Skip to main content
. 2021 Sep 24;7(10):799. doi: 10.3390/jof7100799

Table 1.

Analytical performance of the MVD Histoplasma urinary Ag LFA and MVD EIA.

MVD LFA MVD EIA
Culture Culture
LFA + EIA +
+ 25 3 + 25 17
1 71 1 57
% (CI 95%)
Sensitivity 96% (80–100) 96% (80–100)
Specificity 96% (89–99) 77% (66–86)
Accuracy 96% (90–99) 82% (73–89)
Positive predictive value 89% (73–96) 60% (49–69)
Negative predictive value 99% (91–100) 98% (89–100)
Kappa 0.656 (0.509–0.803)

(LFA) Lateral flow assay; (EIA) Enzyme Immunoassay; (95% CI confidence interval); (+) positive result; (−) negative result.