. 2021 Sep 26;11(10):1771. doi: 10.3390/diagnostics11101771

Table 1.

Summarized key statistical parameters for PP and CC algorithms with/without curve fitting.

Imaging Mode, Sampling Rate Algorithm Without Curve Fit With GV Curve Fit p Value
Bias [Lower Limit, Upper Limit] (mL/min) 95% CI |MQE| (%) Corr.*
(r)
Bias [Lower Limit, Upper Limit] (mL/min) 95% CI |MQE| (%) Corr.*
(r)
DSA acquisition, 6 F/s PP −176 [−842, 490] [29, 43] 0.574 −246 [−676, 184] [21, 32] 0.773 0.175
CC −234 [−691, 223] [25, 35] 0.731 −145 [−484, 195] [18, 25] 0.865 0.017 §
Fluoroscopic, 10 P/s PP −374 [−1098, 351] [42, 67] 0.545 −339 [−891, 213] [34, 51] 0.515 0.027
CC −205 [−930, 519] [43, 59] 0.534 −198 [−722, 327] [30, 44] 0.576 0.031 §

* Pearson’s correlation coefficient (r) of calculated flow rate vs. in-line flow sensor measurement. Statistical test of significance (Student’s t test) between algorithms (peak-to-peak vs. cross-correlation; without curve fitting, § with gamma-variate curve fitting, respectively).