TABLE 4. Blood Oxygen Saturation Devices Implemented Using CMOS in Research.
| [42] | [44] | [45] | [46] | [47] | [48] | [49] | [51] | |
|---|---|---|---|---|---|---|---|---|
Technology ( ) |
0.35 | 0.18 | 0.18 | 0.18 | 0.13 | 0.18 | 0.18 | 0.35 |
| Supply Voltage (V) | 3.3 | 1.8 | 1.2 | 3.3/1.8 | 1.5 | 2.5 | 3.3/1.8 | 2–3.7 |
| Oximeter Type | Transmission | Transmission | Reflective | Reflective | Reflective | Reflective | Reflective | Reflective |
| Sampling Frequency (Hz) | 100 | 165 | 4 | 0.5,1,2,4 | 100 | 100 | 40 | Continuous |
| LED Bias Current (mA) | 7.1 | 25.6 | 0.036 | 0.0033 | 50 | 50 | 9.43 | 3 mW |
| Size (mm2) | 1.265 | 1.84 | 10 | 8.45 | 10c | 2.25c | 20 | 0.64 |
Power Dissipation ( ) a
|
363 | 216b | 171b | ![]() |
68.7b | 180 | 4.6b | 14.85 |
| LED Duty Cycle (%) | 4 | 0.7 | 0.0125 | N/A | 20 | 25 | 0.07 | Continuous |
LED Power is excluded.
ADC power is included.
Estimated value.