Table 1.

Acquisition parameters of contrast-enhanced first-pass cardiovascular magnetic resonance angiography (CMRA) and modified Dixon (mDixon) steady-state CMRA

Parameter First-pass CMRA mDixon steady-state CMRA
Time of echo (msec) 1.60 1.93
Time of repetition (msec) 5.0 5.1
Total scan duration (min) 1:02 (15.2 s./per dynamic) 1:56
Acquired voxel size (mm) 1.7 × 1.7 × 3.4 1.20 × 1.20 × 2.40
Reconstructed voxel size (mm) 1.61 × 1.61 × 1.7 0.62 × 0.62 × 1.2
Turbo field echo factor 19
Field of view (mm) 360 × 360 × 139 240 × 240 × 96
Matrix (slices) 212 × 212 × 82 200 × 199 × 80
Compressed sensing No Yes, factor 6
Parallel imaging factor Yes. P reduction (RL) 4, S reduction (AP) 1 No
Slice orientation Coronal Coronal
Number of slices 82 80
Fat suppression technique Subtraction mDixon
Flip angle 30 20
Cardiac synchronization device ECG
Respiratory compensation Navigator respiratory compensation gating. Gating window 2 (7)
Maximum intensity projection (slice thickness/gap, mm) 3.4/1.7 4/2
Acquisition mode Cartesian Cartesian

ECG, electrocardiogram; AP, Anterior Posterior; RL, Right-left