Fig. 1.
Comparison of RT-SARS and Allplex cycle thresholds in positive SARS-CoV-2 specimens.
Twenty-nine positive samples by Allplex were retested with RT-SARS showing concordantly positive results (100 % PPA). The mean Ct difference (mean of Allplex target regions minus RT-SARS) was 9.6 Ct which is in the range of the 10 unread cycles of RT-SARS. The positive samples in the figure are presented in ascending order of Ct values by RT-SARS. Additionally, 29 SARS-CoV-2 negative specimens by Allplex retested with RT-SARS were found negative (100 % NPA).