. 1999 Sep;37(9):2793–2797. doi: 10.1128/jcm.37.9.2793-2797.1999

TABLE 1.

Precision and reproducibility of Cobas-AM

Sample no. Input HBV DNA (copies/ml) Technician 1
Technician 2
Meana (copies/ml) Rangea (copies/ml) CV (%) Within-run CV range (%) Meana (copies/ml) Rangea (copies/ml) CV (%) Within-run CV range (%)
3 200 258 128–495b 50 25–85 317 201–756b 57 19–84
4 400 544 123–1,030 45 18–97 666 115–948 34 7–64
5 825 1,064 425–2,400 50 18–79 1,562 885–2,060 28 12–33
6 1,650 2,067 1,100–3,360 33 5–35 2,600 2,060–3,210 15 10–21
9 13,300 15,525 10,400–24,000 23 10–16 21,400 17,600–29,000 14 3–22
12 106,000 97,575 75,100–128,000 16 7–19 121,335 95,000–143,000 12 4–17
a

Of 20 analyses; quadruplicates of each sample were analyzed repeatedly over 5 days by each technician. 

b

Two nonreactive results of analyses are excluded.