TABLE 1.

Diagnostic performances of three rapid tests and LE dipstick assay compared to those of two reference standards for detection of C. trachomatis with FVU specimens from men

Rapid test Standard % Sensitivity % Specificity
Testpack Chlamydia (Abbott) EIAa 76.4 (42/55)b 93.2 (68/73)
LCRc 70.9 (44/62) 95.5 (63/66)
Surecell Chlamydia (Kodak) EIA 67.3 (37/55) 97.3 (71/73)
LCR 62.9 (39/62) 100 (66/66)
Clearview Chlamydia (Unipath) EIA 76.4 (42/55) 95.9 (70/73)
LCR 67.7 (42/62) 95.5 (63/66)
LE dipstick (Boehringer) EIA 88.6 (39/44) 89.0 (65/73)
LCR 87.5 (42/48) 92.4 (61/66)
a

Specimens positive by the Chlamydiazyme EIA confirmed with chlamydia blocking reagent. 

b

Values in parentheses are number of positive specimens/number of specimens tested. 

c

Specimens positive by LCR by the LCx Chlamydia assay as confirmed by the presence of three or more EBs by fluorescent-antibody staining.