 10.1128/jcm.37.8.2639-2647.1999

TABLE 1.

Intra-assay precision of testing with 20 replicates of calibrators, controls, and human sera

Sample HBsAg concn (IU/ml)a Chemilumi-nescence signal (counts) Mean s/co ratio SD CV (%)
Negative calibrator 0 236 14 6.1
Positive calibrator 0.5 3,471 43 1.2
Negative control 0 257 0.56 0.04 6.3
Positive control 0.2 1,318 2.88 0.04 1.2
HBsAg-negative serum 263 0.49 0.03 6.9
HBsAg-negative serum spiked with HBsAg 0.05 592 1.1 0.03 2.9
HBsAg-positive serum ND 5,089 9.44 0.16 1.7
a

ND, not determined.