Skip to main content
. 2021 Sep 29;9(10):240. doi: 10.3390/toxics9100240

Table 4.

Method precision using 8 duplicate runs of quality controls (QCs), µg/L.

Precision (%RSD) (n = 8)
Analyte QC Sample Target Mean Repeatability (%) Intermediate Precision (%)
Al Low Spike 7.50 7.48 2.7 3.2
High Spike 40.0 40.9 1.4 2.7
Cr Low Spike 0.375 0.360 1.8 3.7
High Spike 2.00 1.95 1.8 3.1
Fe Low spike 1.88 1.85 2.8 1.1
High spike 10.0 10.1 1.7 4.0
Co Low spike 0.375 0.390 2.8 3.7
High spike 2.00 2.09 2.1 3.8
Ni Low spike 0.375 0.394 3.5 9.6
High spike 2.00 2.01 0.9 2.1
Cu Low spike 0.750 0.744 1.7 2.9
High spike 4.00 3.97 1.2 2.8
Cd Low spike 0.0750 0.0725 2.5 2.2
High spike 0.400 0.394 1.4 2.6
Sn Low spike 0.150 0.169 2.3 13.7
High spike 0.800 0.817 1.0 1.1
Ba Low spike 0.375 0.385 2.1 0.0
High spike 2.00 2.02 0.5 2.4
Pb Low spike 0.375 0.368 1.5 0.4
High spike 2.00 2.01 1.5 0.8