. 2021 Oct 7;13:746982. doi: 10.3389/fnagi.2021.746982

TABLE 4.

Intra-scanner variability results per brain structure for three different MRI systems.

Brain structures CV (%; mean ± SD) AVD (mL; mean ± SD) ICC [95% CI] DSC (mean ± SD)
Whole brain
Achieva 0.18 ± 0.14 3.11 ± 2.50 1.000 [0.999,1.000] 0.92 ± 0.00
Ingenia 0.21 ± 0.14 3.32 ± 2.17 1.000 [0.999,1.000] 0.98 ± 0.00
GE 0.29 ± 0.29 4.70 ± 4.60 0.999 [0.996,1.000] 0.98 ± 0.00
p-value 0.418 0.453 0.144
Gray matter
Achieva 0.16 ± 0.12 1.49 ± 1.13 1.000 [0.999,1.000] 0.93 ± 0.01
Ingenia 0.52 ± 0.42 5.07 ± 4.27 0.997 [0.991,0.999] 0.92 ± 0.01
GE 0.43 ± 0.42 3.82 ± 3.62 0.998 [0.995,1.000] 0.93 ± 0.01
p-value 0.051 0.045 0.822
Cortical gray matter
Achieva 0.29 ± 0.19 2.61 ± 1.75 0.999 [0.998,1.000] 0.93 ± 0.01
Ingenia 0.54 ± 0.38 4.96 ± 3.74 0.997 [0.992,0.999] 0.92 ± 0.01
GE 0.38 ± 0.48 3.24 ± 3.82 0.998 [0.995,0.999] 0.93 ± 0.02
p-value 0.287 0.229 0.893
White matter
Achieva 0.46 ± 0.18 2.97 ± 1.24 0.998 [0.995,0.999] 0.94 ± 0.01
Ingenia 1.06 ± 0.80 6.64 ± 4.61 0.991 [0.971,0.997] 0.94 ± 0.01
GE 0.74 ± 0.34 5.16 ± 2.29 0.994 [0.983,0.998] 0.94 ± 0.01
p-value 0.034 0.028 0.088
Frontal cortex
Achieva 0.60 ± 0.35 1.46 ± 0.77 0.998 [0.993,0.999] 0.90 ± 0.01
Ingenia 0.63 ± 0.55 1.52 ± 1.32 0.996 [0.989,0.999] 0.90 ± 0.02
GE 0.99 ± 0.57 2.36 ± 1.41 0.994 [0.982,0.998] 0.89 ± 0.02
p-value 0.132 0.164 0.940
Parietal cortex
Achieva 1.58 ± 1.19 2.44 ± 1.69 0.980 [0.936,0.994] 0.87 ± 0.01
Ingenia 1.52 ± 1.24 2.33 ± 1.71 0.981 [0.940,0.994] 0.87 ± 0.02
GE 1.78 ± 1.22 2.73 ± 2.17 0.980 [0.938,0.994] 0.87 ± 0.03
p-value 0.829 0.850 0.464
Temporal cortex
Achieva 1.94 ± 1.19 3.16 ± 1.88 0.964 [0.890,0.989] 0.89 ± 0.02
Ingenia 1.56 ± 1.30 2.61 ± 2.16 0.974 [0.921,0.992] 0.90 ± 0.01
GE 1.11 ± 0.74 1.82 ± 1.15 0.990 [0.970,0.997] 0.90 ± 0.02
p-value 0.148 0.151 0.165
Hippocampus, total
Achieva 1.44 ± 0.87 0.14 ± 0.09 0.986 [0.957,0.996] 0.90 ± 0.01
Ingenia 1.63 ± 0.91 0.16 ± 0.08 0.985 [0.953,0.995] 0.91 ± 0.01
GE 0.95 ± 0.67 0.09 ± 0.06 0.994 [0.981,0.998] 0.93 ± 0.01∧•
p-value 0.142 0.119 <0.001
Hippocampus, left
Achieva 2.90 ± 1.87 0.14 ± 0.09 0.956 [0.865,0.986] 0.90 ± 0.01
Ingenia 3.14 ± 2.15 0.15 ± 0.10 0.941 [0.823,0.981] 0.90 ± 0.02
GE 1.46 ± 1.14 0.07 ± 0.05 0.988 [0.963,0.996] 0.92 ± 0.01∧•
p-value 0.049 0.063 <0.001
Hippocampus, right
Achieva 1.35 ± 1.45 0.06 ± 0.06 0.988 [0.963,0.996] 0.91 ± 0.01
Ingenia 1.11 ± 0.88 0.05 ± 0.04 0.992 [0.974,0.997] 0.91 ± 0.01
GE 1.16 ± 0.71 0.06 ± 0.04 0.992 [0.973,0.997] 0.93 ± 0.01
p-value 0.848 0.957 <0.001
Lateral ventricles
Achieva 0.66 ± 0.84 0.28 ± 0.29 0.999 [0.998,1.000] 0.96 ± 0.01
Ingenia 0.75 ± 1.09 0.34 ± 0.46 0.999 [0.997,1.000] 0.96 ± 0.01
GE 1.13 ± 1.33 0.55 ± 0.78 0.997 [0.973,0.997] 0.97 ± 0.01
p-value 0.407 0.475 0.054

All volumes Mean CV (%; mean ± SD) Mean AVD (mL; mean ± SD) Mean ICC [95% CI] Mean DSC (mean ± SD)

Achieva 1.05 ± 0.87 1.63 ± 1.29 0.988 [0.977,0.998] 0.91 ± 0.01
Ingenia 1.15 ± 0.81 2.47 ± 2.30 0.987 [0.975,0.998] 0.92 ± 0.01
GE 0.95 ± 0.46 2.24 ± 1.88 0.993 [0.989,0.997] 0.93 ± 0.04

Brain volumes used to calculate measurements of precision were computed by the icobrain dm segmentation software. Coefficients of variation (CV, %), AVD (mL), intraclass correlation coefficients (ICC, [95% CI]), and Dice similarity coefficients (DSC, mean ± SD) were reported. The highest CV and lowest ICC and DSC values amongst all structures were highlighted in “bold.” Achieva: Philips Medical Systems Achieva dStream 1.5T. Ingenia: Philips Medical Systems Ingenia 3T. GE: GE Discovery MR750w 3T. p-value < 0.005 for “all scanners” differences were highlighted in “bold.” p-value < 0.005 between Achieva and GE p-value < 0.005 between Ingenia and GE.