Fig. 3.

Fig. 3

Comparison of cycle thresholds in clinical specimens positive for SARS-CoV-2 across Allplex, Alin4Plex, and AlinSARS.

Positive SARS-CoV-2 samples by Allplex were retested with Alin4Plex (n = 20) and AlinSARS (n = 11) and showed concordant positive results (100 % PPA). The mean Ct differences between the assays were 1.0 Ct (mean of Allplex target regions minus Alin4Plex), 4.6 Ct (AlinSARS minus Alin4Plex), and 4.0 Ct (AlinSARS minus mean of Allplex target regions), respectively. The positive samples in the figure are presented in ascending order of Allplex Ct values. Additionally, a set of 19 SARS-CoV-2 negative specimens by Allplex was retested by Alin4Plex confirming all negative results for SARS-CoV-2 (100 % NPA).